Skip to content

Kreatech FISH probes 40V016V495Recalls & reported problems

Leica Biosystems Newcastle Ltd · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 1, 2024Class IIOpen, Classified

    The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.

    Z-0522-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Reagents, specific, analyte devices

All Reagents, specific, analyte devices →