MINI TREK 1012268-12URecalls & reported problems
Abbott Vascular Inc. · MINI TREK Coronary Dilatation Catheter 1.20 mm x 12 mm / Rapid-Exchange · Class 2
All models of MINI TREK (25)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08717648176555), received between Sep 7, 2017 and Sep 30, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Tear, Rip or Hole in Device Packaging2
- Break1
- Crack1
- Deformation Due to Compressive Stress1
- Difficult to Advance1
- Difficult to Remove1
- Failure to Advance1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break; Failure to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949580 ↗MalfunctionDec 30, 2025Material Separation; Deformation Due to Compressive Stress; Difficult to Advance; Material Separation; Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE PERFORMED WAS TO TREAT A CALCIFIED, MILDLY TORTUOUS, LEFT CORONARY DIAGONAL ARTERY THAT IS 70% STENOSED WITH A PREVIOUSLY IMPLANTED VESSEL. DURING ADVANCEMENT OF THE 1.20X12MM MINI TREK BALLOON DILATATION CATHETER (BDC) INSIDE A GUIDELINER RESISTANCE WAS MET WITH THE GUIDELINER AND THE SHAFT OF THE MINI TREK BDC KINKED AND BROKE INTO TWO PIECES INSIDE THE GUIDELINER. THE MINI TREK BDC NEVER MADE IT OUT OF THE GUIDELINER AND WAS REMOVED AS ONE UNIT. A NEW TREK BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23928059 ↗MalfunctionJan 6, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE LEFT ANTERIOR DESCENDING CORONARY ARTERY THAT WAS MILDLY CALCIFIED AND 70% STENOSED. THE MINI TREK BALLOON RUPTURED IN THE PATIENT AT 10 ATMOSPHERES DURING THE FIRST INFLATION. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER MINI TREK WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13184103 ↗MalfunctionJul 14, 2020Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE 1.20X12MM MINI TREK POUCH WAS CREASED AND THE DEVICE STERILITY WAS COMPROMISED. THERE WAS NO DAMAGE TO THE CHIPBOARD BOX. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 10273131 ↗MalfunctionJun 29, 2020Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE MILDLY TORTUOUS RIGHT CORONARY ARTERY (RCA). THE 1.2X12 MM TREK BALLOON DILATATION CATHETER (BDC) WAS USED IN THE RCA AND THEN AFTER INFLATION IT WAS NOTED THAT THE BDC COULD NOT MOVE BACK ON THE GUIDE WIRE SO THE BDC WAS REMOVED. ANOTHER 2.0X15 MM MINI TREK BDC WAS USED IN THE RCA AND AFTER INFLATION IT WAS NOTED THE BDC COULD NOT MOVE BACK ON THE GUIDE WIRE SO ANOTHER BDC WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ISSUES WITH INFLATION OR DEFLATION WITH EITHER BALLOON. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10207372 ↗MalfunctionJun 26, 2020Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT PRIOR TO BEING GIVEN TO A HOSPITAL, THE DISTRIBUTOR NOTED THAT A 1.2X12MM MINI TREK BALLOON DILATATION CATHETER (BDC) HAD A HOLE IN THE PACKAGING POUCH, WHICH LIKELY LED TO COMPROMISED STERILITY. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 10201493 ↗MalfunctionSep 7, 2017Crack; Leak/Splash
IT WAS REPORTED THAT DURING DEVICE PREPARATION AND PRIOR TO USE IN THE ANATOMY, THE TREK LEAKED AT THE HUB AREA DUE TO A CRACK IN THE LUER. THERE WAS NO PATIENT INVOLVEMENT; THE DEVICE WAS NOT USED. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 6850590 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page