Skip to content

Mi1250 SYNCHRONY 2 MEDIUM 36674Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
1
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 09008737366742), received between May 6, 2021 and Sep 24, 2026 — classified as 1 death, 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Impedance Problem” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Impedance Problem3
  • Device Appears to Trigger Rejection2
  • Insufficient Device Problem Information1
  • Migration1

Patient problems

  • Failure of Implant3
  • Skin Tears2
  • Bacterial Infection1
  • Cerebrospinal Fluid Leakage1
  • Insufficient Information1
  • Meningitis1
  • Swelling/ Edema1
  • Tissue Breakdown1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • DeathSep 24, 2026Insufficient Device Problem Information

    The recipient was implanted on (b)(6) 2026. Approximately 3 weeks post-OP, she developed meningitis. She was hospitalized and died within a few days.

    Patient problems: Meningitis

    MAUDE report 50883832 ↗
  • InjuryNov 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER IS UNABLE TO WEAR THE EXTERNAL DEVICE DUE TO SWELLING CAUSED BY INFLAMMATION AROUND THE IMPLANT AREA. THE USER HAS BEEN REIMPLANTED.

    Patient problems: Swelling/ Edema

    MAUDE report 23591413 ↗
  • InjuryMay 2, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    REPORTEDLY, THE USER HAD AN INFECTION AND EXTRUSION OF THE IMPLANT. THE USER WAS EXPLANTED AND WILL BE RE-IMPLANTED AS SOON AS AN IMPLANT IS AVAILABLE.

    Patient problems: Skin Tears

    MAUDE report 21947820 ↗
  • MalfunctionSep 30, 2024Impedance Problem; Impedance Problem

    IN SITU TESTING SHOWED AFFECTED CHANNELS. THE USER WAS REIMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 20331551 ↗
  • MalfunctionAug 29, 2024Migration; Migration

    THE CLINIC REPORTS THAT THE REASON FOR EXPLANTATION IS DUE TO A MASTOID ABSCESS WHICH THE IMPLANT HAD SLIPPED DOWN INTO.

    Patient problems: Tissue Breakdown

    MAUDE report 20094701 ↗
  • MalfunctionJun 7, 2024Impedance Problem; Adverse Event Without Identified Device or Use Problem; Impedance Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE USER UNDERWENT REVISION A FEW DAYS AFTER IMPLANTATION DUE TO CSF LEAKAGE TO CLOSE THE SEALING. A HEMATOMA WAS PRESENT AS WELL. DURING THE REVISION SURGERY, THE ELECTRODE ARRAY WAS FOUND OUTSIDE THE COCHLEA. REIMPLANTATION WITH A FORM ELECTRODE IS BEING CONSIDERED DUE TO POSSIBLE DAMAGE TO THE ACTIVE ELECTRODE.

    Patient problems: Cerebrospinal Fluid Leakage; Failure of Implant

    MAUDE report 19489457 ↗
  • InjuryOct 31, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    SKIN PROBLEMS APPEARED 6 DAYS POST IMPLANTATION (REDNESS). THE ENTS TREATED THE SKIN AND HEALING IMPROVED GRADUALLY. BUT IN (B)(6) 2023 THE PARENTS CAME TO THE CLINIC AND THE ENTS NOTED THAT THE IMPLANT HAD EXTRUDED. EXPLANTATION WILL BE PLANNED.

    Patient problems: Skin Tears

    MAUDE report 18038017 ↗
  • MalfunctionApr 24, 2023Impedance Problem; Impedance Problem

    THE PARENTS REPORTED A DECREASE IN THE CHILD'S HEARING BENEFIT WITH THE DEVICE. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant

    MAUDE report 16793839 ↗
  • InjuryMay 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER SUFFERED FROM AN INFECTION. IN (B)(6) 2021 AN INFECTION OF THE MASTOID STARTED AND SOME PURULENT FLUID CAME OUT OF THE WOUND FROM THE COCHLEAR IMPLANT SURGERY. THE INFECTION WAS DETERMINED TO BE A STAPHYLOCOCCUS INFECTION, AS CONFIRMED BY BACTERIOLOGICAL EXAMINATION. THE RECIPIENT WAS TREATED WITH ANTIBIOTICS, BUT THE INFECTION RETURNED ON ABOUT (B)(6) 2021. THE MASTOID AND SKIN FLAP WERE AFFECTED BY THE INFECTION. THE USER WAS EXPLANTED ON THE CONCERNED LEFT SIDE ON THE (B)(6) 2021 AND WAS IMPLANTED ON THE CONTRA-LATERAL SIDE DURING THE SAME SURGERY.

    Patient problems: Bacterial Infection; Insufficient Information

    MAUDE report 11782520 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCE
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge

Other Implant, Cochlear devices

All Implant, Cochlear devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page