XIENCE Skypoint™ 1804400-08Recalls & reported problems
Abbott Vascular Inc. · XIENCE Skypoint™ Everolimus Eluting Coronary Stent System 4.00 mm x 08 mm / Rapid-Exchange · Class 3 · Implant
All models of XIENCE Skypoint (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648233456), received between Jul 19, 2021 and Oct 22, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove2
- Failure to Advance2
- Material Separation1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 22, 2024Migration; Migration
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE LEFT MAIN CORONARY ARTERY WITH MILD CALCIFICATION AND WHERE A PREVIOUSLY IMPLANTED, UNSPECIFIED STENT, WAS LOCATED. THE 4.0X8MM XIENCE SKYPOINT STENT WAS DEPLOYED TO OVERLAP THE IMPLANTED STENT; HOWEVER, THE OVERLAP WAS MISSED BY 1MM. THE IMPLANTED STENT REMAINED WITHIN THE TARGET LESION AND NOT IN HEALTH TISSUE. POST-DILATION WAS PERFORMED WITH AN UNSPECIFIED 4.5X08 NON-COMPLIANT BALLOON WITHOUT ISSUE. INTRAVASCULAR ULTRASOUND (IVUS) WAS THEN USED, NOTING THAT THE STENT HAD MIGRATED OVER THE TOP OF THE GUIDING CATHETER AND INTO THE FEMORAL ARTERY WHERE IT REMAINED. AN UNSPECIFIED STENT WAS IMPLANTED IN THE TARGET LESION. ADDITIONAL HOSPITALIZATION OCCURRED FOR VASCULAR SURGERY TO REMOVE THE MIGRATED STENT BY CUT DOWN. THERE WERE NO ADVERSE PATIENT SEQUELA. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20507179 ↗MalfunctionJul 15, 2022Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE LEFT ANTERIOR DESCENDING CORONARY ARTERY THAT WAS BOTH MILDLY CALCIFIED AND TORTUOUS AND 90% STENOSED. THE 4.00X08MM XIENCE SKYPOINT STENT SYSTEM WAS ADVANCED, BUT MET RESISTANCE AND FAILED TO REACH THE LESION. DURING REMOVAL OF THE DEVICE, RESISTANCE WAS MET, BUT THE DEVICE WAS REMOVED WITHOUT ISSUE. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. A NON-ABBOTT STENT WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15032908 ↗MalfunctionApr 26, 2022Material Separation; Material Separation
IT WAS REPORTED THAT DURING PREPARATION OF THE 4.0 X 8 MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) , WHEN THE STYLET WAS PULLED OUT OF THE SDS, THE STYLET WAS NOTED TO BE BROKEN IN TWO PIECES. THEREFORE, THE SDS WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14207421 ↗MalfunctionJul 19, 2021Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED, HEAVILY TORTUOUS MID RIGHT CORONARY ARTERY THAT IS 100% STENOSED. A 4.0X08MM XIENCE SKYPOINT WAS ATTEMPTED TO BE ADVANCE; HOWEVER, MET RESISTANCE WITH ANATOMY. IT WAS NOTED DURING REMOVAL RESISTANCE WAS ALSO MET WITH ANATOMY. THE PROCEDURE WAS COMPLETED WITH A NON-ABBOTT STENT. THERE WAS NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY REPORTED. SUBSEQUENTLY AFTER INITIAL WAS FILED IT WAS CONFIRMED THE DEVICE FAILED TO CROSS DUE TO ANATOMY AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12193207 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page