Skip to content

MINI TREK 1012268-06URecalls & reported problems

Abbott Vascular Inc. · MINI TREK Coronary Dilatation Catheter 1.20 mm x 06 mm / Rapid-Exchange · Class 2

All models of MINI TREK (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648176531), received between Mar 12, 2020 and Jan 19, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Improper or Incorrect Procedure or Method1
  • Material Rupture1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT DURING REMOVAL OF THE 1.2X6MM MINI TREK BALLOON DILATATION CATHETER (BDC) FROM THE DISPENSER HOOP, THE SHAFT WAS NOTED TO HAVE BROKEN INTO 2 SEPARATE PIECES. ANOTHER SAME SIZE MINI TREK BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DEVICE USE OR PATIENT INVOLVEMENT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18548160 ↗
  • InjuryMar 12, 2020Material Rupture; Material Separation; Improper or Incorrect Procedure or Method; Material Rupture; Material Separation; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A LESION IN THE LEFT MAIN (LM) ARTERY. THE 1.20X6MM RX MINI TREK BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE LESION AND THE BALLOON INFLATED THREE TIMES TO 20, 15, AND 15 ATMOSPHERES (ATM) RESPECTIVELY WHEN THE BALLOON RUPTURED. THE DEVICE WAS REMOVED HOWEVER IT WAS NOTED THE SHAFT HAD SEPARATED, THE DISTAL 10CM OF THE DEVICE REMAINED IN THE ANATOMY. ATTEMPTS TO RETRIEVE THE SEPARATED PORTION WERE UNSUCCESSFUL AND THE DEVICE REMAINS IN THE ANATOMY. THE PATIENT WAS TRANSFERRED TO THE INTENSIVE CARE UNIT (ICU) TO BE FURTHER MONITORED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 9823562 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates