XIENCE Skypoint™ 1804500-15Recalls & reported problems

Abbott Vascular Inc. · XIENCE Skypoint™ Everolimus Eluting Coronary Stent System 5.00 mm x 15 mm / Rapid-Exchange · Class 3 · Implant

All models of XIENCE Skypoint (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648233630), received between Aug 2, 2022 and Nov 27, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove3
  • Difficult to Advance2
  • Device Dislodged or Dislocated1
  • Failure to Advance1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2023Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE MILDLY CALCIFIED, MILDLY TORTUOUS PROXIMAL LEFT ANTERIOR DESCENDING ARTERY (LAD) WITH 90% STENOSIS. AFTER IMPLANTING A 4.5X23MM XIENCE SKYPOINT STENT, A 5.0X15MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) WAS ATTEMPTED TO BE ADVANCED THROUGH A NON-ABBOTT GUIDE CATHETER; HOWEVER, BECAME STUCK IN THE GUIDING CATHETER AND COULD NOT BE REMOVED. THEREFORE, THE XIENCE SKYPOINT AND GUIDING CATHETER WERE REMOVED TOGETHER AS ONE UNIT. A 4.5X15MM XIENCE SKYPOINT WAS IMPLANTED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18213185 ↗
  • MalfunctionMay 24, 2023Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT LESIONS IN THE MID AND PROXIMAL RIGHT CORONARY ARTERY (RCA) WITH SEVERE CALCIFICATION. THE LESIONS WERE PRE-DILATED. A 5X15 MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) FAILED TO CROSS THE PROXIMAL RCA LESION DUE TO THE ANATOMY AND THE STENT WAS DISLODGED BUT REMAINED ON THE SDS. THE DEVICE WAS REMOVED WITH RESISTANCE WITH A NON-ABBOTT GUIDE EXTENSION CATHETER. A SAME SIZED SDS WAS USED TO TREAT THE LESION. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16994764 ↗
  • MalfunctionAug 2, 2022Difficult to Remove; Improper or Incorrect Procedure or Method; Difficult to Advance; Difficult to Remove; Improper or Incorrect Procedure or Method; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE LEFT MAIN (LM) ARTERY WITH 75% STENOSIS, NO CALCIFICATION AND NO TORTUOSITY. THE 5.0X15MM XIENCE SKYPOINT STENT DELIVERY SYSTEM (SDS) MET RESISTANCE AND FAILED TO CROSS DUE TO DEVICE INTERACTION WITH THE GUIDE WIRE. AFTER REMOVAL, AN ATTEMPT WAS MADE AND FAILED TO CROSS THE LESION AGAIN. THERE WAS RESISTANCE DURING REMOVAL WITH THE GUIDE WIRE. ALL DEVICES WERE REMOVED AS A SINGLE UNIT. THERE WERE NO OTHER DEVICE ISSUES SUCH AS SHAFT SEPARATION AND STENT DISLODGEMENT. A NON-ABBOTT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15148168 ↗