NC TREK NEO™ 1400325-08Recalls & reported problems
Abbott Vascular Inc. · NC TREK NEO™ Coronary Dilatation Catheter 3.25 mm x 8 mm / Rapid-Exchange · Class 2
All models of NC TREK NEO (45)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08717648232565), received between Jul 14, 2023 and Oct 23, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture5
- Difficult to Advance3
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 23, 2025Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE NC TREK BALLOON DILATATION CATHETER (BDC) WAS PREPARED FOR USE ACCORDING TO THE INSTRUCTIONS FOR USE AND NO ISSUES OCCURRED. THE NC TREK BDC WAS ADVANCED TO THE TARGET LESION FOR POST-DILATION IN THE MODERATELY CALCIFIED, MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING CORONARY ARTERY AND RESISTANCE FROM THE ANATOMY WAS NOTED. DURING THE SECOND INFLATION TO 8 ATMOSPHERES (ATM), THE BALLOON RUPTURED. THE BDC WAS THEN REMOVED FROM THE PATIENT. THE PROCEDURE WAS CONTINUED USING ANOTHER NC TREK BALLOON. NO ADVERSE EVENTS OCCURRED TO THE PATIENT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374629 ↗MalfunctionJul 24, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE LEFT MAIN TRUNK TO PROXIMAL LEFT ANTERIOR DESCENDING ARTERY WITH MODERATE CALCIFICATION, MILD TORTUOSITY AND 75% STENOSIS. AFTER A NON-ABBOTT 3.0MM STENT WAS DEPLOYED, THE 3.25X8 MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS USED FOR PRE-DILATATION, BUT THE BALLOON RUPTURED DURING THE FIRST INFLATION TO 12 ATMOSPHERES. A NON-ABBOTT BALLOON WAS USED TO CONTINUE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19818535 ↗MalfunctionJun 25, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO A MID RIGHT CORONARY ARTERY THAT IS 75% STENOSED. A 3.0MM STENT WAS DEPLOYED AND 3.25X8MM NC TREK NEO BALLOON DILATATION CATHETER WAS ATTEMPTED TO BE USED FOR POST DILATATION; HOWEVER THE BDC RUPTURED AT 12 ATMOSPHERES DURING THE FIRST INFLATION. ANOTHER 3.5MM BALLOON WAS USED TO SUCCESSFULLY POST DILATE THE STENT AT 16 ATMOSPHERES. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19606978 ↗MalfunctionMar 12, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 90% STENOSED DE NOVO LESION IN THE PROXIMAL RIGHT CORONARY ARTERY (PRCA) WITH MODERATE CALCIFICATION AND MODERATE TORTUOSITY. FOR POST-DILATATION, 3.25X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION WITH RESISTANCE DUE TO THE ANATOMY AND THE BALLOON WAS INFLATED; HOWEVER, THE BALLOON RUPTURED AT 10 ATMOSPHERES (ATM) DURING THE FIRST INFLATION. THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18891251 ↗MalfunctionJul 14, 2023Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A DE NOVO LESION IN THE PROXIMAL RIGHT CORONARY ARTERY (PRCA) WITH 90% STENOSIS, MILD CALCIFICATION AND MODERATE TORTUOSITY. A GUIDE WIRE CROSSED THE LESION AND A 2.75X13MM BALLOON WAS USED FOR PRE-DILATATION. A XIENCE SKYPOINT STENT WAS DEPLOYED. THE 3.25X8MM NC TREK NEO BALLOON DILATATION CATHETER (BDC) HAD RESISTANCE DURING ADVANCEMENT AND WAS USED TO DILATE THE PROXIMAL END OF THE STENT DEPLOYED, HOWEVER, RUPTURED DURING THE FIRST INFLATION AT 10 ATMOSPHERES (ATMS). THE CATHETER WAS SOAKED IN SALINE AND THE DEVICE WAS PREPPED (AIR ASPIRATION) OUTSIDE THE ANATOMY PRIOR TO USE. NO RESISTANCE WAS MET WHEN THE PROTECTIVE SHEATH WAS REMOVED. THERE WAS NO RESISTANCE UPON REMOVAL OF THE NC TREK NEO BALLOON. ANOTHER 3.25X10MM NON-ABBOTT BALLOON WAS USED FOR POST DILATATION PER STANDARD PROCEDURE AND COMPLETED THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17329561 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page