Omnilink Elite 1012629-29Recalls & reported problems

Abbott Vascular Inc. · Omnilink Elite Vascular Balloon-Expandable Stent System 6.0 mm x 29 mm x 135 cm Over-The-Wire · Class 3 · Implant

All models of Omnilink Elite (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08717648179525), received between Mar 2, 2018 and Sep 24, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Failure to Advance4
  • Difficult to Remove2
  • Device Operates Differently Than Expected1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2025Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A TYPE 2 AORTIC DISSECTION EXTENDING INTO THE LEFT RENAL ARTERY, EXTENDING INTO THE SUPERIOR MESENTERIC ARTERY (SMA) AND ILIAC ARTERIES. A 6.0X29MM OMNILINK ELITE STENT DELIVERY SYSTEM WAS ATTEMPTED TO BE ADVANCED; HOWEVER FAILED TO ADVANCE ACROSS THE SMA DUE TO THE DISSECTION. DURING REMOVAL RESISTANCE WITH FELT WITH THE NON-ABBOTT SHEATH AND THE STENT DISLODGED JUST ABOVE THE NON-ABBOTT SHEATH IN THE AORTA. THE STENT WAS REMOVED VIA SNARE. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23137201 ↗
  • MalfunctionAug 3, 2023Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY CALCIFIED LEFT EXTERNAL ILIAC ARTERY. AN UNSPECIFIED STENT WAS IMPLANTED AND A 6.0X29MM OMNILINK STENT WAS ATTEMPTED TO BE ADVANCE THROUGH THE STENT; HOWEVER, FELT RESISTANCE WITH THE PREVIOUSLY IMPLANTED STENT AND THE PHYSICIAN DECIDED TO REMOVE THE STENT. ONCE REMOVED IT WAS NOTED THAT THE STENT WAS STILL ON THE DELIVERY CATHETER; HOWEVER, NOTED TO BE LOOSE. THEREFORE, THE UNSPECIFIED SHEATH WAS REPOSITIONED AND A NEW 6X39MM OMNILINK STENT CROSSED WITHOUT ISSUE AND WAS SUCCESSFULLY IMPLANTED. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17452249 ↗
  • MalfunctionApr 19, 2023Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY CALCIFIED, 95% STENOSED LESION IN THE SUBCLAVIAN ARTERY. A 6.0X29MM OMNI ELITE VASCULAR BALLOON-EXPANDABLE STENT (BES) WAS ADVANCED BUT COULD NOT CROSS THE ANATOMY TO REACH THE TARGET LESION. DURING WITHDRAW OF THE BES, THE STENT BECAME CAUGHT ON THE 6F SHEATH, AND THE STENT AND THE SHEATH WERE REMOVED AS A SINGLE UNIT WITHOUT ISSUE. A SAME SIZE NON-ABBOTT DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAYS IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16770488 ↗
  • MalfunctionDec 3, 2019Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MODERATELY TORTUOUS AND HEAVILY CALCIFIED LESION IN THE RIGHT COMMON ILIAC ARTERY. A 6.0X29MM OTW OMNI ELITE BALLOON EXPANDABLE STENT SYSTEM (BESS) WAS BEING ADVANCED; HOWEVER, FAILED TO CROSS DUE TO RESISTANCE WITH THE ANATOMY. THE BESS WAS REMOVED FROM THE PATIENT ANATOMY WITHOUT RESISTANCE. ONCE THE BESS WAS OUTSIDE THE PATIENT ANATOMY IT WAS NOTED THAT THE STENT WAS MISSING. THE STENT WAS LOCATED INSIDE THE DISTAL TIP OF THE 6F SHEATH. THE SHEATH WAS REMOVED WITH THE STENT IN IT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 6F SHEATH AND A 7X27MM UNSPECIFIED NON-ABBOTT STENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9405597 ↗
  • MalfunctionMar 2, 2018Device Operates Differently Than Expected; Material Deformation

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN 80% STENOSED AND HEAVILY CALCIFIED LESION IN THE ILIAC ARTERY. IT WAS NOTICED THAT THE STENT OF A 6X29MM OMNI ELITE UNRAVELED (FLARED) WHILE ATTEMPTING TO INSERT IT INTO THE SHEATH. IT WAS CONFIRMED THAT THE STENT DID NOT DISLODGE; HOWEVER, THE SPECIFIC STENT DAMAGE COULD NOT BE CONFIRMED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW OMNI ELITE STENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED. THE RETURN DEVICE ANALYSIS FOUND THAT THE STENT WAS STRETCHED OUT AND DISLOCATED ON THE BALLOON YET STATIONARY, WITH THE PROXIMAL END OF STENT AT 12.5MM PROXIMAL TO DISTAL BALLOON MARKER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7311214 ↗