Skip to content

Kozee M21505Recalls & reported problems

Sourcemark, LLC. · Warming Unit · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08973920006730), received between Aug 27, 2025 and Aug 27, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “No Audible Prompt/Feedback” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Audible Prompt/Feedback1
  • Output Problem1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2025No Audible Prompt/Feedback; Output Problem; No Audible Prompt/Feedback; Output Problem

    A PATIENT WENT UNDER A 6HR HEART SURGERY. THE HOSPITAL USED 505 BLOWER WITHOUT BLANKET. THE PATIENT EXPERIENCED A SECOND DEGREE BURN ON BOTH LEGS. THE HOSPITAL IS CLAIMING THAT OUR EQUIPMENT DID NOT HAVE ALARM OR AUTOMATICALLY STOP WHEN OVERHEATING. THE HOSPITAL'S BIOMED GROUP DID A TEST SHOWING THAT OUR BLOWER CONTINUED BLOWING EVEN WHEN THE TEMPERATURE AT THE AIR HOSE HITTING 63 DEGREES.

    Patient problems: Burn(s)

    MAUDE report 22902938 ↗

Other System, Thermal Regulating devices

All System, Thermal Regulating devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page