Skip to content

Mi1200 SYNCHRONY PIN +COMPRESSED 09397Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 09008737093976), received between Jul 2, 2019 and Apr 3, 2024 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Output Problem” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem2
  • Circuit Failure1
  • Device Handling Problem1
  • Impedance Problem1
  • Insufficient Device Problem Information1
  • Malposition of Device1
  • No Device Output1

Patient problems

  • Failure of Implant6
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2024Circuit Failure; No Device Output; Circuit Failure; No Device Output

    THE USER REPORTS THAT SINCE 2 WEEKS HE HEARS 3 "BEEPS" WHEN HE PUTS ON THE EXTERNAL DEVICE. EVERYTHING ELSE IS WORKING NORMALLY AND HIS HEARING IS NORMAL. IN SITU MEASUREMENTS COULD NOT BE PERFORMED DUE TO INTEGRITY FAILURE. THERE HAVE BEEN NO RECENT CHANGES IN HEALTH OR MEDICATION. ACCORDING TO NEW INFORMATION FROM 26.MAR.2024 THERE WAS SOME LEAKAGE OF THE DEVICE.

    Patient problems: Failure of Implant; Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19032533 ↗
  • MalfunctionMar 25, 2024Malposition of Device; Malposition of Device

    DURING THE INITIAL SURGICAL PROCEDURE, THE ACTIVE ELECTRODE ARRAY COULDN'T BE FULLY INSERTED INTO THE USER'S COCHLEA DUE TO ANATOMICAL ISSUES. THE CLINIC WAS ONLY ABLE TO INSERT 4 CHANNELS WHICH WAS CONFIRMED BY IMAGING. THE CLINIC NOTICED MULTIPLE CHOLESTEATOMAS AND OTHER EAR DISEASES. THE USER DID NOT FOLLOW THE RECOMMENDED CLINICAL AFTERCARE, RECENTLY PRESENTED AT THE CLINIC FOR EXPLANTATION, HAVING RELIED ON HEARING FROM THE CONTRALATERAL SIDE AND HAVING NEVER EXPERIENCED ANY HEARING PERCEPTION POST-SURGERY WITH THE DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 18966825 ↗
  • MalfunctionJan 26, 2024Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED AFTER A TRAUMA TO THE HEAD. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 18585814 ↗
  • InjuryApr 20, 2020Output Problem; Output Problem

    THE USER SUSTAINED A TRAUMA ON THE REGION OF THE IMPLANT. FOLLOWING THE INCIDENT THE USER COULD NO LONGER HEAR WITH THE DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 9977062 ↗
  • InjuryOct 24, 2019Output Problem; Output Problem

    IN (B)(6) 2019, THE RECIPIENT UNDERWENT SURGERY TO REPOSITION THE STIMULATOR HOUSING. DURING THIS PROCEDURE THE ACTIVE ELECTRODE WAS REPORTEDLY NOT MOVED. THIS PROCEDURE WAS CONDUCTED AS THE USER WAS UNABLE TO WEAR THE AUDIO PROCESSOR FOR LONG, DUE TO FEELING PAIN AT THE MAGNET SITE. THE SKIN BECAME THINNER AND AN ULCER FORMED. THE EXTERNAL MAGNET STRENGTH WAS THOUGHT TO BE APPROPRIATE FOR THE USER. SINCE THE REPOSITIONING SURGERY THE RECIPIENT HAS NO AUDITORY SENSATION WITH THE DEVICE. ADDITIONALLY, THE USER REPORTEDLY HAD NO AUDITORY SENSATION WITH THE DEVICE BEFORE REPOSITIONING SURGERY. THE USER HAS BEEN RE-IMPLANTED. ACCORDING TO THE DEVICE EXPLANTATION REPORT, THE ELECTRODE ARRAY WAS OUT OF THE COCHLEA.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 9232822 ↗
  • InjuryJul 2, 2019Insufficient Device Problem Information; Device Handling Problem; Insufficient Device Problem Information; Device Handling Problem

    AFTER INITIAL IMPLANTATION, IMAGING SHOWED ELECTRODES PASSING THROUGH THE ROUND WINDOW WITH KINKING AND FOLDING OF THE ELECTRODE ARRAY IN THE FIRST 3 MM OF THE BASAL TURN. ALSO, DURING THE INSERTION, THE ELECTRODE CONTACTS APPEARED TO BE DISCONNECTED FROM THE SILICON. PER THE IMPLANT REGISTRATION CARD, 2 ELECTRODES WERE LEFT EXTRA-COCHLEAR. THE DEVICE WAS NEVER ACTIVATED, AND THE RECIPIENT WAS RE-IMPLANTED 5 DAYS AFTER INITIAL SURGERY.

    Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient

    MAUDE report 8752060 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100

Other IMPLANT, COCHLEAR devices

All IMPLANT, COCHLEAR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page