Technomed Europe TE/S50715-001Recalls & reported problems
Technomed Engineering B.V. · Disposable Subdermal Needle Electrode, corkscrew, 4 x 6 colors, 1.2m wire · Class 2
All models of Technomed Europe (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08718375863442), received between Nov 15, 2021 and Dec 10, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Dec 8, 2021Class IITerminated
There is a high risk of the blue hub detaching from the corkscrew needle, resulting the the needle remaining in the patient's scalp.
Z-0465-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2021Material Separation; Material Separation
THE PROBLEM OCCURRED WITH THE SECOND CORKSCREW THEY USED DURING THE PROCEDURE, THE FIRST ONE WENT FINE. THE USER INSERTED IT A HALF TURN, BUT IT COULD NOT BE INSERTED COMPLETELY INTO THE SKIN AS THE HUB STARTED TURNING. SHE HAD NO GRIP ON THE NEEDLE ANYMORE. THEN, SHE TRIED TO PULL IT OUT, BUT THE NEEDLE DID NOT COME OUT DUE TO LACK OF GRIP BECAUSE OF THE LOOSE HUB. THE SURGEON WAS IN A HURRY, WHICH IS WHY THEY PLACED A NEW CORKSCREW AND LEFT THE DEFECTIVE ONE AS IS. AFTER THE SURGERY, THE SURGEON USED A LITTLE SURGICAL INSTRUMENT (FORCEPS) TO REMOVE THE NEEDLE. 1MM AT A TIME, HE PULLED IT OUT. THE PATIENT BLED QUITE A LOT AFTER THE NEEDLE CAME OUT. THE NEEDLE WAS STILL ATTACHED TO THE LEAD WIRE. DUE TO USING PRESSURE, THE NEEDLE WAS DEFORMED. THE WOUND WAS STAUNCHED AND BLEEDING STOPPED AFTER A WHILE. NO STITCHING WAS NECESSARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12974279 ↗InjuryNov 15, 2021Material Separation; Material Separation
THE CORKSCREW ELECTRODE WAS INSERTED INTO A PATIENT'S SCALP, WAS REMOVED, AND SUBSEQUENTLY REPOSITIONED IN ANOTHER POSITION IN THE PATIENT'S SCALP. UPON SCREWING THE ELECTRODE BACK IN, THE HEALTH PROFESSIONAL NOTICED THAT IT WAS SCREWING BUT NOT STOPPING (INDICATES IT IS IN), TOOK HER HAND AWAY TO CHECK IF IT WAS IN, AND THE PLASTIC CAP AND CORD CAME AWAY WITH NO NEEDLE INSIDE. THE HEALTH PROFESSIONAL OBSERVED THAT THE NEEDLE ELECTRODE WAS STILL LODGED IN THE PATIENT'S SCALP. THEREAFTER, ATTEMPTS WERE MADE TO REMOVE THE NEEDLE ELECTRODE WITH FORCEPS, BUT EVENTUALLY A SMALL INCISION HAD TO BE MADE IN ORDER TO BE ABLE TO REMOVE THE NEEDLE ELECTRODE FROM THE SCALP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12811651 ↗
Other Electrode, Needle, Diagnostic Electromyograph devices
- Technomed Europe TE/S46-638Technomed Engineering B.V. · Electrode, Needle, Diagnostic Electromyograph1 recall2 reports
- Technomed Europe TE/S50715-001-CTechnomed Engineering B.V. · Electrode, Needle, Diagnostic Electromyograph1 recall0 reports
- Technomed Europe TE/S50715-004Technomed Engineering B.V. · Electrode, Needle, Diagnostic Electromyograph1 recall0 reports
- N/A 8227410Medtronic Xomed, Inc. · Electrode, needle, diagnostic electromyograph12 reports
- Teca Elite S53153Natus Manufacturing Limited · Electrode, Needle, Diagnostic Electromyograph7 reports
- Dantec 9013S0032Natus Manufacturing Limited · Electrode, Needle, Diagnostic Electromyograph6 reports
- Dantec 9013S0042Natus Manufacturing Limited · Electrode, Needle, Diagnostic Electromyograph4 reports
- Teca Elite 902-DMF25-SNatus Manufacturing Limited · Electrode, Needle, Diagnostic Electromyograph3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page