Skip to content

SONATATI100 +FLEX24 04918Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This MED-EL cochlear implant is placed in the inner ear to aid hearing. The 26 reports mostly describe implants losing sound output or hearing performance and later being removed and replaced, sometimes after a head injury or with electrodes out of place and facial stimulation. The reports mention no deaths and 8 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 09008737049188), received between Aug 8, 2019 and Sep 25, 2026 — classified as 9 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Output Problem” (14 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem14
  • Impedance Problem10
  • Migration5
  • Adverse Event Without Identified Device or Use Problem1
  • Circuit Failure1
  • Device Appears to Trigger Rejection1

Patient problems

  • Failure of Implant20
  • Head Injury3
  • No Clinical Signs, Symptoms or Conditions2
  • Extraskeletal ossification1
  • Headache1
  • Implant Pain1
  • Undesired Nerve Stimulation1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Impedance Problem

    In situ measurements from (b)(6) 2023, there was an identified abnormal impedance across 5 channels following a bump. However, no marked deterioration in hearing performance was observed at that time. Three years later, the number of channels with abnormal impedance rose to seven.The user has been re-implanted.

    Patient problems: Head Injury

    MAUDE report 50894912 ↗
  • MalfunctionMay 29, 2026Impedance Problem

    THE USER HAS NOT USED THE DEVICE FOR 6 YEARS AFTER LOSING THE AP. HE WAS UNABLE TO AFFORD A REPLACEMENT. VIA PUBLIC FUNDING HE WOULD BE ABLE TO GET A NEW PROCESSOR. NEXT STEPS INCLUDE PERFORMING INTEGRITY TEST _ AWAITING INTERNAL DEVICE REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25328688 ↗
  • MalfunctionMay 7, 2026Impedance Problem

    THE DEVICE WAS EXPLANTED.

    Patient problems: Head Injury

    MAUDE report 25112056 ↗
  • MalfunctionJun 24, 2025Impedance Problem; Impedance Problem

    THE RECIPIENT HAS NOT BEEN USING THE AUDIO PROCESSOR FOR 2 YEARS DUE TO MAINTENANCE DIFFICULTIES. ON THE DAY OF UPGRADE ACTIVATION, AFFECTED CHANNELS WERE SEEN DURING TESTING. FURTHER PROCEEDINGS ARE UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22313553 ↗
  • MalfunctionSep 19, 2024Migration; Migration

    ON (B)(6) 2024 THE RECIPIENT REPORTED EXPERIENCING A HEADACHE. THE FOLLOWING DAY THE RECIPIENT WENT TO THE HOSPITAL FOR EXAMINATION. THE RECIPIENT HAS BEEN RE-IMPLANTED.

    Patient problems: Headache

    MAUDE report 20253220 ↗
  • MalfunctionJul 11, 2024Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    THE RECIPIENT WAS IMPLANTED AS SINGLE SIDED DEAFNESS (SSD) SINCE A COUPLE OF YEARS DUE TO OSSIFYING LABYRINTHITIS. DUE TO SEVERE OSSIFICATION, ONLY 3 CHANNELS WERE FUNCTIONAL. REPORTEDLY, SOME CHANNELS WERE EXTRA-COCHLEAR AND OTHERS CAUSE SEVERE FACIAL NERVE STIMULATION. THE RECIPIENT HAD A REVISION SURGERY ON (B)(6) 2021 DUE TO THE EXTRA-COCHLEAR CHANNELS, HOWEVER, WITHOUT ANY IMPROVEMENT. THE RECIPIENT VISITED ANOTHER CLINIC FOR EXPLANTATION.

    Patient problems: Extraskeletal ossification; Undesired Nerve Stimulation

    MAUDE report 19717482 ↗
  • MalfunctionMay 19, 2023Impedance Problem; Impedance Problem

    SINCE (B)(6) 2023, THE USER WAS COMPLAINING ABOUT DECREASED HEARING PERFORMANCE WITH THE DEVICE. RE-IMPLANTATION IS CONSIDERED BUT HAS NOT BEEN SCHEDULED YET.

    Patient problems: Failure of Implant

    MAUDE report 16960282 ↗
  • MalfunctionMay 8, 2023Impedance Problem; Impedance Problem

    DURING A FITTING APPOINTMENT ON 06-AUG-2021 IT WAS OBSERVED THAT SOME CHANNELS HAD FLUCTUATING IMPEDANCES BUT THE USER DID NOT REPORT ANY PROBLEM WITH THE HEARING PERFORMANCE WHEN USING THE DEVICE. THE USER_S PARENT REPORTED THAT ON 13-AUG-2022, THE USER COULD NOT HEAR NORMALLY WITH THE DEVICE. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16885198 ↗
  • MalfunctionApr 13, 2023Impedance Problem; Impedance Problem

    THE PARENTS REPORTED THAT THE RECIPIENT HAD A BLOW TO HER HEAD ON 01-MAR-2023. AFTER THE ACCIDENT, THE RECIPIENT_S HEARING PERFORMANCE WITH THE DEVICE WAS NOT GOOD.THE RECIPIENT HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16730178 ↗
  • MalfunctionFeb 15, 2023Circuit Failure; Circuit Failure

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED. REPORTEDLY, THE SURGEON HAS RECOMMENDED TO RE-IMPLANT THE RECIPIENT DUE TO THE COMPLETE LOSS OF CONNECTION BETWEEN THE IMPLANT AND EXTERNAL DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 16377183 ↗
  • MalfunctionJan 4, 2023Impedance Problem; Impedance Problem

    THE PARENTS REPORTED THAT THE USER HAD A HEAD TRAUMA WITH BLEEDING IN THE (B)(6) 2022, AFTERWARDS HE SHOWED POOR AUDITORY RESPONSES WITH THE DEVICE. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16094415 ↗
  • MalfunctionDec 15, 2022Impedance Problem; Output Problem; Impedance Problem; Output Problem

    THE USER HEARING PERFORMANCE WITH THE DEVICE WAS LIMITED. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15985754 ↗
  • MalfunctionAug 30, 2022Output Problem; Output Problem

    THE USER HAS STOPPED HEARING WITH THE DEVICE AFTER A HEAD TRAUMA OCCURRED ON (B)(6) 2022 AND CHANNELS WITH ABNORMAL IMPEDANCES WERE FOUND. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15318501 ↗
  • MalfunctionMar 18, 2022Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AND CHANNELS WITH ABNORMAL IMPEDANCES WERE FOUND.

    Patient problems: Failure of Implant

    MAUDE report 13804077 ↗
  • MalfunctionFeb 21, 2022Output Problem; Output Problem

    THE USER BUMPED HIS HEAD ON (B)(6) 2022 AND THEN COULD NO LONGER HEAR WITH THE DEVICE. THE USER WAS RE-IMPLANTED ON (B)(6) 2022.

    Patient problems: Failure of Implant

    MAUDE report 13567295 ↗
  • MalfunctionFeb 21, 2022Output Problem; Output Problem

    THE USER ATTENDED FOR A FIRST REVIEW MAPPING APPOINTMENT IN (B)(6), AFTER RELOCATING FROM (B)(6) AND CHANNELS WITH ABNORMAL IMPEDANCES WERE FOUND.

    Patient problems: Failure of Implant

    MAUDE report 13567253 ↗
  • MalfunctionJan 3, 2022Output Problem; Output Problem

    THE USER HAD ABNORMAL IMPEDANCE LEVELS AND POOR HEARING PERFORMANCE WITH THE DEVICE DURING A APPOINTMENT IN (B)(6) 2021. THE USER WAS RE-IMPLANTED ON (B)(6) 2021.

    Patient problems: Failure of Implant

    MAUDE report 13142032 ↗
  • InjuryAug 19, 2021Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE HAS BEEN DECREASING OVER THE PAST MONTHS AND SHE HAS NO ACCESS TO SOUND NOW. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12341047 ↗
  • MalfunctionAug 2, 2021Impedance Problem; Output Problem; Impedance Problem; Output Problem

    THE USER WAS SEEN ON (B)(6) 2021 FOR HIS ANNUAL FITTING. IN SITU TESTING DONE DURING THIS APPOINTMENT SHOWED ALL CHANNELS AFFECTED WITH HIGH IMPEDANCE OR SHORT CIRCUITS. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12257703 ↗
  • InjuryJun 14, 2021Output Problem; Output Problem

    THE USER'S GUARDIANS NOTED A SIGNIFICANT LOSS OF HEARING WITH THE DEVICE AFTER A TRAUMA TO THE HEAD. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11991454 ↗
  • InjuryFeb 10, 2021Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED. IN SITU MEASUREMENTS SHOWED AFFECTED CHANNELS. THE USER WAS RE-IMPLANTED ON (B)(6) 2021.

    Patient problems: Failure of Implant

    MAUDE report 11307670 ↗
  • InjuryDec 29, 2020Output Problem; Output Problem

    WHEN THE USER CAME FOR A REGULAR CHECK-UP, IN SITU TESTING SHOWED AFFECTED CHANNELS. HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE PARENTS HAD PREVIOUSLY NOT NOTICED ANY CHANGE OF HEARING OR BEHAVIOR AT HOME.

    Patient problems: Failure of Implant

    MAUDE report 11085585 ↗
  • InjuryOct 27, 2020Migration; Migration

    SEVERAL MONTHS AFTER IMPLANTATION A MIGRATION OF THE IMPLANT HOUSING DOWN TO THE LEVEL OF THE AUDIO PROCESSOR WAS NOTICED. BECAUSE OF PAIN, THE IMPLANT HOUSING WAS REPOSITIONED ON (B)(6) 2020. AFTER REVISION SURGERY AFFECTED CHANNELS AND EXTRA-COCHLEA CHANNELS WERE OBSERVED. THE RECIPIENT WAS RE-IMPLANTED ON (B)(6) 2020.

    Patient problems: Implant Pain

    MAUDE report 10740469 ↗
  • InjuryAug 26, 2020Device Appears to Trigger Rejection; Migration; Device Appears to Trigger Rejection; Migration

    WHILE THE USER WAS UNDERGOING AN EAR CLEANING PROCEDURE THE ELECTRODE LEAD WAS FOUND TO BE IN THE EXTERNAL AUDITORY CANAL. REPAIR OF THE EXTERNAL AUDITORY CANAL AND RE-IMPLANTATION HAD BEEN SUGGESTED. THE USER WAS RE-IMPLANTED ON (B)(6) 2020.

    Patient problems: Failure of Implant

    MAUDE report 10452678 ↗
  • InjuryJun 3, 2020Output Problem; Output Problem

    THE USER'S FATHER REPORTED THAT THE USER FELL ON THE SIDE OF THE IMPLANT AROUND THE MIDDLE OF (B)(6) 2020. THE USER DEVELOPED A HAEMATOMA ON THE EAR. SINCE THEN SHE CAN NO LONGER HEAR WITH THE DEVICE. THE USER'S HEARING PERFORMANCE WITH THE DEVICE BEFORE THE TRAUMA WAS GOOD.

    Patient problems: Failure of Implant; Head Injury

    MAUDE report 10113583 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCE

Other Implant, Cochlear devices

All Implant, Cochlear devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page