Mi1260 SONATA 2 FLEX26 38544Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 09008737385446), received between May 22, 2024 and Sep 8, 2026 — classified as 6 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Impedance Problem” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Impedance Problem3
  • Migration2
  • Device Appears to Trigger Rejection1
  • Device Dislodged or Dislocated1
  • Insufficient Device Problem Information1
  • Microbial Contamination of Device1
  • No Device Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Failure of Implant3
  • Skin Infection2
  • Bacterial Infection1
  • Facial Paralysis1
  • Fibrosis1
  • Head Injury1
  • Hematoma1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2026No Device Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50670065 ↗
  • MalfunctionApr 15, 2026Device Dislodged or Dislocated

    The user, who is bilaterally implanted, presented two months ago with a spontaneous hematoma, without trauma, in the right temporal region over the magnet of the right cochlear implant. A revision surgery was performed to clean and drain the hematoma in (b)(6) 2026. During the procedure, it was observed that the magnet was outside the pocket.

    Patient problems: Hematoma; Swelling/ Edema

    MAUDE report 24882533 ↗
  • InjuryMar 12, 2026Adverse Event Without Identified Device or Use Problem

    THE LEFT IMPLANT WAS EXPLANTED DUE TO AN INFECTION THAT COULD NOT BE CONTROLLED. ACCORDING TO THE EXPLANTATION REPORT, THE USER ALSO HAD PAIN.

    Patient problems: Post Operative Wound Infection

    MAUDE report 24577708 ↗
  • MalfunctionJan 14, 2026Impedance Problem

    THE USER SUFFERED FROM A TRAUMA.

    Patient problems: Head Injury

    MAUDE report 24065671 ↗
  • MalfunctionSep 11, 2025Impedance Problem; Impedance Problem

    THE USER REPORTS A DECREASE IN SOUND PERCEPTION. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Fibrosis; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23023971 ↗
  • InjuryAug 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A RECENT FITTING SESSION, THE AUDIOLOGIST ADJUSTED THE STIMULATION PARAMETERS BY MODIFYING THE MINIMUM PULSE DURATION AND SWITCHING FROM BIPHASIC TO TRIPHASIC PULSES. DESPITE THESE ADJUSTMENTS, THE USER CONTINUED TO EXPERIENCE FACIAL NERVE STIMULATION, BOTH WITH AND WITHOUT THE AUDIO PROCESSOR (AP) IN PLACE. FOLLOWING A CALIBRATION SESSION, THE USER INITIALLY REPORTED AN IMPROVEMENT. HOWEVER, THE FACIAL STIMULATION SYMPTOMS REAPPEARED A FEW DAYS LATER AND THE USER IS STILL SUFFERING BY FNS AT THE MOMENT. FURTHER PROCEEDINGS ARE CURRENTLY UNKNOWN.

    Patient problems: Facial Paralysis

    MAUDE report 22892979 ↗
  • InjuryJun 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT DEVELOPED A SKIN INFECTION AROUND THE AUDIO PROCESSOR. REPORTEDLY, THERE WAS A SUBCUTANEOUS INFLAMMATION. THE SKIN INFECTION STARTED IN (B)(6) 2025. ANTIBIOTIC TREATMENT WAS GIVEN, BUT DID NOT RESOLVE THE ISSUE, AND THEREFORE THE DEVICE WAS EXPLANTED. RE-IMPLANTATION IS PLANNED AT A LATER POINT IN TIME.

    Patient problems: Skin Infection

    MAUDE report 22183323 ↗
  • MalfunctionMay 9, 2025Migration; Migration

    DURING THE LAST FITTING APPOINTMENT HIGH IMPEDANCES IN THE BASAL CHANNELS WERE OBSERVED AND THE RECIPIENT HAD NOT HEARING SENSATION WITH THE DEVICE. A REVISION SURGERY WAS PERFORMED TO REINSERT THE ELECTRODE AND USED A FIXATION CLIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21992586 ↗
  • InjuryApr 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT HAD A SKIN INFECTION AT THE IMPLANT SITE FOR THE SECOND TIME. THERE WAS A SKIN NECROSIS WITH OPENING OF THE SKIN. ANTIBIOTIC TREATMENT WAS PROVIDED, HOWEVER DID NOT RESOLVE THE ISSUE. CULTURES WERE PERFORMED AND SHOWED PSEUDOMONAS AERUGINOSA. THE RECIPIENT HAS BEEN EXPLANTED.

    Patient problems: Skin Infection

    MAUDE report 21785009 ↗
  • MalfunctionFeb 14, 2025Impedance Problem; Impedance Problem

    DURING A FOLLOW-UP APPOINTMENT, IN SITU TESTING WAS PERFORMED WHERE AFFECTED CHANNELS WERE OBSERVED. THE RECIPIENT WILL BE MONITORED, RE-IMPLANTATION MIGHT BE CONSIDERED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21382732 ↗
  • InjuryDec 9, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED. THE USER VISITED THE CLINIC WITH EAR SUPPURATION AND AN EXTRUDED ELECTRODE LEAD IN THE EXTERNAL AUDITORY CANAL (EAC). THE USER HAS BEEN EXPLANTED. ACCORDING TO THE EXPLANT REPORT FORM, THE ELECTRODE ARRAY WAS FOUND OUTSIDE THE COCHLEA PRIOR TO REMOVAL.

    Patient problems: Skin Tears

    MAUDE report 20875822 ↗
  • MalfunctionAug 30, 2024Migration; Migration

    THE RECIPIENT'S SPEECH UNDERSTANDING AND PERFORMANCE WITH THE DEVICE HAS GONE POORER ALL THE TIME AND SOUND WAS NOT GOOD AT ALL. RE-INSERTION SCHEDULED FOR (B)(6) 2024.

    Patient problems: Failure of Implant

    MAUDE report 20104956 ↗
  • MalfunctionAug 9, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE RECIPIENT REPORTED THAT HE DOES NOT HEAR ANYTHING AND HAS NO BENEFIT FROM THE DEVICE. CURRENTLY, THE RECIPIENT DOES NOT RESPOND TO ANY CHARGE INCREASE AND REPORTED DISCOMFORT UNDER CHARGE INCREASE. EABR TESTING IS CONSIDERED. THE SURGEON HAS SCHEDULED AN APPOINTMENT FOR (B)(6) 2024.

    Patient problems: Failure of Implant

    MAUDE report 19947354 ↗
  • InjuryMay 22, 2024Microbial Contamination of Device; Microbial Contamination of Device

    THE RECIPIENT HAD A SKIN INFECTION WITH NECROSIS AT THE IMPLANT SITE AND WAS TREATED WITH ANTIBIOTICS FOR 1 MONTH. THE DEVICE WAS VISIBLE THROUGH THE WOUND. REPORTEDLY, THE SURGICAL WOUND COMPLETELY HEALED AFTER INITIAL IMPLANTATION. THE RECIPIENT HAS BEEN EXPLANTED. THE RECIPIENT HAS BEEN IMPLANTED ON THE CONTRA-LATERAL SIDE.

    Patient problems: Bacterial Infection

    MAUDE report 19369443 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi1000 CONCE
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product

Other Implant, Cochlear devices

All Implant, Cochlear devices →