Mi1200 SYNCHRONY +COMPRESSED 09408Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This MED-EL Synchrony cochlear implant is placed to help with hearing. Reports mostly describe implants stopping working or losing sound, sometimes needing removal and replacement, and infections including meningitis. Of 24 reports naming its identifier, there are no deaths and 14 mention injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 09008737094089), received between Sep 10, 2019 and Aug 31, 2026 — classified as 14 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Output Problem” (10 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem10
  • Impedance Problem5
  • Adverse Event Without Identified Device or Use Problem3
  • Device Appears to Trigger Rejection2
  • Microbial Contamination of Device2
  • Migration1
  • No Device Output1

Patient problems

  • Failure of Implant13
  • No Clinical Signs, Symptoms or Conditions5
  • Meningitis4
  • Head Injury2
  • Bacterial Infection1
  • Skin Tears1
  • Tissue Breakdown1
  • Undesired Nerve Stimulation1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 31, 2026Impedance Problem

    The user was diagnosed with cochlear malformation. Insertion resistance was encountered during the implantation procedure. The user was reimplanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50516895 ↗
  • MalfunctionJan 30, 2026Impedance Problem

    DURING IMPLANTATION SURGERY IN SITU MEASUREMENTS REPORTEDLY SHOWED AFFECTED CHANNELS. A BACK-UP IMPLANT WAS SUCCESSFULLY IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24219747 ↗
  • MalfunctionNov 24, 2025Impedance Problem; Impedance Problem

    THE RECIPIENT HAS NOT WORN AUDIO PROCESSORS BILATERALLY SINCE 2019 AND WAS NOW ABOUT TO BE REACTIVATED AGAIN. FURTHER PROCEEDINGS ARE CURRENTLY UNKNOWN.

    Patient problems: Head Injury; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23632259 ↗
  • MalfunctionAug 18, 2025Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. FURTHER PROCEEDINGS ARE CURRENTLY UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22827581 ↗
  • InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CHILD IS BILATERALLY IMPLANTED AND HAD MENINGITIS MULTIPLE TIMES, SO IT WAS DECIDED TO REMOVE THE IMPLANT ON THE LEFT SIDE. THE USER WAS EXPLANTED APPROXIMATELY ON (B)(6) 2024. AFTER THE IMPLANT WAS REMOVED, THE CHILD NO LONGER HAD ANY CASES OF MENINGITIS.

    Patient problems: Meningitis

    MAUDE report 22229191 ↗
  • MalfunctionJun 5, 2025No Device Output; No Device Output

    THE DEVICE WAS RECEIVED AT THE MANUFACTURER. ACCORDING TO ADDITIONAL INFORMATION THE CAUSE OF THE EXPLANTATION WAS A TRAUMA. A GATE FELL ON THE USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22147873 ↗
  • InjuryFeb 26, 2024Microbial Contamination of Device; Microbial Contamination of Device

    THE EXPLANTED DEVICE HAS BEEN RECEIVED. ACCORDING TO THE DEVICE EXPLANT REPORT FORM THE USER HAD A PNEUMOCOCCAL MENINGITIS WHERE THE ELECTRODE ARRAY ACTED AS BACTERIAL CONDUCTOR TO THE INTRACRANIAL AREA. REPORTEDLY, THE USER WAS NOT IMPLANTED WITH A NEW DEVICE. FURTHER INFORMATION HAS BEEN REPEATEDLY REQUESTED, BUT HAS NOT BEEN RECEIVED.

    Patient problems: Meningitis

    MAUDE report 18778167 ↗
  • InjuryDec 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER REPORTEDLY HAD 3 INSTANCES OF MENINGITIS WITH CSF GUSHER. FURTHER PROCEEDINGS ARE CURRENTLY UNKNOWN.

    Patient problems: Meningitis

    MAUDE report 18409308 ↗
  • InjuryAug 8, 2023Microbial Contamination of Device; Microbial Contamination of Device

    THE USER REPORTEDLY HAS AN INFECTION AND AN ABSCESS NEAR THE SITE OF THE IMPLANT. THE INFECTION HAS REPORTEDLY SPREAD TO THE LEVEL OF THE IMPLANT. IN (B)(6) 2023, GRANULATION APPEARED AND THE USER'S PARENTS SOUGHT MEDICAL ASSISTANCE EVERY 2-4 WEEKS. ON (B)(6) AN ABSCESS APPEARED AND THE CLINIC PRESCRIBED ANTIBIOTICS TO AVOID A SECOND SURGERY. BETWEEN (B)(6) AND (B)(6), THE ABSCESS BROKE AND THE CLINIC AND THE PARENTS DISCUSSED REMOVING THE DEVICE. THE DEVICE WAS EXPLANTED ON (B)(6) 2023.

    Patient problems: Bacterial Infection

    MAUDE report 17481034 ↗
  • MalfunctionJun 30, 2023Migration; Migration

    THE USER WAS RE-IMPLANTED. PER EXPLANTATION REPORT THERE WERE APPROX. 8 EXTRA-COCHLEAR CHANNELS.

    Patient problems: Failure of Implant

    MAUDE report 17236616 ↗
  • MalfunctionMay 10, 2023Impedance Problem; Impedance Problem

    THE USER_S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED.THE USER WAS REIMPLANTED ON (B)(6)2023

    Patient problems: Failure of Implant

    MAUDE report 16902351 ↗
  • InjuryOct 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT THE DEVICE HAS BEEN EXPLANTED. THE USER HAD 3 POST-OPERATIVE EPISODES OF MENINGITIS ((B)(6) 2022) FOR WHICH HE HAD TO BE HOSPITALIZED EACH TIME AND TREATED ACCORDINGLY. IT WAS THEN DECIDED TO PERFORM REVISION SURGERY AND TO REMOVE THE DEVICE. REPORTEDLY, A CT SCAN SHOWED AN ANOMALY (COMMON CAVITY). AN TRANSORBITAL X-RAY SHOWED THE ELECTRODE INSIDE THE COCHLEA.

    Patient problems: Meningitis

    MAUDE report 15689358 ↗
  • MalfunctionApr 15, 2022Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. IN SITU TESTING SHOWED AFFECTED CHANNELS. REIMPLANTATION IS BEING DISCUSSED.

    Patient problems: Failure of Implant

    MAUDE report 14118052 ↗
  • MalfunctionFeb 3, 2022Output Problem; Output Problem

    IN SITU MEASUREMENTS SHOWED AFFECTED CHANNELS. THE USER DID NOT HAVE GOOD BENEFIT WITH THE DEVICE BEFORE. THE USER HAS HYPOPLASIA ON BOTH SIDES. REPORTEDLY THE USER MAY HAVE SUFFERED A TRAUMA. REIMPLANTATION OR EXPLANTATION SURGERY IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 13438908 ↗
  • InjuryJul 28, 2021Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. NO ACCIDENT OR TRAUMA HAS BEEN REPORTED BUT PER PARENTS THE USER IS WOBBLY ON HER FEET. RECIPIENT WAS REIMPLANTED ON (B)(6) 2021. THIS REPORT REFERS TO 9710014-2020-00656. FOR FURTHER INFORMATION PLEASE REFER TO THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 12237464 ↗
  • InjuryMay 7, 2021Output Problem; Output Problem

    THE USERS PERFORMANCE WITH THE DEVICE IS AFFECTED. RE-IMPLANTATION IS PLANNED BUT NO DATE HAS BEEN MADE AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 11788601 ↗
  • MalfunctionApr 22, 2021Output Problem; Output Problem

    THE USER_S PERFORMANCE WITH THE DEVICE IS AFFECTED. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11705527 ↗
  • InjuryNov 12, 2020Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. NO ACCIDENT OR TRAUMA HAS BEEN REPORTED BUT PER PARENTS THE USER IS WOBBLY ON HER FEET.

    Patient problems: Failure of Implant

    MAUDE report 10827998 ↗
  • InjuryOct 14, 2020Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE RECIPIENT WAS PLANNED TO BE EXPLANTED DUE TO ELECTRODE LEAD EXTRUSION THROUGH THE TYMPANIC MEMBRANE AND NON-USE. THE RECIPIENT NEVER HAD BENEFIT FROM THE DEVICE DUE TO CYSTIC COCHLEA DEFORMITY AND COCHLEAR NERVE DEFICIENCY. EXPECTATIONS AT IMPLANTATION WERE MINIMAL BECAUSE OF THESE CONDITIONS. THE CLINIC HAD BEEN PLANNING EXPLANTATION FOR A WHILE. NO NEW DEVICE WILL BE IMPLANTED. THE DEVICE WAS EXPLANTED (B)(6) 2020.

    Patient problems: Skin Tears

    MAUDE report 10676268 ↗
  • InjurySep 29, 2020Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE ELECTRODE LEAD EXTRUDED THROUGH THE TYMPANIC MEMBRANE. THE DEVICE WAS EXPLANTED ON (B)(6) 2020.

    Patient problems: Failure of Implant; Tissue Breakdown

    MAUDE report 10599679 ↗
  • InjuryAug 5, 2020Output Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A SEVERE LEFT-SIDED HEAD TRAUMA. THE USER WILL BE RE-IMPLANTED ON THE (B)(6).

    Patient problems: Failure of Implant; Head Injury

    MAUDE report 10366104 ↗
  • InjurySep 25, 2019Output Problem; Output Problem

    REPORTEDLY THE SOUND WITH THE DEVICE BECAME TO SOFT STARTING IN (B)(6) 2019. PRIOR THE RECIPIENT WAS DOING WELL WITH PERFORMANCE INCREASING OVER TIME. THERE HAVE BEEN NO RECENT CHANGES IN HEALTH, BUMPS ON THE HEAD OR PAIN. ADDITIONALLY, THERE WAS NO SWELLING OR REDNESS AT IMPLANT SITE. THERE ARE NO CONCERNS WITH HIS HEARING ON THE CONTRALATERAL EAR. RE-IMPLANTATION TOOK PLACE ON (B)(6) 2019.

    Patient problems: Failure of Implant

    MAUDE report 9114655 ↗
  • MalfunctionSep 23, 2019Output Problem; Output Problem

    WHEN THE USER ATTENDED A FITTING SESSION SHE REPORTED THAT HEARING PERFORMANCE WITH THE DEVICE HAD CHANGED. NO CLEAR SPECIFICATION ON REDUCED PERFORMANCE WAS MENTIONED. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 9107446 ↗
  • InjurySep 10, 2019Output Problem; Output Problem

    REPORTEDLY THE SOUND WITH THE DEVICE BECAME TOO SOFT STARTING IN AUGUST 2019. PRIOR THE RECIPIENT WAS DOING WELL WITH PERFORMANCE INCREASING OVER TIME. THE RECIPIENT HAD 6 CHANNELS LEFT FOR STIMULATION, BUT DURING MOST COMFORTABLE LOUDNESS MEASURES THE REMAINING CHANNELS (1-6) CAUSED FACIAL NERVE STIMULATION AND THE LEVELS WERE STILL TOO SOFT. TRIPHASIC STIMULATION AND RE-FITTINGS DID NOT RESOLVE STIMULATION OF THE FACIAL NERVE.RE-IMPLANTATION TOOK PLACE ON (B)(6), 2019. ONE WEEK POST ACTIVATION THERE WAS AN IMPROVEMENT IN AIDED THRESHOLDS AND THE USER REPORTED IMPROVED CLARITY. THE USER REPORTED THE SOUND WAS INITIALLY TOO SOFT BUT AFTER MAPPING AN IMPROVEMENT IN LOUDNESS AND CLARITY WAS REPORTED.

    Patient problems: Failure of Implant; Undesired Nerve Stimulation

    MAUDE report 8978934 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCE

Other IMPLANT, COCHLEAR devices

All IMPLANT, COCHLEAR devices →