Omnilink Elite 1012626-59Recalls & reported problems
Abbott Vascular Inc. · Omnilink Elite Vascular Balloon-Expandable Stent System 10.0 mm x 59 mm x 80 cm Over-The-Wire · Class 3 · Implant
All models of Omnilink Elite (42)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648179396), received between Jan 2, 2019 and Dec 6, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Activation Failure1
- Device Dislodged or Dislocated1
- Difficult to Advance1
- Difficult to Remove1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 6, 2019Difficult to Advance; Activation Failure; Difficult to Advance; Activation Failure
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE HEAVILY CALCIFIED ILIAC ARTERY. THE 10.0 X 59 MM OMNILINK ELITE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED, WITH RESISTANCE, TO THE TARGET LESION. THE SDS BALLOON WAS ATTEMPTED TO BE INFLATED, BUT THE STENT REMAINED ON THE STENT DELIVERY SYSTEM. THE SDS WAS REMOVED WITHOUT RESISTANCE AND THE STENT REMAINED ON THE DELIVERY SYSTEM. ADDITIONAL PREDILATATION WAS PERFORMED AND A SECOND OMNILINK ELITE WAS USED WITHOUT ISSUE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9434299 ↗MalfunctionJan 2, 2019Difficult to Remove; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated; Difficult to Remove; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED COMMON ILIAC ARTERY AND PLAQUE AT THE OSTIUM WAS NOTED. PRE-DILATATION WAS NOT PERFORMED. A 10X59 MM OMNILINK ELITE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED SLIGHTLY PAST THE LESION AND WHEN PULLING THE SDS BACK, RESISTANCE WAS FELT WITH THE ANATOMY AND THE STENT PARTIALLY CAME OFF THE BALLOON APPROXIMATELY 5MM. THE PHYSICIAN MADE THE DECISION TO DEPLOY THE STENT AT THAT LOCATION TO PREVENT DISLODGEMENT OF THE STENT. THE STENT WAS PARTIALLY IN THE LESION AND PARTIALLY IN HEALTHY TISSUE. A 10X29 MM OMNILINK ELITE STENT WAS IMPLANTED TO SUCCESSFULLY TREAT THE REMAINING PORTION OF THE LESION. THERE WAS NO ADVERSE PATIENT EFFECT AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8209503 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page