Mi1200 SYNCHRONY +FLEX26 36617Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 09008737366179), received between May 14, 2021 and Nov 20, 2025 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Impedance Problem” (7 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Impedance Problem7
  • Adverse Event Without Identified Device or Use Problem2
  • Output Problem2
  • Migration1

Patient problems

  • Failure of Implant7
  • No Clinical Signs, Symptoms or Conditions3
  • Bacterial Infection1
  • Dizziness1
  • Head Injury1
  • Pain1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2025Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. FURTHER PROCEEDINGS ARE CURRENTLY UNKNOWN. THE CLINIC WILL PERFORM X-RAY IMAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23605825 ↗
  • MalfunctionSep 19, 2025Migration; Migration

    DURING MAPPING IN (B)(6) 2025 ONLY LIMITED RESPONSES WERE OBSERVED AND THE USER REPORTED DIZZINESS UPON INCREASING THE STIMULATION LEVELS.

    Patient problems: Dizziness

    MAUDE report 23097094 ↗
  • MalfunctionDec 3, 2024Impedance Problem; Impedance Problem

    THE USER REPORTED A GRADUAL DECLINE IN HEARING PERFORMANCE WITH THE DEVICE. THE USER WAS EXPLANTED ON (B)(6) 2024.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20832962 ↗
  • MalfunctionSep 2, 2024Impedance Problem; Impedance Problem

    AFTER A TRAUMA THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED. THE USER WAS REIMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 20118065 ↗
  • MalfunctionAug 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER_S REHABILITATION PROGRESS IS NOT MEETING EXPECTATIONS. THE USER WAS RE-IMPLANTED ON (B)(6) 2024. REPORTEDLY, THE REASON FOR EXPLANTATION WAS PAIN, WHICH WAS PRESENT WHEN THE DEVICE WAS USED, EVEN WITH LOW QU LEVELS.

    Patient problems: Pain

    MAUDE report 19979451 ↗
  • MalfunctionJul 24, 2024Impedance Problem; Impedance Problem

    THE HEARING PERFORMANCE WITH THE LEFT-SIDED DEVICE IS AFFECTED. TRAUMA WAS REPORTED. THERE WAS NO SWELLING OR REDNESS AROUND THE IMPLANT AREA. THE USER WAS RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 19814913 ↗
  • MalfunctionOct 3, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER A TRAUMA TO THE LEFT SIDE OF THE HEAD. THE USER'S PARENTS ALSO REPORTED REDNESS AT THE SIDE OF THE IMPLANT. REIMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 17859888 ↗
  • MalfunctionAug 21, 2023Impedance Problem; Impedance Problem

    THE USER'S HEARING PERFORMANCE IS AFFECTED AFTER A TRAUMA TO THE HEAD. THE USER HAS COCHLEAR ANAPLASIA. RE-IMPLANTATION IS BEING CONSIDERED.

    Patient problems: Failure of Implant; Head Injury; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17585528 ↗
  • MalfunctionMay 30, 2022Impedance Problem; Output Problem; Impedance Problem; Output Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE USER'S MOTHER NOTICED A CHANGE IN HEARING WITH THE DEVICE AFTER TRAUMA TO THE HEAD. CT IMAGING WILL BE PERFORMED AND REIMPLANTATION SURGERY IS BEING CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 14539632 ↗
  • InjurySep 8, 2021Output Problem; Output Problem

    DURING THE FIRST MAPPING HIGH IMPEDANCES WERE SEEN ON MULTIPLE CHANNELS, THEREFORE THE CLINIC DECIDED TO REIMPLANT THE USER. REPORTEDLY, NO PROBLEMS WERE ENCOUNTERED DURING IMPLANTATION SURGERY. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12438043 ↗
  • InjuryMay 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER HAD A CHRONIC INFECTION IN THE BLIND SAC CLOSURE SITE. THE USER WAS EXPLANTED ON (B)(6) 2021 AND IT IS PLANNED TO RE-IMPLANT AT A LATER DATE.

    Patient problems: Bacterial Infection

    MAUDE report 11824216 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCE
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product

Other Implant, Cochlear devices

All Implant, Cochlear devices →