Skip to content

Mi1210 SYNCHRONY ST +MEDIUM 33183Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This MED-EL SYNCHRONY cochlear implant is a surgically implanted device that restores hearing through electrical signals. Most of the 72 reports describe the implant ceasing to work properly, with reduced hearing performance, some after head injuries, and in a few cases infection that led to the device being removed. The reports include 33 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
73
Deaths reported
0
Injuries reported
33
Recalls naming it
0

Summary

The FDA's MAUDE database contains 73 adverse event reports that list this exact device (DI 09008737331832), received between Oct 22, 2018 and Sep 11, 2026 — classified as 33 injuries, 40 malfunctions.

The most frequently reported specific device problem is “Output Problem” (50 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem50
  • Impedance Problem24
  • Migration5
  • Circuit Failure1
  • Device Appears to Trigger Rejection1

Patient problems

  • Failure of Implant69
  • Head Injury6
  • No Clinical Signs, Symptoms or Conditions4
  • Medical device site infection1

Recent adverse event reports

The 25 most recent of 73 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Impedance Problem

    The user's hearing performance with the device is affected.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50726094 ↗
  • InjuryMay 7, 2026Device Appears to Trigger Rejection

    SINCE THE IMPLANTATION SURGERY, THE USER HAD RECURRING INFECTION. THE DEVICE WAS EXPLANTED.

    Patient problems: Medical device site infection

    MAUDE report 25108680 ↗
  • MalfunctionFeb 3, 2026Impedance Problem

    HEARING PERFORMANCE IS AFFECTED AFTER TRAUMA.

    Patient problems: Head Injury

    MAUDE report 24245044 ↗
  • MalfunctionNov 10, 2025Impedance Problem; Impedance Problem

    IN SITU TESTING SHOWED AFFECTED CHANNELS. A RE-IMPLANTATION IS BEING CONSIDERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23508549 ↗
  • MalfunctionNov 6, 2024Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A TRAUMA TO THE HEAD. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 20618840 ↗
  • MalfunctionSep 26, 2024Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED.

    Patient problems: Failure of Implant

    MAUDE report 20309455 ↗
  • MalfunctionSep 20, 2024Impedance Problem; Impedance Problem

    HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20265177 ↗
  • MalfunctionAug 21, 2024Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A HEAD TRAUMA IN MAY 2024. RE-IMPLANTATION IS PLANNED, BUT NO DATE HAS BEEN SCHEDULED YET.

    Patient problems: Failure of Implant

    MAUDE report 20010349 ↗
  • MalfunctionAug 9, 2024Migration; Migration

    THE USER WAS FITTED WITH 8 ACTIVE CHANNELS DURING HER FIRST FITTING. OVER TIME, THE NUMBER OF CHANNELS USED DECREASED. CURRENTLY SHE IS USING 4 ACTIVE CHANNELS. REPORTEDLY, THERE ARE SEVERAL EXTRA-COCHLEAR CHANNELS. THE SURGEON PLANS TO PERFORM A REVISION SURGERY, BUT ALSO A BACK-UP DEVICE HAS BEEN REQUESTED.

    Patient problems: Failure of Implant

    MAUDE report 19947584 ↗
  • MalfunctionMay 21, 2024Impedance Problem; Impedance Problem

    THE USERS HEARING PERFORMANCE DECREASED. THERE IS NO REPORT OF SPECIFIC TRAUMA. THERE IS NO SWELLING OR REDNESS OVER THE IMPLANT AREA. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 19359981 ↗
  • MalfunctionJan 15, 2024Impedance Problem; Impedance Problem

    THE USER FELL OUT OF HIS BED AND HIT HIS HEAD ON THE WARDROBE.

    Patient problems: Failure of Implant

    MAUDE report 18510108 ↗
  • MalfunctionDec 27, 2023Impedance Problem; Impedance Problem

    APPROXIMATELY ONE MONTH AGO (B)(6) 2023), THE USER HAD A HEAD TRAUMA ON THE CONTRA-LATERAL SIDE. DESPITE THIS, THE USER CONTINUED TO USE THE IMPLANT WITHOUT INTERRUPTION. ONE WEEK AGO (MID (B)(6) 2023), THE USER HAD AN OTITIS INFECTION ON THE CONTRA-LATERAL SIDE. THE USER ATTENDED A ROUTINE APPOINTMENT ON (B)(6) 2023, REPORTING NO ISSUES WITH THE IMPLANT'S FUNCTIONALITY. DURING THE APPOINTMENT, AN AUDIOLOGIST CONDUCTED A FREE FIELD THRESHOLD AUDIOMETRY TEST ON BOTH SIDES INDIVIDUALLY, YIELDING SIMILAR RESULTS FOR EACH EAR: 20DB, 20DB, 15DB, 20DB, 15DB, 15DB, 15DB, 15DB.

    Patient problems: Failure of Implant

    MAUDE report 18398073 ↗
  • MalfunctionDec 20, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE HAS BEEN DECREASING OVER THE PAST 2 YEARS. CURRENTLY THE USER DOES NOT HAVE ACCESS TO SOUND. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 18362081 ↗
  • MalfunctionOct 18, 2023Impedance Problem; Impedance Problem

    ON (B)(6) 2023, THE USER CAME TO THE CLINIC FOR A ROUTINE APPOINTMENT WHERE CHANNELS WITH ABNORMAL IMPEDANCES WERE DISCOVERED. IN ADDITION, A GOOD ECAP RESPONSE COULDN'T BE OBSERVED.

    Patient problems: Failure of Implant

    MAUDE report 17955617 ↗
  • MalfunctionSep 21, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A HEAD TRAUMA OCCURRED. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17786842 ↗
  • MalfunctionSep 14, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A TRAUMA OCCURRED ON (B)(6) 2023.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17743733 ↗
  • MalfunctionAug 16, 2023Impedance Problem; Impedance Problem

    HEARING PERFORMANCE WITH THE DEVICE SHOWED AFFECTED CHANNELS. THE USER WAS IN THE EARTHQUAKE IN HATAY IN 2023 AND FELL DOWN SOME STAIRS WHILE ESCAPING. ABOUT 10-15 DAYS LATER THERE WAS NO MORE SOUND FROM THE IMPLANT ON THE LEFT SIDE. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 17555111 ↗
  • MalfunctionJul 5, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE HUSBAND REPORTED THAT THE USER HAS NOT BEEN HAVING GOOD SPEECH COMPREHENSION FOR THE PAST 12-18 MONTHS.

    Patient problems: Failure of Implant

    MAUDE report 17256662 ↗
  • MalfunctionFeb 6, 2023Impedance Problem; Impedance Problem

    A ROUTINE FITTING WAS PERFORMED ON (B)(6) 2023 WHERE AFFECTED CHANNELS WERE SEEN. THE PARENTS DID NOT NOTICE ANY HEARING DEFICIT ON THE LEFT SIDE.

    Patient problems: Failure of Implant

    MAUDE report 16302895 ↗
  • MalfunctionFeb 2, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE IS AFFECTED AFTER A HEAD TRAUMA OCCURRED IN (B)(6) 2022.

    Patient problems: Failure of Implant

    MAUDE report 16282314 ↗
  • MalfunctionJan 11, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AND CHANNELS WITH ABNORMAL IMPEDANCES HAVE BEEN OBSERVED. RE-IMPLANTATION IS CONSIDERED, HOWEVER A DATE FOR SURGERY HAS NOT BEEN SCHEDULED YET.

    Patient problems: Failure of Implant

    MAUDE report 16136381 ↗
  • MalfunctionOct 27, 2022Impedance Problem; Output Problem; Impedance Problem; Output Problem

    AROUND (B)(6) 2022, THE USER REPORTED THAT SHE COULD NO LONGER HEAR WITH THE DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 15680257 ↗
  • MalfunctionSep 9, 2022Migration; Migration

    THE USER REPORTED THAT RECENTLY THE IMPLANT HOUSING HAS MOVED DOWN TOWARDS THE PINNA. A REVISION SURGERY FOR AN IMPLANT BED IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 15388783 ↗
  • MalfunctionSep 7, 2022Output Problem; Output Problem

    THE MOTHER REPORTED THAT THE USER COULD NO LONGER HEAR WITH THE DEVICE. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 15368162 ↗
  • MalfunctionSep 7, 2022Output Problem; Output Problem

    THE USER REPORTED THAT HE COULD HARDLY HEAR WITH THE DEVICE. AN X-RAY WILL BE TAKEN. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 15368045 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard

Other Implant, Cochlear devices

All Implant, Cochlear devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page