NC TREK NEO™ 1400275-08Recalls & reported problems
Abbott Vascular Inc. · NC TREK NEO™ Coronary Dilatation Catheter 2.75 mm x 8 mm / Rapid-Exchange · Class 2
All models of NC TREK NEO (45)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08717648232459), received between Dec 15, 2023 and Sep 29, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture10
- Difficult to Advance7
- Improper or Incorrect Procedure or Method1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Material Rupture
It was reported that the procedure was to treat the left circumflex coronary artery with heavy calcification, moderate tortuosity and 90% stenosis. The 2.75x8 mm NC Trek Neo balloon dilatation catheter (BDC) was inflated 4 times to 12 atmospheres and ruptured. Another device was used to complete the procedure. There were no adverse patient effects. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935001 ↗MalfunctionMar 12, 2026Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 90% STENOSED LESION IN THE RIGHT CORONARY ARTERY (RCA) WITH MILD CALCIFICATION AND MILD TORTUOSITY. FOR POST-DILATATION, THE 2.75X8MM NC TREK NEO BALLOON DILATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION WITH RESISTANCE DUE TO THE ANATOMY. THE BALLOON WAS INFLATED; HOWEVER, THE BALLOON RUPTURED DURING THE INITIAL INFLATION AT 6 ATMOSPHERES (ATM). THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS CONTINUED WITH ANOTHER TREK BDC. THERE WAS NO ADVERSE PATIENT EFFECT, AND NO CLINICALLY SIGNIFICANT PROCEDURAL DELAY WAS REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24587047 ↗MalfunctionFeb 12, 2026Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 90% STENOSED DE NOVO LESION IN THE LEFT ANTERIOR DESCENDING (LAD) ARTERY. FOR POST-DILATATION, THE 2.75X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS ADVANCED WITH RESISTANCE WITH THE UNSPECIFIED IMPLANTED STENT. THE BALLOON WAS INFLATED; HOWEVER, THE DURING THE INITIAL INFLATION, THE BALLOON RUPTURED AT 12 ATMOSPHERES (ATM). THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS CONTINUED WITH ANOTHER NC TREK BALLOON. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24322482 ↗MalfunctionOct 24, 2025Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A DE NOVO, HEAVILY CALCIFIED, MILDLY TORTUOUS RIGHT CORONARY ARTERY WITH 90% STENOSIS. THE 2.75 X 8MM TREK BALLOON DILATATION CATHETER (BDC) WAS PREPPED OUTSIDE THE ANATOMY PRIOR TO USE. THE TREK BDC WAS ADVANCED WITH RESISTANCE NOTED. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 14 ATMOSPHERES. THE PROCEDURE WAS COMPLETED WITH ANOTHER TREK BDC. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23381252 ↗MalfunctionAug 12, 2025Inflation Problem; Improper or Incorrect Procedure or Method; Inflation Problem; Improper or Incorrect Procedure or Method
IT WAS REPORTED THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED AND TORTUOUS LESION IN THE RIGHT CORONARY ARTERY. THE 2.75X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS PREPARED FOR USE OUTSIDE THE PATIENT ANATOMY WITHOUT ISSUE; BUT WAS NOT SOAKED PRIOR TO USE. THE BDC WAS ADVANCED WITH NO RESISTANCE AND INFLATED TO NOMINAL PRESSURE; HOWEVER, THE BALLOON WOULD INFLATE. THE DEVICE WAS REMOVED AND ANOTHER NC TREK NEO WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. RETURN DEVICE ANALYSIS NOTED A TEAR IN THE OUTER MEMBER ON THE SIDE OF THE GUIDE WIRE EXIT NOTCH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22781479 ↗MalfunctionJun 3, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A DE NOVO LESION IN THE MID LEFT ANTERIOR DESCENDING (MLAD) WITH 90% STENOSIS, MODERATE CALCIFICATION AND MODERATE TORTUOSITY. THE 2.75X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS ADVANCED WITHOUT RESISTANCE BUT RUPTURED DURING THE FIRST INFLATION AT AROUND 10 ATMOSPHERES (ATMS). ANOTHER NON-ABBOTT BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22138201 ↗MalfunctionNov 6, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE PERFORMED WAS TO TREAT A HEAVILY CALCIFIED, MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING ARTERY THAT IS 99% STENOSED. THE 2.75X8MM NC TREK NEO BALLOON DILATATION MET CATHETER (BDC) MET RESISTANCE WITH ANATOMY DURING ADVANCEMENT; HOWEVER, ONCE AT THE LESION THE BDC WAS INFLATED. AT THE SECOND INFLATION THE BALLOON RUPTURED AT 8 ATMOSPHERES. ANOTHER NC TREK BALLOON WAS USED TO CONTINUE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20625399 ↗MalfunctionAug 26, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 90% STENOSED DE NOVO LESION IN THE PROXIMAL LEFT ANTERIOR DESCENDING (LAD) ARTERY WITH MODERATE CALCIFICATION AND MILD TORTUOSITY. THE 2.75X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION WITH RESISTANCE DUE TO THE ANATOMY. THE BALLOON WAS INFLATED; HOWEVER, THE BALLOON RUPTURED AT 12 ATMOSPHERES (ATM) DURING THE FIRST INFLATION. THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20065607 ↗MalfunctionAug 5, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT IN STENT RESTENOSIS (ISR) IN THE MID LEFT ANTERIOR DESCENDING (LAD) CORONARY ARTERY WITH HEAVY CALCIFICATION, MILD TORTUOSITY AND 90% STENOSIS. THE 2.75X8 MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE LESION WITH RESISTANCE WITH THE ANATOMY NOTED. THE BDC WAS USED FOR PRE-DILATATION, BUT THE BALLOON RUPTURED DURING THE FIRST INFLATION TO 10 ATMOSPHERES. A NON-ABBOTT BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19906949 ↗MalfunctionMay 8, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 90% STENOSED DE NOVO LESION IN THE MID LEFT ANTERIOR DESCENDING (MLAD) ARTERY WITH HEAVY CALCIFICATION AND MILD TORTUOSITY. FOR PRE-DILATATION, THE 2.75X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION WITH RESISTANCE DUE TO THE LESION. THE BALLOON WAS INFLATED; HOWEVER, THE BALLOON RUPTURED AT 12 ATMOSPHERES (ATM) AT THE FIRST INFLATION. THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. A NON-ABBOTT BALLOON FOLLOWED BY IMPLANTATION OF A XIENCE STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19274111 ↗MalfunctionDec 15, 2023Material Rupture; Material Rupture
IT WAS REPORTED THE 2.75 X 8MM NC TREK NEO RX WAS TO BE USED IN THE PROXIMAL LEFT ANTERIOR DESCENDING (PLAD) LESION. THE DEVICE WAS INFLATED FOR 6 SECONDS, BUT RUPTURED AT 8 ATMOSPHERES (ATM) DURING THE FIRST INFLATION. THEREFORE, THE DEVICE WAS REMOVED AND A NC EUPHORA OF THE SAME SIZE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18339736 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page