Mi1000 MED-EL CONCERT +MEDIUM 06272Recalls & reported problems
Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This MED-EL CONCERT+ Medium cochlear implant is an implanted hearing device. The 47 reports mostly describe the implant not working as expected: rising electrical resistance in some channels, reduced or lost hearing benefit, and a few cases where the implant had to be removed or replaced. Reports mention 13 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 09008737062729), received between Nov 27, 2015 and Aug 24, 2026 — classified as 13 injuries, 34 malfunctions.
The most frequently reported specific device problem is “Impedance Problem” (22 reports).
4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Impedance Problem22
- Output Problem19
- Adverse Event Without Identified Device or Use Problem4
- Insufficient Device Problem Information4
- Migration3
- Device Appears to Trigger Rejection2
- Lack of Effect2
- Device Operates Differently Than Expected1
Patient problems
- Failure of Implant36
- No Clinical Signs, Symptoms or Conditions7
- Head Injury4
- Skin Tears2
- Device Overstimulation of Tissue1
- Extraskeletal ossification1
- Implant Pain1
- Pain1
Recent adverse event reports
The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.
InjuryAug 24, 2026Impedance Problem
The user's parents reported sudden hearing loss with the left CI following head impact on (b)(6) 2026. After one month of observation with no improvement in the hearing performance, reimplantation was performed on (b)(6) 2026.
Patient problems: Head Injury
MAUDE report 50403182 ↗MalfunctionApr 9, 2026Impedance Problem
HOWEVER, DURING THE FIRST DEVICE ADJUSTMENT ON (B)(6) 2025, MEASUREMENTS SHOWED ABNORMAL IMPEDANCE IN SOME CHANNELS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24830011 ↗MalfunctionFeb 26, 2026Impedance Problem
MalfunctionJan 29, 2026Impedance Problem
THE HEARING PERFORMANCE OF THE DEVICE IS AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24207342 ↗MalfunctionJul 18, 2025Impedance Problem; Impedance Problem
IN SITU MEASUREMENTS SHOWED AFFECTED CHANNELS. THE USER NEVER SHOWED ANY DISCOMFORT. RE-IMPLANTATION IS BEING CONSIDERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22539519 ↗MalfunctionJul 18, 2025Impedance Problem; Impedance Problem
THE USER'S PARENT REPORTED IN EARLY (B)(6) 2025 THAT A RECENT DECLINE IN HEARING PERFORMANCE WITH THE RIGHT COCHLEAR IMPLANT WAS OBSERVED. THE USER HAS BEEN RE-IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22539513 ↗MalfunctionDec 30, 2024Electrical Shorting; Lack of Effect; Electrical Shorting; Lack of Effect
THE DEVICE WAS EXPLANTED. THE USER WAS DISSATISFIED WITH ITS PERFORMANCE.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 21034141 ↗MalfunctionDec 30, 2024Migration; Lack of Effect; Migration; Lack of Effect
THE DEVICE WAS EXPLANTED. THE USER WAS DISSATISFIED WITH ITS PERFORMANCE.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 21034026 ↗MalfunctionSep 13, 2024Impedance Problem; Impedance Problem
THE PARENT REPORTED THAT THE USER WAS SUDDENLY UNABLE TO HEAR SOUNDS WITH THE DEVICE. THE USER HAS BEEN RE-IMPLANTED.
Patient problems: Failure of Implant
MAUDE report 20209980 ↗MalfunctionSep 3, 2024Impedance Problem; Impedance Problem
THE USER REPORTED THAT HE COULD NOT HEAR WITH HIS DEVICE ANYMORE. EXPERT MEASUREMENTS WILL BE PLANNED TO CONFIRM A POTENTIAL DEVICE FAILURE.
Patient problems: Failure of Implant
MAUDE report 20123397 ↗MalfunctionMar 29, 2024Impedance Problem; Impedance Problem
THE PARENTS OF THE USER REPORTED A TRAUMA IN THE REGION OF THE IMPLANT. EXPERT TELEMETRY WILL BE SCHEDULED BY THE DISTRIBUTOR TO CONFIRM THE FAILURE. RE-IMPLANTATION IS CONSIDERED. NO DATE HAS BEEN SCHEDULED YET.
Patient problems: Failure of Implant
MAUDE report 19002341 ↗MalfunctionFeb 5, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
THE EXTRUSION OF THE DEVICE WAS REPORTED WHICH STARTED IN MID-JANUARY, RESULTING IN AN EXPLANTATION AND CLOSING OF THE WOUND. THE USER HAS BEEN EXPLANTED.
Patient problems: Skin Tears
MAUDE report 18640522 ↗MalfunctionOct 10, 2023Impedance Problem; Impedance Problem
THE PARENTS REPORTED THAT THE USER WAS NO LONGER ABLE TO HEAR WITH THE DEVICE. RE-IMPLANTATION HAS BEEN SUGGESTED TO THE MEDICAL TEAM, BUT NO DATE HAS BEEN SCHEDULED YET.
Patient problems: Failure of Implant
MAUDE report 17904326 ↗MalfunctionAug 28, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE USER HAD PAIN AROUND THE COIL STARTING TWO YEARS AGO. THE USER HARDLY USED THE PROCESSOR DUE TO THE PAIN. THE USER HAS BEEN RE-IMPLANTED.
Patient problems: Extraskeletal ossification; Implant Pain
MAUDE report 17630522 ↗MalfunctionJul 20, 2023Impedance Problem; Impedance Problem
THE USER_S HEARING PERFORMANCE SUDDENLY DECREASED.
Patient problems: Failure of Implant
MAUDE report 17361065 ↗MalfunctionJul 19, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. RE-IMPLANTATION IS PLANNED.
Patient problems: Failure of Implant
MAUDE report 17348380 ↗MalfunctionMay 15, 2023Malposition of Device; Insufficient Device Problem Information; Malposition of Device; Insufficient Device Problem Information
THE USER REPORTED SHE HAD NEVER PERFORMED WELL WITH HER DEVICE. THE USER HAS BEEN RE-IMPLANTED.
Patient problems: Failure of Implant
MAUDE report 16929472 ↗MalfunctionMay 12, 2023Impedance Problem; Impedance Problem
IN (B)(6) 2022, THE USER HAD A FALL AND HIT HIS HEAD ON THE IMPLANT SIDE. FURTHER MEDICAL INTERVENTION IS CONSIDERED, BUT NO DATE HAS BEEN PROVIDED YET.
Patient problems: Failure of Implant
MAUDE report 16917415 ↗MalfunctionApr 25, 2023Impedance Problem; Impedance Problem
THE USER FELL AND HIT HER HEAD ON THE IMPLANT SITE. FOLLOWING THIS EVENT, THE PARENTS IMMEDIATELY NOTICED THAT THE CHILD WAS NOT HEARING ANYMORE WITH THE DEVICE. A RE-IMPLANTATION WILL BE PLANNED BUT NO DATE HAS BEEN SCHEDULED YET.
Patient problems: Failure of Implant
MAUDE report 16805792 ↗MalfunctionMar 28, 2023Impedance Problem; Impedance Problem
ON (B)(6) 2023 THE USER ATTENDED A FITTING SESSION AND DURING THE IN SITU MEASUREMENTS, AFFECTED CHANNELS HAVE BEEN SEEN. FURTHERMORE, THE HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED. THE USER HAS BEEN RE-IMPLANTED.
Patient problems: Failure of Implant
MAUDE report 16626279 ↗MalfunctionMar 23, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE IS AFFECTED.
Patient problems: Failure of Implant
MAUDE report 16596639 ↗MalfunctionJan 30, 2023Impedance Problem; Impedance Problem
THE USER'S FATHER REPORTED THAT AFTER A FALL, THE USER SUDDENLY STOPPED HEARING AND RESPONDING TO SPEECH ON THE AFFECTED SIDE. IT WAS SUGGESTED TO DO NOT USE THE PROCESSOR ON THAT SIDE.
Patient problems: Failure of Implant
MAUDE report 16263722 ↗MalfunctionJan 27, 2023Impedance Problem; Insufficient Device Problem Information; Impedance Problem; Insufficient Device Problem Information
THE USER'S HEARING PERFORMANCE WITH THE DEVICE HAS DETERIORATED AFTER A HEAD TRAUMA WHICH OCCURRED IN (B)(6) 2020. THE USER_S PERFORMANCE WITH THE DEVICE WAS GOOD BEFORE THE EVENT OCCURRED. RE-IMPLANTATION HAS BEEN DONE ON (B)(6)2023.
Patient problems: Failure of Implant
MAUDE report 16252138 ↗MalfunctionJan 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE RECIPIENT REQUESTED THE EXPLANTATION OF THE DEVICE DUE TO PAIN MORE THAN 7 YEARS AFTER IMPLANTATION. THE RECIPIENT HAS BEEN EXPLANTED ON (B)(6) 2022.
Patient problems: Pain
MAUDE report 16245330 ↗MalfunctionSep 6, 2022Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration
THE USER PRESENTED SUDDEN DEAFNESS IN THE LEFT EAR IN 2020.
Patient problems: Device Overstimulation of Tissue; Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 15362130 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.
- Z-2356-2024 · Apr 18, 2024 · Open, ClassifiedMi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
- Z-2137-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
- Z-2138-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
- Z-2136-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
Other Implant, Cochlear devices
- HiRes™ Ultra CI HiFocus™ MS Electrode CI-1600-04Advanced Bionics AG · Implant, Cochlear2,918 reports
- Cochlear® Nucleus® CI512 cochlear ImplantCochlear Limited · IMPLANT, COCHLEAR2,771 reports
- Cochlear® Nucleus® CI622 cochlear Implant with Slim Straight electrodeCochlear Limited · Implant, cochlear2,115 reports
- Cochlear® Nucleus® CI632 cochlear Implant with Slim Modiolar electrodeCochlear Limited · Implant, cochlear2,057 reports
- Nucleus Z60353Cochlear Limited · IMPLANT, COCHLEAR2 deaths1,815 reports
- Cochlear® Nucleus® CI522 cochlear Implant with Slim Straight electrodeCochlear Limited · IMPLANT, COCHLEAR1,475 reports
- Cochlear® Nucleus® CI612 cochlear ImplantCochlear Limited · Implant, cochlear1,150 reports
- Nucleus Z190574Cochlear Limited · IMPLANT, COCHLEAR981 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page