Mi1000 MED-EL CONCERT PIN +FLEX24 06273Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 09008737062736), received between Dec 5, 2018 and Nov 19, 2024 — classified as 6 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Output Problem” (7 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem7
  • Migration3
  • Adverse Event Without Identified Device or Use Problem2
  • Impedance Problem2
  • Circuit Failure1
  • Insufficient Device Problem Information1

Patient problems

  • Failure of Implant12
  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body Reaction1
  • Inflammation1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1
  • Tinnitus1
  • Undesired Nerve Stimulation1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2024Impedance Problem; Impedance Problem

    THE USER HAD A SUDDEN LOSS OF HEARING SENSATION. RE-IMPLANTATION IS BEING CONSIDERED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20721245 ↗
  • MalfunctionAug 21, 2024Impedance Problem; Impedance Problem

    THE USER HAS FLUCTUATING IMPEDANCES. THE PERFORMANCE WITH THE DEVICE IS AFFECTED. NEXT STEPS WILL BE DISCUSSED.

    Patient problems: Failure of Implant

    MAUDE report 20012871 ↗
  • MalfunctionDec 4, 2023Migration; Migration

    BASED ON IMAGING FOUR CHANNELS ARE EXTRA COCHLEAR ON THE CONCERNED SIDE AND HAVE BEEN DISABLED. THE RECIPIENT HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18254491 ↗
  • MalfunctionSep 27, 2022Circuit Failure; Output Problem; Circuit Failure; Output Problem

    THE USER DOES NOT HAVE HEARING SENSATION. A RE-IMPLANTATION IS CONSIDERED, BUT NO DATE HAS BEEN SCHEDULED YET.

    Patient problems: Failure of Implant

    MAUDE report 15494466 ↗
  • MalfunctionSep 12, 2022Migration; Migration

    THE USER UNDERWENT A REVISION SURGERY TO REPOSITION THE DEVICE, DURING WHICH THE DEVELOPMENT OF A CHOLESTEATOMA WAS VERIFIED. IT WAS THEN DECIDED TO EXPLANT THE DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 15402125 ↗
  • MalfunctionFeb 21, 2022Output Problem; Output Problem

    HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED.

    Patient problems: Failure of Implant

    MAUDE report 13568738 ↗
  • InjuryMar 25, 2021Output Problem; Output Problem

    IN SITU MEASUREMENTS SHOW ABNORMAL RESULTS.

    Patient problems: Failure of Implant

    MAUDE report 11567700 ↗
  • InjuryMar 22, 2021Output Problem; Output Problem

    THE RECIPIENT HAD EXPERIENCED A DECREASING OF THE HEARING IMPRESSION FOR APPROXIMATELY HALF A YEAR. THE RECIPIENT ALSO REPORTED UNPLEASANT PRESSURE SENSATION OVER THE IMPLANT AREA WHEN WEARING THE AUDIO PROCESSOR. REPORTEDLY, THE MAGNET STRENGTH SEEMED TO BE APPROPRIATE WITHOUT BEING TOO STRONG AND THERE WAS NO VISIBLE SWELLING OR SKIN REACTION. THE RECIPIENT HAS BEEN RE-IMPLANTED ON (B)(6) 2021.

    Patient problems: Failure of Implant

    MAUDE report 11540682 ↗
  • InjuryFeb 3, 2021Migration; Migration

    IT HAS BEEN REPORTED A MIGRATION OF THE ELECTRODE ARRAY AND ELECTRODES FROM SEVEN TO TWELVE ARE OUT OF COCHLEA. THE USER HAS BEEN EXPLANTED ON (B)(6) 2021.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 11271478 ↗
  • InjuryFeb 7, 2020Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. IT IS REPORTED THAT THE PATIENT DID NOT PRESENT ANY RESPONSES IN THE FREE FIELD AUDIOMETRY (NOT ABLE TO DETECT LOUD SOUNDS).

    Patient problems: Failure of Implant

    MAUDE report 9680465 ↗
  • InjuryNov 5, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    SINCE IMPLANTATION, THE RECIPIENT HADN'T SHOWED ANY CLEAR ADVANCEMENT WITH THE DEVICE. THERE WAS NO CLEAR RESPONSE TO THE DEVICE ACCORDING TO REHABILITATION INFORMATION. ON (B)(6)2019 A SMALL BUMP WAS SEEN OVER HER SCAR. THE RECIPIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2019. REPORTEDLY, THE RECIPIENT WAS RE-ACTIVATED WITH GOOD RESULTS.

    Patient problems: Foreign Body Reaction; Inflammation

    MAUDE report 9279043 ↗
  • MalfunctionApr 7, 2019Adverse Event Without Identified Device or Use Problem; Output Problem; Adverse Event Without Identified Device or Use Problem; Output Problem

    USER REPORTED A DECREASED IN HEARING PERFORMANCE WITH THE DEVICE FOLLOWING A CAR ACCIDENT ON THE IN (B)(6) 2019. THE USER WAS HIT HARD FROM BEHIND RESULTING IN SWELLING OVER THE FACE. SHE WAS ALSO TREATED FOR A MIDDLE EAR INFECTION. AS PER NEW INFORMATION, USER'S HEARING PERFORMANCE IS CURRENTLY GOOD FOLLOWING THE RESOLUTION OF THE SWELLING.

    Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient

    MAUDE report 8489963 ↗
  • InjuryDec 5, 2018Output Problem; Output Problem

    THE USER EXPERIENCED A SUDDEN LOSS OF SOUND WITH THE DEVICE. IN THE FOLLOWING FEW DAYS THE DEVICE WOULD WORK INTERMITTENTLY. ON THE THIRD DAY, AFTER 3-4 HOURS OF USE, HE HAD AN UNCOMFORTABLE PINCHING FEELING IN HIS EAR AND FACIAL NERVE STIMULATION. THIS WAS FOLLOWED BY A TINNITUS THAT CHANGED FROM HIGH PITCHED TO LOW PITCHED. NO TRAUMA OR ACCIDENT IS REPORTED.

    Patient problems: Failure of Implant; Tinnitus; Undesired Nerve Stimulation

    MAUDE report 8132359 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge

Other Implant, Cochlear devices

All Implant, Cochlear devices →