Skip to content

CastleLoc-P Anterior Cervical Plate System 6712-1741Recalls & reported problems

L&K Biomed CO. ,Ltd. · CastleLoc-P Anterior Cervical Plate (2Level-Contour Type) 41mm 1EA · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08806189893509), received between Jun 26, 2017 and Jun 26, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Disintegration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Disintegration1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2017Material Disintegration; Material Disintegration

    ON (B)(6) 2017, MED3 ASSOCIATES INC. (US DISTRIBUTOR OF AEGIS SPINE) REPORTED THE COMPLAINT TO AEGIS SPINE (US AGENCY OF L&K BIOMED), AND AEGIS SPINE REPORTED IT TO L&K BIOMED ON (B)(6) 2017. ACCORDING TO MED3 ASSOCIATES INC., DR. (B)(6) PERFORMED ACDF SURGERY ON (B)(6), 2016 USING CASTLELOC-P MODEL# IS 6712-1741 (2 LEVEL 41MM) AND LOT# IS UNKNOWN, AND UPON INVESTIGATION IT WAS DISCOVERED THAT THE CAM AT ONE OF THE C7 SCREWS HAD BROKEN OFF AT SOME POINT SINCE THE SURGERY IN (B)(6). THE PATIENT WAS ABOUT 7 WEEKS POST-OP AND DENIED FALLING OR ANY MOTOR VEHICLE ACCIDENT BUT HAS BEEN NON-COMPLIANT WITH HER NECK BRACE. DR.(B)(6) HAD RETRIEVED THE FREE-FLOATING CAM FROM THE PATIENT AND TOOK OUT THAT 16MM SCREW. HE DID NOT PUT IN A RESCUE SCREW BECAUSE THERE WAS NO LOCKING MECHANISM. MED3 ASSOCIATES INC. SAYS THAT THE PATIENT DOES NOT HAVE GREAT LORDOSIS AND IS RATHER STRAIGHT AND ALMOST KYPHOTIC ACCORDING TO PATIENT X-RAY PICTURE PROVIDED BY HOSPITAL, AND IT WAS SAID IT IS DIFFICULT TO ASSESS BUT I BELIEVE THERE MAY ALSO HAVE BEEN A NARROW DISK HEIGHT OF C7 - T1 ALONG WITH NOT WEARING HER NECK BRACE WHICH MAY HAVE CONTRIBUTED TO THE FORCES THAT WERE GENERATED TO CAUSE THE SCREW TO BACK O

    Patient problems: Failure of Implant

    MAUDE report 6668960 ↗

Other Appliance, Fixation, Spinal Intervertebral Body devices

All Appliance, Fixation, Spinal Intervertebral Body devices →

Sources and updates