Omnilink Elite 1012633-59Recalls & reported problems
Abbott Vascular Inc. · Omnilink Elite Vascular Balloon-Expandable Stent System 10.0 mm x 59 mm x 135 cm Over-The-Wire · Class 3 · Implant
All models of Omnilink Elite (42)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648178603), received between Feb 22, 2018 and Nov 25, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated4
- Failure to Advance1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
- No Consequences Or Impact To Patient1
- Obstruction/Occlusion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 25, 2024Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE MILDLY TORTUOUS, HEAVILY CALCIFIED COMMON ILIAC ARTERY. WHEN ADVANCING THE 10X59MM OMNILINK ELITE STENT DELIVERY SYSTEM RESISTANCE WAS FELT WITH ANATOMY AND THE STENT DISLODGED. AN UNSPECIFIED BALLOON DILATATION CATHETER WAS ADVANCED AND INFLATED IN THE ATTEMPT TO BRING BACK THE STENT; HOWEVER, WAS UNSUCCESSFUL. THEREFORE, THE STENT WAS HELD IN PLACE WITH DEVICE AND ANOTHER 9.0X59MM OMILINK ELITE STENT WAS ATTEMPTED TO BE ADVANCED TO TREAT THE LESION; HOWEVER, MET RESISTANCE WITH ANATOMY AND DISLODGED. THE DISLODGED STENT WAS ATTEMPTED TO BE SNARED BUT THE SNARE DEVICE BECAME STUCK WITH ANATOMY AND THE STENT. IT WAS NOTED THAT THE STENTS HAD OCCLUDED IN THE PROXIMAL EXTERNAL ILIAC ARTERY BLOCKING MOST OF THE BLOOD FLOW THROUGH THE ILIAC. THE SNARE DEVICE COULD NOT BE PULLED BACK. THE SNARE DEVICE WAS BROKEN ON PURPOSE AND THE STUCK PORTION ALONG WITH BOTH STENTS WERE CRUSHED INTO THE VESSEL WALL WITH A NON-ABBOTT STENT OVER THEM. BLOOD FLOW WAS RESTORED WHEN THE STENTS WERE CRUSHED AND CLINICALLY SIGNIFICANT DELAY WAS NOTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Device Embedded In Tissue or Plaque; Obstruction/Occlusion
MAUDE report 20774720 ↗MalfunctionJan 11, 2023Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING PREPARATION OF THE 10.00X59MM OMNILINK ELITE STENT DELIVERY SYSTEM (SDS) THE WIRE LUMEN WAS FLUSHED; HOWEVER, NEGATIVE PRESSURE WAS NOT HELD. THE PROCEDURE WAS TO TREAT A LESION IN THE COMMON ILIAC ARTERY. WHILE BACKLOADING THE DEVICE ONTO THE GUIDEWIRE, OUTSIDE THE PATIENT ANATOMY, AND BEFORE ENTERING THE ROTATING HEMOSTATIC VALVE (RHV), THE STENT BECAME PARTIALLY DISLODGED FROM THE BALLOON AND WAS NOT USED IN THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. ANOTHER SAME SIZE OMNILINK WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16142824 ↗MalfunctionMar 30, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MILDLY CALCIFIED LESION IN THE RIGHT ILIAC ARTERY. A 10X59MM OTW OMNI ELITE BALLOON EXPANDABLE STENT SYSTEM (BESS) WAS PREPARED PER THE INSTRUCTIONS FOR USE WITHOUT ISSUES. THE BESS WAS ADVANCED ON AN UNKNOWN GUIDE WIRE; HOWEVER, IT WAS NOTED THAT THE STENT WAS LOOSE ON THE BESS BEFORE IT WAS ADVANCED THROUGH THE INTRODUCER SHEATH. THE BESS WAS SIMPLY REMOVED FROM THE GUIDE WIRE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 10X59MM OTW OMNI ELITE STENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593090 ↗MalfunctionFeb 22, 2018Device Dislodged or Dislocated
IT WAS REPORTED THAT WHILE LOADING AN OMNI ELITE STENT DELIVERY SYSTEM ONTO THE GUIDE WIRE, THE DISTAL END OF THE DEVICE WAS HANDLED AND STENT WAS POSSIBLY PULLED; HOWEVER, THIS COULD NOT BE CONFIRMED. JUST AFTER LOADING AND BEFORE ADVANCING THE DEVICE INTO THE ANATOMY, THE STENT DISLODGED. THE DEVICE WAS SET ASIDE AND NOT USED FURTHER. ANOTHER DEVICE WAS USED IN REPLACEMENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY. THERE WAS NO ADDITIONAL INFORMATION PROVIDED REGARDING THIS ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7289929 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page