Skip to content

Mi1200 SYNCHRONY +MEDIUM 09398Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 09008737093983), received between Nov 29, 2018 and Feb 26, 2026 — classified as 8 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Impedance Problem” (10 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Impedance Problem10
  • Output Problem5
  • Device Appears to Trigger Rejection2
  • Adverse Event Without Identified Device or Use Problem1
  • Circuit Failure1
  • Insufficient Device Problem Information1
  • Migration1

Patient problems

  • Failure of Implant13
  • No Clinical Signs, Symptoms or Conditions3
  • Head Injury2
  • Infectious Otitis Media1
  • Insufficient Information1
  • Skin Tears1
  • Tissue Breakdown1
  • Unspecified Infection1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Impedance Problem

    AFFECTED CHANNELS WERE NOTICED DURING IN SITU MEASUREMENTS. REIMPLANTATION SURGERY IS CONSIDERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452090 ↗
  • MalfunctionFeb 3, 2026Impedance Problem

    THE USER'S HEARING PERFORMANCE IS AFFECTED. NO INCIDENT OR TRAUMA REPORTED. REIMPLANTATION IS CONSIDERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246894 ↗
  • MalfunctionJan 26, 2026Circuit Failure

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED AFTER BEING HIT BY AN ELECTRIC BIKE. THE USER WAS REIMPLANTED.

    Patient problems: Head Injury

    MAUDE report 24176233 ↗
  • MalfunctionJun 10, 2025Impedance Problem; Impedance Problem

    THE USER HAS NO SOUND PERCEPTION WITH THE DEVICE SINCE APPROXIMATELY (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22176285 ↗
  • InjuryMay 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT HAD OTITIS MEDIA WHICH HAD SPREAD TO THE LEVEL OF THE IMPLANT AND INVADED THE ELECTRODES. THE USER WAS EXPLANTED ON (B)(6) 2025. AS THERE WAS SEVERE INFLAMMATION RE-IMPLANTATION WAS PUT ON HOLD UNTIL IT IS COMPLETELY HEALED.

    Patient problems: Infectious Otitis Media

    MAUDE report 22000564 ↗
  • MalfunctionJul 18, 2024Impedance Problem; Impedance Problem

    ACCORDING TO LIMITED AVAILABLE INFORMATION RE-IMPLANTATION IS BEING CONSIDERED. FURTHER INFORMATION HAS BEEN REQUESTED.

    Patient problems: Failure of Implant

    MAUDE report 19770337 ↗
  • MalfunctionMay 22, 2024Impedance Problem; Impedance Problem

    HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER TRAUMA. THE USER WAS RE-IMPLANTED ON (B)(6) 2024.

    Patient problems: Failure of Implant

    MAUDE report 19369257 ↗
  • MalfunctionMay 22, 2024Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. FURTHER INFORMATION WILL FOLLOW FROM THE CLINIC.

    Patient problems: Failure of Implant

    MAUDE report 19369122 ↗
  • MalfunctionApr 23, 2024Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A TRAUMA. THE USER NOTICED A SUDDEN CHANGE IN HEARING WITH THE DEVICE.

    Patient problems: Failure of Implant; Head Injury

    MAUDE report 19163859 ↗
  • MalfunctionJan 9, 2024Impedance Problem; Impedance Problem

    THE USER_S HEARING PERFORMANCE WITH THE DEVICE HAS BEEN DECLINING OVER THE YEARS, WITH A SIGNIFICANT DROP SINCE AUGUST 2023. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 18475596 ↗
  • MalfunctionDec 21, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED AFTER A TRAUMA TO THE HEAD ON (B)(6)2023. RE-IMPLANTATION WAS PERFORMED.

    Patient problems: Failure of Implant

    MAUDE report 18370800 ↗
  • MalfunctionJul 7, 2023Impedance Problem; Impedance Problem

    THE USER REPORTS LOW AUDIO VOLUME AND SHOWS NO RESPONSE TO STIMULATION DURING IN SITU TESTING. NO TRAUMA WAS REPORTED. "FURTHER" MEDICAL INTERVENTION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 17276602 ↗
  • InjurySep 10, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE USER UNDERWENT A REVISION SURGERY TO REPAIR THE SKIN FLAP ON THE (B)(6) 2021 DUE TO EXPOSURE OF THE DEVICE. THE IMPLANT HOUSING HAD PARTIALLY EXTRUDED THROUGH THE SKIN.

    Patient problems: Skin Tears; Unspecified Infection

    MAUDE report 12451978 ↗
  • InjuryAug 11, 2021Migration; Migration

    REPORTEDLY THE USER'S ELECTRODE ARRAY HAD MIGRATED COMPLETELY OUT OF THE COCHLEA AND WAS SEEN THROUGH OTOSCOPY. THE DEVICE WAS EXPLANTED ON (B)(6) 2021 FOR NOW THERE ARE NO PLANS TO RE-IMPLANT THE USER.

    Patient problems: Failure of Implant

    MAUDE report 12306645 ↗
  • InjuryJan 25, 2021Device Appears to Trigger Rejection; Insufficient Device Problem Information; Device Appears to Trigger Rejection; Insufficient Device Problem Information

    DEVICE WAS EXPLANTED ON (B)(6), 2021 DUE TO EXTRUSION OF THE DEVICE THROUGH THE SKIN. AN EXTRUSION WAS FIRST SEEN ON THE (B)(6) 2020. BEFORE DEVICE EXTRUSION THE HEARING PERFORMANCE WITH THE DEVICE WAS GOOD. THE USER WAS RE-IMPLANTED.

    Patient problems: Insufficient Information; Tissue Breakdown

    MAUDE report 11221470 ↗
  • InjuryDec 9, 2020Output Problem; Output Problem

    REPORTEDLY, THE USER STARTED TO EXPERIENCE A DECREASE IN HEARING PERFORMANCE WITH THE DEVICE IN SUMMER. THE USER WAS RE-IMPLANTED ON (B)(6) 2020.

    Patient problems: Failure of Implant

    MAUDE report 10974322 ↗
  • MalfunctionDec 11, 2019Output Problem; Output Problem

    DURING A ROUTINE FOLLOW-UP APPOINTMENT AFFECTED CHANNELS WERE FOUND. RE-IMPLANTATION IS CONSIDERED BUT NO DATE HAS BEEN SCHEDULED DUE TO THE GENERAL HEALTH CONDITION OF THE RECIPIENT.

    Patient problems: Failure of Implant

    MAUDE report 9450027 ↗
  • InjurySep 12, 2019Output Problem; Output Problem

    USER REPORTED DECREASED HEARING PERFORMANCE FOLLOWING A SHUNT REMOVAL. FLUID IS VISIBLY POOLED AT THE BASE OF THE RECIPIENTS NECK BELOW THE IMPLANT. THERE IS HOWEVER NO FLUID ACCUMULATION OVER THE IMPLANT ITSELF. THE USER'S DETECTION LEVELS ARE UNCHANGED HOWEVER SPEECH MEASURES ARE UNRELIABLE. THE DEVICE HAS BEEN EXPLANTED. THIS REPORT REFERS TO 9710014-2019-00234. FOR FURTHER INFORMATION PLEASE REFER TO THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 9005838 ↗
  • InjuryMar 19, 2019Output Problem; Output Problem

    USER REPORTED DECREASED HEARING PERFORMANCE FOLLOWING A SHUNT REMOVAL. FLUID IS VISIBLY POOLED AT THE BASE OF THE RECIPIENTS NECK BELOW THE IMPLANT. THERE IS HOWEVER NO FLUID ACCUMULATION OVER THE IMPLANT ITSELF. THE USER'S DETECTION LEVELS ARE UNCHANGED HOWEVER SPEECH MEASURES ARE UNRELIABLE.

    Patient problems: Failure of Implant

    MAUDE report 8432152 ↗
  • InjuryNov 29, 2018Output Problem; Output Problem

    HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. DURING A SPEECH THERAPY SESSION ON (B)(6) 2018, THE USER WAS UNABLE TO REPEAT LING SOUNDS. THERE WAS NO REPORT OF A HEAD INJURY. THE PARENT NOTICED THAT DIFFICULTIES WITH HEARING STARTED THAT DAY.

    Patient problems: Failure of Implant

    MAUDE report 8114923 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCE
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product

Other IMPLANT, COCHLEAR devices

All IMPLANT, COCHLEAR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page