XIENCE Skypoint™ 1804300-08Recalls & reported problems
Abbott Vascular Inc. · XIENCE Skypoint™ Everolimus Eluting Coronary Stent System 3.00 mm x 08 mm / Rapid-Exchange · Class 3 · Implant
All models of XIENCE Skypoint (69)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 08717648233210), received between Jan 4, 2022 and Feb 25, 2025 — classified as 4 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove5
- Failure to Advance5
- Device Dislodged or Dislocated2
- Material Deformation2
- Activation Failure1
- Adverse Event Without Identified Device or Use Problem1
- Improper or Incorrect Procedure or Method1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Thrombosis/Thrombus2
- Angina1
- Myocardial Infarction1
- Non specific EKG/ECG Changes1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryFeb 25, 2025Patient-Device Incompatibility; Patient-Device Incompatibility
SUBSEQUENT TO THE INITIAL FILED REPORT, THE FOLLOWING INFORMATION WAS PROVIDED: THE XIENCE SKYPOINT STENT WAS IMPLANTED ON (B)(6) 2022. THE DATE THE PATIENT RETURNED TO THE HOSPITAL WITH SUDDEN ONSET ANGINA IS (B)(6) 2022. PERCUTANEOUS CORONARY INTERVENTION (PCI) WAS PERFORMED AS TREATMENT. THE DRAGONFLY DEVICE WAS AN OPTIS AND NOT AN OPSTAR. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Angina; Non specific EKG/ECG Changes; Thrombosis/Thrombus
MAUDE report 21462611 ↗MalfunctionApr 2, 2024Stretched; Device Dislodged or Dislocated; Material Deformation; Stretched; Device Dislodged or Dislocated; Material Deformation
USER FACILITY MEDWATCH REPORT [MW5152005] RECEIVED THAT STATES: CATHETER DEFECTIVE AT JUNCTION OF RX. IT WAS REPORTED PRIOR TO THE PROCEDURE, THE 3.0X08MM XIENCE SKYPOINT STENT WAS NOTED TO BE DEFECTIVE AND STARTED TO UNRAVEL. THE DEVICE WAS NOT USED. AN UNSPECIFIED DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19026965 ↗InjurySep 1, 2023Material Rupture; Improper or Incorrect Procedure or Method; Activation Failure; Material Rupture; Improper or Incorrect Procedure or Method; Activation Failure
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE CIRCUMFLEX CORONARY ARTERY WITH NO CALCIFICATION AND NO TORTUOSITY. THE 3.0X08 XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LESION AND THEN PREPPED, AND THEN INFLATION WAS ATTEMPTED. THE BALLOON WAS ABLE TO GET TO 16 ATMOSPHERES AFTER ONE INFLATION; HOWEVER, THE DEVICE WAS NOT HOLDING PRESSURE AND A RUPTURE WAS NOTED. THE STENT WAS 90% DEPLOYED. A NON-COMPLIANT (NC) BALLOON WAS USED FOR POST-DILATATION TO COMPLETE DEPLOYMENT OF THE STENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17671692 ↗InjuryDec 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT EXPERIENCED A NON-ST ELEVATION MYOCARDIAL INFARCTION (NSTEMI) AND 2 XIENCE SKYPOINT STENTS (3.0X08MM, 2.5X15MM) WERE IMPLANTED. ON (B)(6) 2022, THE PATIENT EXPERIENCED POST-PROCEDURAL CARDIAC DISTURBANCES AND AN ST ELEVATION MYOCARDIAL INFARCTION. THE PATIENT WAS HOSPITALIZED. THROMBUS MAY HAVE BEEN NOTED IN ONE OF THE STENTS, SO PERCUTANEOUS CORONARY INTERVENTION WAS PERFORMED. ON (B)(6) 2022, THE PATIENT EXPERIENCED A NON-ST ELEVATION MYOCARDIAL INFARCTION. STENOSIS WAS NOTED IN AN UNSPECIFIED STENT, UNRELATED TO THE XIENCE SKYPOINT STENTS AND A NON-ABBOTT STENT WAS IMPLANTED. THERE WAS NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Myocardial Infarction; Thrombosis/Thrombus
MAUDE report 16038023 ↗MalfunctionNov 17, 2022Difficult to Remove; Failure to Advance; Material Deformation; Difficult to Remove; Failure to Advance; Material Deformation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A RIGHT CORONARY ARTERY (RCA). THE 3.0X08MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) FAILED TO CROSS; HOWEVER, THE CAUSE IS UNKNOWN. RESISTANCE WAS NOTED DURING REMOVAL AND THE STENT BECAME BENT AS THE PROXIMAL EDGE BECAME CAUGHT ON SOMETHING. ANOTHER UNSPECIFIED XIENCE SKYPOINT STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED AND NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15811719 ↗InjuryJul 29, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LESION IN THE PROXIMAL LEFT ANTERIOR DESCENDING (PLAD) ARTERY. A 3X8MM XIENCE SKYPOINT DRUG ELUTING STENT (DES) WAS ADVANCED THROUGH A XIENCE STENT THAT WAS IMPLANTED 2 DAYS EARLIER. WHEN PULLING BACK TO POSITION THE STENT AT THE TARGET LESION, THE 3X8MM XIENCE SKYPOINT STENT INTERACTED WITH THE PREVIOUSLY IMPLANTED STENT AND DISLODGED FROM THE BALLOON, DISTAL TO THE IMPLANTED STENT. A SNARE WAS USED, AND THE DISLODGED STENT WAS SUCCESSFULLY REMOVED. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE LESION WAS BALLOONED WITH AN UNSPECIFIED BALLOON TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15136368 ↗MalfunctionJul 12, 2022Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE LEFT ANTERIOR DESCENDING (LAD) ARTERY WITH 90% STENOSIS AND MODERATE CALCIFICATION. THE 3.0X8MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) MET RESISTANCE WITH THE LESION DURING ADVANCEMENT AND FAILED TO CROSS THE LESION DUE TO THE ANATOMY. THERE WAS RESISTANCE WITH THE LESION DURING REMOVAL. THE SDS WAS REMOVED INDEPENDENTLY. THERE WERE NO OTHER DEVICE ISSUES SUCH AS SHAFT SEPARATION OR STENT DISLODGEMENT. A NON-ABBOTT STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14991154 ↗MalfunctionJun 29, 2022Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 99% STENOSED, HEAVILY CALCIFIED AND HEAVILY TORTUOUS VESSEL IN THE MID LEFT ANTERIOR DESCENDING (MLAD) ARTERY. THE 3.0X8MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) MET RESISTANCE WITH THE ANATOMY DURING ADVANCEMENT AND FAILED TO CROSS THE LESION. THERE WAS NO DEVICE INTERACTION AND THERE WERE NO OTHER DEVICE ISSUES SUCH AS SHAFT SEPARATION OR STENT DISLODGEMENT. THE SDS WAS REMOVED INDEPENDENTLY; HOWEVER, THERE WAS RESISTANCE WITH THE ANATOMY DURING REMOVAL. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14855904 ↗MalfunctionJun 23, 2022Difficult to Remove; Material Too Rigid or Stiff; Failure to Advance; Difficult to Remove; Material Too Rigid or Stiff; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED, HEAVILY TORTUOUS LEFT CIRCUMFLEX ARTERY THAT IS 99% STENOSED. DURING ADVANCEMENT OF THE 3.8X08MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) MET RESISTANCE WITH ANATOMY AS WELL AS DURING REMOVAL OF THE SDS. IN THE PHYSICIAN'S OPINION THE NO CROSS WAS ALSO DUE TO THE STENT NOT BEING AS FLEXIBLE. THE PROCEDURE WAS COMPLETED WITH A NON-ABBOTT STENT. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14793410 ↗MalfunctionJan 4, 2022Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A 99% STENOSED, HEAVILY CALCIFIED AND HEAVILY TORTUOUS LESION IN THE PROXIMAL CIRCUMFLEX (CX) CORONARY ARTERY. THE 3.0 X 8 MM XIENCE SKYPOINT STENT DELIVERY SYSTEM WAS ADVANCED TO THE TARGET LESION BUT COULD NOT CROSS THE LESION AND DURING REMOVAL OF THE SDS THERE WAS RESISTANCE WITH THE LESION. A NEW NON-ABBOTT SDS WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13154372 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page