Omnilink Elite 1012632-59Recalls & reported problems

Abbott Vascular Inc. · Omnilink Elite Vascular Balloon-Expandable Stent System 9.0 mm x 59 mm x 135 cm Over-The-Wire · Class 3 · Implant

All models of Omnilink Elite (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648178566), received between Jul 17, 2019 and Nov 25, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Activation Failure1
  • Difficult to Remove1
  • Failure to Advance1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Obstruction/Occlusion1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2024Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE MILDLY TORTUOUS, HEAVILY CALCIFIED COMMON ILIAC ARTERY. WHEN ADVANCING THE 10X59MM OMNILINK ELITE STENT DELIVERY SYSTEM RESISTANCE WAS FELT WITH ANATOMY AND THE STENT DISLODGED. AN UNSPECIFIED BALLOON DILATATION CATHETER WAS ADVANCED AND INFLATED IN THE ATTEMPT TO BRING BACK THE STENT; HOWEVER, WAS UNSUCCESSFUL. THEREFORE, THE STENT WAS HELD IN PLACE WITH DEVICE AND ANOTHER 9.0X59MM OMILINK ELITE STENT WAS ATTEMPTED TO BE ADVANCED TO TREAT THE LESION; HOWEVER, MET RESISTANCE WITH ANATOMY AND DISLODGED. THE DISLODGED STENT WAS ATTEMPTED TO BE SNARED BUT THE SNARE DEVICE BECAME STUCK WITH ANATOMY AND THE STENT. IT WAS NOTED THAT THE STENTS HAD OCCLUDED IN THE PROXIMAL EXTERNAL ILIAC ARTERY BLOCKING MOST OF THE BLOOD FLOW THROUGH THE ILIAC. THE SNARE DEVICE COULD NOT BE PULLED BACK. THE SNARE DEVICE WAS BROKEN ON PURPOSE AND THE STUCK PORTION ALONG WITH BOTH STENTS WERE CRUSHED INTO THE VESSEL WALL WITH A NON-ABBOTT STENT OVER THEM. BLOOD FLOW WAS RESTORED WHEN THE STENTS WERE CRUSHED AND CLINICALLY SIGNIFICANT DELAY WAS NOTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque; Obstruction/Occlusion

    MAUDE report 20774565 ↗
  • MalfunctionJun 9, 2022Activation Failure; Activation Failure

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED VESSEL WITHOUT TORTUOSITY IN THE LEFT ILIAC ARTERY. THE 9.0X59MM OMNILINK ELITE STENT DELIVERY SYSTEM (SDS) WAS INFLATED AT 8 AND 11 ATMOSPHERES (ATMS) AND THE STENT DID NOT COME OFF THE BALLOON AND COMPLETELY FAILED TO DEPLOY. THE SDS WAS REMOVED AND A 9.0X39MM OMNI ELITE SDS WAS USED AND COMPLETED THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14644014 ↗
  • MalfunctionJul 17, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE STENT OF A 9.00 X 59 MM OMNILINK ELITE STENT DELIVERY SYSTEM (SDS) DISLODGED OFF THE BALLOON DURING USE; HOWEVER, REMAINED ON THE SDS. THERE WAS NO REPORTED RESISTANCE NOTED PRIOR TO THE DISLODGEMENT. THE STENT WAS REMOVED WITH THE SDS AND A NEW UNSPECIFIED DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8800608 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →