Mi1210 SYNCHRONY ST +MEDIUM 33183Recalls & reported problems
Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This MED-EL SYNCHRONY cochlear implant is a surgically implanted device that restores hearing through electrical signals. Most of the 72 reports describe the implant ceasing to work properly, with reduced hearing performance, some after head injuries, and in a few cases infection that led to the device being removed. The reports include 33 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 73
- Deaths reported
- 0
- Injuries reported
- 33
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 73 adverse event reports that list this exact device (DI 09008737331832), received between Oct 22, 2018 and Sep 11, 2026 — classified as 33 injuries, 40 malfunctions.
The most frequently reported specific device problem is “Output Problem” (50 reports).
4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem50
- Impedance Problem24
- Migration5
- Circuit Failure1
- Device Appears to Trigger Rejection1
Patient problems
- Failure of Implant69
- Head Injury6
- No Clinical Signs, Symptoms or Conditions4
- Medical device site infection1
Recent adverse event reports
The 25 most recent of 73 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Impedance Problem
The user's hearing performance with the device is affected.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50726094 ↗InjuryMay 7, 2026Device Appears to Trigger Rejection
SINCE THE IMPLANTATION SURGERY, THE USER HAD RECURRING INFECTION. THE DEVICE WAS EXPLANTED.
Patient problems: Medical device site infection
MAUDE report 25108680 ↗MalfunctionFeb 3, 2026Impedance Problem
MalfunctionNov 10, 2025Impedance Problem; Impedance Problem
IN SITU TESTING SHOWED AFFECTED CHANNELS. A RE-IMPLANTATION IS BEING CONSIDERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23508549 ↗MalfunctionNov 6, 2024Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A TRAUMA TO THE HEAD. RE-IMPLANTATION IS CONSIDERED.
Patient problems: Failure of Implant
MAUDE report 20618840 ↗MalfunctionSep 26, 2024Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED.
Patient problems: Failure of Implant
MAUDE report 20309455 ↗MalfunctionSep 20, 2024Impedance Problem; Impedance Problem
HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. RE-IMPLANTATION IS CONSIDERED.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 20265177 ↗MalfunctionAug 21, 2024Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A HEAD TRAUMA IN MAY 2024. RE-IMPLANTATION IS PLANNED, BUT NO DATE HAS BEEN SCHEDULED YET.
Patient problems: Failure of Implant
MAUDE report 20010349 ↗MalfunctionAug 9, 2024Migration; Migration
THE USER WAS FITTED WITH 8 ACTIVE CHANNELS DURING HER FIRST FITTING. OVER TIME, THE NUMBER OF CHANNELS USED DECREASED. CURRENTLY SHE IS USING 4 ACTIVE CHANNELS. REPORTEDLY, THERE ARE SEVERAL EXTRA-COCHLEAR CHANNELS. THE SURGEON PLANS TO PERFORM A REVISION SURGERY, BUT ALSO A BACK-UP DEVICE HAS BEEN REQUESTED.
Patient problems: Failure of Implant
MAUDE report 19947584 ↗MalfunctionMay 21, 2024Impedance Problem; Impedance Problem
THE USERS HEARING PERFORMANCE DECREASED. THERE IS NO REPORT OF SPECIFIC TRAUMA. THERE IS NO SWELLING OR REDNESS OVER THE IMPLANT AREA. RE-IMPLANTATION IS CONSIDERED.
Patient problems: Failure of Implant
MAUDE report 19359981 ↗MalfunctionJan 15, 2024Impedance Problem; Impedance Problem
THE USER FELL OUT OF HIS BED AND HIT HIS HEAD ON THE WARDROBE.
Patient problems: Failure of Implant
MAUDE report 18510108 ↗MalfunctionDec 27, 2023Impedance Problem; Impedance Problem
APPROXIMATELY ONE MONTH AGO (B)(6) 2023), THE USER HAD A HEAD TRAUMA ON THE CONTRA-LATERAL SIDE. DESPITE THIS, THE USER CONTINUED TO USE THE IMPLANT WITHOUT INTERRUPTION. ONE WEEK AGO (MID (B)(6) 2023), THE USER HAD AN OTITIS INFECTION ON THE CONTRA-LATERAL SIDE. THE USER ATTENDED A ROUTINE APPOINTMENT ON (B)(6) 2023, REPORTING NO ISSUES WITH THE IMPLANT'S FUNCTIONALITY. DURING THE APPOINTMENT, AN AUDIOLOGIST CONDUCTED A FREE FIELD THRESHOLD AUDIOMETRY TEST ON BOTH SIDES INDIVIDUALLY, YIELDING SIMILAR RESULTS FOR EACH EAR: 20DB, 20DB, 15DB, 20DB, 15DB, 15DB, 15DB, 15DB.
Patient problems: Failure of Implant
MAUDE report 18398073 ↗MalfunctionDec 20, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE HAS BEEN DECREASING OVER THE PAST 2 YEARS. CURRENTLY THE USER DOES NOT HAVE ACCESS TO SOUND. THE USER WAS RE-IMPLANTED.
Patient problems: Failure of Implant
MAUDE report 18362081 ↗MalfunctionOct 18, 2023Impedance Problem; Impedance Problem
ON (B)(6) 2023, THE USER CAME TO THE CLINIC FOR A ROUTINE APPOINTMENT WHERE CHANNELS WITH ABNORMAL IMPEDANCES WERE DISCOVERED. IN ADDITION, A GOOD ECAP RESPONSE COULDN'T BE OBSERVED.
Patient problems: Failure of Implant
MAUDE report 17955617 ↗MalfunctionSep 21, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A HEAD TRAUMA OCCURRED. THE USER WAS RE-IMPLANTED.
Patient problems: Failure of Implant
MAUDE report 17786842 ↗MalfunctionSep 14, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A TRAUMA OCCURRED ON (B)(6) 2023.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 17743733 ↗MalfunctionAug 16, 2023Impedance Problem; Impedance Problem
HEARING PERFORMANCE WITH THE DEVICE SHOWED AFFECTED CHANNELS. THE USER WAS IN THE EARTHQUAKE IN HATAY IN 2023 AND FELL DOWN SOME STAIRS WHILE ESCAPING. ABOUT 10-15 DAYS LATER THERE WAS NO MORE SOUND FROM THE IMPLANT ON THE LEFT SIDE. RE-IMPLANTATION IS CONSIDERED.
Patient problems: Failure of Implant
MAUDE report 17555111 ↗MalfunctionJul 5, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE HUSBAND REPORTED THAT THE USER HAS NOT BEEN HAVING GOOD SPEECH COMPREHENSION FOR THE PAST 12-18 MONTHS.
Patient problems: Failure of Implant
MAUDE report 17256662 ↗MalfunctionFeb 6, 2023Impedance Problem; Impedance Problem
A ROUTINE FITTING WAS PERFORMED ON (B)(6) 2023 WHERE AFFECTED CHANNELS WERE SEEN. THE PARENTS DID NOT NOTICE ANY HEARING DEFICIT ON THE LEFT SIDE.
Patient problems: Failure of Implant
MAUDE report 16302895 ↗MalfunctionFeb 2, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE IS AFFECTED AFTER A HEAD TRAUMA OCCURRED IN (B)(6) 2022.
Patient problems: Failure of Implant
MAUDE report 16282314 ↗MalfunctionJan 11, 2023Impedance Problem; Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AND CHANNELS WITH ABNORMAL IMPEDANCES HAVE BEEN OBSERVED. RE-IMPLANTATION IS CONSIDERED, HOWEVER A DATE FOR SURGERY HAS NOT BEEN SCHEDULED YET.
Patient problems: Failure of Implant
MAUDE report 16136381 ↗MalfunctionOct 27, 2022Impedance Problem; Output Problem; Impedance Problem; Output Problem
AROUND (B)(6) 2022, THE USER REPORTED THAT SHE COULD NO LONGER HEAR WITH THE DEVICE.
Patient problems: Failure of Implant
MAUDE report 15680257 ↗MalfunctionSep 9, 2022Migration; Migration
THE USER REPORTED THAT RECENTLY THE IMPLANT HOUSING HAS MOVED DOWN TOWARDS THE PINNA. A REVISION SURGERY FOR AN IMPLANT BED IS CONSIDERED.
Patient problems: Failure of Implant
MAUDE report 15388783 ↗MalfunctionSep 7, 2022Output Problem; Output Problem
THE MOTHER REPORTED THAT THE USER COULD NO LONGER HEAR WITH THE DEVICE. RE-IMPLANTATION IS CONSIDERED.
Patient problems: Failure of Implant
MAUDE report 15368162 ↗MalfunctionSep 7, 2022Output Problem; Output Problem
THE USER REPORTED THAT HE COULD HARDLY HEAR WITH THE DEVICE. AN X-RAY WILL BE TAKEN. RE-IMPLANTATION IS CONSIDERED.
Patient problems: Failure of Implant
MAUDE report 15368045 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.
- Z-2356-2024 · Apr 18, 2024 · Open, ClassifiedMi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
- Z-2136-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
- Z-2138-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
- Z-2137-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
Other Implant, Cochlear devices
- HiRes™ Ultra CI HiFocus™ MS Electrode CI-1600-04Advanced Bionics AG · Implant, Cochlear2,918 reports
- Cochlear® Nucleus® CI512 cochlear ImplantCochlear Limited · IMPLANT, COCHLEAR2,771 reports
- Cochlear® Nucleus® CI622 cochlear Implant with Slim Straight electrodeCochlear Limited · Implant, cochlear2,115 reports
- Cochlear® Nucleus® CI632 cochlear Implant with Slim Modiolar electrodeCochlear Limited · Implant, cochlear2,057 reports
- Nucleus Z60353Cochlear Limited · IMPLANT, COCHLEAR2 deaths1,815 reports
- Cochlear® Nucleus® CI522 cochlear Implant with Slim Straight electrodeCochlear Limited · IMPLANT, COCHLEAR1,475 reports
- Cochlear® Nucleus® CI612 cochlear ImplantCochlear Limited · Implant, cochlear1,150 reports
- Nucleus Z190574Cochlear Limited · IMPLANT, COCHLEAR981 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page