Skip to content

Mi1250 SYNCHRONY 2 FLEXsoft S-Vector Magnet 39541Recalls & reported problems

Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Med-El Synchrony 2 cochlear implant has 42 adverse event reports, with 13 injuries and no deaths. Reports mostly describe the implant migrating or the electrode array coming out of place, device failures after head trauma, and wound-healing problems, often leading to removal and reimplantation. There is 1 open recall, started April 2024, because some devices may develop a crack in the internal electronics that could stop them working.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
14
Recalls naming it
1

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 09008737395414), received between Dec 15, 2020 and Sep 25, 2026 — classified as 14 injuries, 28 malfunctions.

The most frequently reported specific device problem is “Migration” (11 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 3 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 18, 2024Class IIOpen, Classified

    A small number of devices may have a higher probability of a crack within the internal electronics substrate which could result in a non-functional device.

    Z-2356-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration11
  • Impedance Problem10
  • Adverse Event Without Identified Device or Use Problem8
  • Device Appears to Trigger Rejection6
  • Circuit Failure3
  • Insufficient Device Problem Information3
  • Output Problem3
  • Device Dislodged or Dislocated2

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Failure of Implant12
  • Skin Tears4
  • Head Injury3
  • Bacterial Infection2
  • Discomfort2
  • Undesired Nerve Stimulation2
  • Unspecified Infection2

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Impedance Problem

    The user's hearing performance is affected. During the fitting session the IFT showed 7 channels in Hi impedance. The user will go back to the clinic to have a second check.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895636 ↗
  • InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem

    Due to a wound-healing complication, it was decided to remove the implant housing. A new implant will be inserted after the healing phase.

    Patient problems: Post Operative Wound Infection

    MAUDE report 50447104 ↗
  • MalfunctionJun 12, 2026Migration

    POST-OP IMAGING TAKEN ON THE DAY OF IMPLANTATION PRIOR TO DISCHARGE SHOWED ONLY 5 CHANNELS STILL IN THE COCHLEA. THIS ARRAY WAS REPOSITIONED ON (B)(6), BY THE IMPLANTING SURGEON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25483751 ↗
  • MalfunctionJun 9, 2026Circuit Failure; No Device Output

    CONNECTION TO THE IMPLANT CANNOT BE ESTABLISHED AFTER USER SUSTAINED A TEMPORAL BONE FRACTURE. THE USER WAS REIMPLANTED.

    Patient problems: Head Injury

    MAUDE report 25439540 ↗
  • InjuryMar 4, 2026Device Appears to Trigger Rejection; Improper or Incorrect Procedure or Method

    THE DEVICE HAS BEEN EXPLANTED. THE USER HAS BEEN REIMPLANTED WITH A NEW DEVICE. REPORTEDLY, DURING AN EAR CLEANING THE ELECTRODE ARRAY WAS PULLED OUT OF THE COCHLEA AND WAS DISLOCATED IN THE EAR CANAL. THE USER THEN COULD NO LONGER HEAR WITH THE CI.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Skin Tears

    MAUDE report 24506239 ↗
  • InjuryFeb 27, 2026Adverse Event Without Identified Device or Use Problem

    THE USER WAS EXPLANTED ON (B)(6) 2026 DUE TO INFLAMMATION OF THE CAVITY.

    Patient problems: Inflammation

    MAUDE report 24468263 ↗
  • InjuryDec 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE DEVICE WAS EXPLANTED DUE TO POOR SCAR HEALING AND POST-OPERATIVE DEHISCENCE WITH INFECTION.

    Patient problems: Unspecified Infection; Wound Dehiscence

    MAUDE report 23920574 ↗
  • MalfunctionNov 25, 2025Migration; Migration

    THE USER HAS REPORTED THAT THE SOUND QUALITY HAS DECREASED AND IS EXPERIENCING FACIAL NERVE STIMULATION. THE USER HAS BEEN REIMPLANTED.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 23645886 ↗
  • InjuryNov 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THERE IS AN INFECTION SURROUNDING THE IMPLANT. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 23496145 ↗
  • MalfunctionOct 30, 2025Migration; Migration

    THE USER REPORTED POOR SOUND QUALITY. REINSERTION IS PLANNED FOR (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23423556 ↗
  • MalfunctionOct 21, 2025Impedance Problem; Impedance Problem

    THE USER HAD A FALL. IT IS UNCERTAIN IF HE HIT HIS HEAD. HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED SINCE THEN.

    Patient problems: Head Injury

    MAUDE report 23346341 ↗
  • MalfunctionOct 21, 2025Migration; Migration

    DUE TO A MIGRATION OF THE IMPLANT HOUSING A REVISION SURGERY TOOK PLACE IN WHICH THE IMPLANT WAS RE-POSITIONED. SINCE REPOSITION SURGERY THE USER HAS NO SOUND PERCEPTION ON THE MOST BASAL CHANNEL, WHICH WAS SUBSEQUENTLY DEACTIVATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23345420 ↗
  • MalfunctionSep 19, 2025Impedance Problem; Impedance Problem

    DURING THE FITTING SESSION, AFFECTED CHANNELS WERE FOUND. THERE IS A DECREASE IN THE USER_S PERCEPTION. THE USER WAS RE-IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095584 ↗
  • InjurySep 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CSF WAS LEAKING OUT OF THE COCHLEA.

    Patient problems: Cerebrospinal Fluid Leakage

    MAUDE report 23013692 ↗
  • InjuryAug 21, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE USER HAS BEEN EXPLANTED DUE TO AN EXTRUSION OF THE IMPLANT.

    Patient problems: Skin Tears

    MAUDE report 22858299 ↗
  • MalfunctionJul 7, 2025Migration; Migration

    THE USER HAS NO AUDITORY IMPRESSION. THE USER HAS BEEN REIMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22411926 ↗
  • MalfunctionApr 30, 2025Impedance Problem; Impedance Problem

    THE USER HAS HAD GOOD PERFORMANCE WITH THE IMPLANT, HOWEVER HE DOES NOTICE A PINGING SOUND WITH STIMULATION. THE USER WAS RE-IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21932651 ↗
  • MalfunctionApr 24, 2025Impedance Problem; Impedance Problem

    THE USER HAD AN EPILEPTIC SEIZURE AND FELL ON HER HEAD. SHE THEN HAD A DIFFERENT AUDITORY IMPRESSION. THE USER WAS RE-IMPLANTED ON (B)(6) 2025.

    Patient problems: Head Injury

    MAUDE report 21897090 ↗
  • MalfunctionApr 23, 2025Migration; Migration

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE USER WAS PROGRESSING WELL WITH THE IMPLANT BUT APPROXIMATELY 4 WEEKS AGO (ON (B)(6) 2025) A SCRATCHY SOUND AND ECHO WAS REPORTED. THE USER HAD REVISION SURGERY ON (B)(6) 2025 TO REINSERT THE IMPLANT. THE SURGEON OBSERVED ELECTRODES OUTSIDE OF THE COCHLEA. HE WAS ABLE TO REINSERT THE DEVICE WITHOUT ANY ISSUES.

    Patient problems: Discomfort

    MAUDE report 21887747 ↗
  • MalfunctionMar 28, 2025Impedance Problem; Impedance Problem

    ON (B)(6) 2025, THE USER WAS ADMITTED TO THE EMERGENCY ROOM AS HE BLEW HIS NOSE AND FORCED AIR UP UNDER HIS FLAP. THE USER IS 5 DAYS POST CI SURGERY. HE REPORTED PAIN IMMEDIATELY AFTER DOING THIS. AT INITIAL ACTIVATION ON (B)(6) 2025, THE USER REPORTED VARIOUS_ FEELING_ SENSATIONS, BUT LEFT THE SESSION BEING ABLE TO HEAR SOUND WITH THE DEVICE. THE USER WAS SEEN AGAIN ON (B)(6) 2025 FOR FOLLOW UP. INITIALLY HE STATED THAT IT FELT LIKE A "9 VOLT BATTERY" IN HIS HEAD WHICH IS WHY HE HAD NOT BEEN WEARING THE DEVICE. RE-MAPPING AS DONE, CHANNELS WERE DEACTIVATED AND HE REPORTED IMPROVEMENT. HE WILL BE SEEN AGAIN FOR FOLLOW-UP.

    Patient problems: Pain

    MAUDE report 21717395 ↗
  • MalfunctionMar 20, 2025Impedance Problem; Impedance Problem

    DURING SURGERY, THE SURGEON ENCOUNTERED UNEXPECTED FIBROSIS, AND IT WAS NOT POSSIBLE TO ACHIEVE FULL INSERTION (CHANNELS 11 AND 12 WERE EXTRA-COCHLEAR). THE SURGEON DECIDED TO REPLACE THE IMPLANT WITH A SHORTER ELECTRODE ARRAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21649998 ↗
  • MalfunctionMar 3, 2025Improper or Incorrect Procedure or Method; Impedance Problem; Improper or Incorrect Procedure or Method; Impedance Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. DUE TO THE IMPLANT POSITION THE COIL TOUCHED THE PROCESSOR. IN ORDER TO FIX THIS ISSUE, THE CLINIC PERFORMED A REPOSITION OF THE IMPLANT TWO WEEKS AGO. DURING/AFTER THE SURGERY THE ELECTRODE MIGRATED AND CHANNELS 11 AND 12 ARE NOW OUTSIDE OF THE COCHLEA AS CONFIRMED BY X-RAY). AN IMPLANT CHECK IS SCHEDULED FOR (B)(6) 2025.

    Patient problems: Discomfort; Failure of Implant

    MAUDE report 21504597 ↗
  • MalfunctionFeb 21, 2025Circuit Failure; Circuit Failure

    DURING A FITTING SESSION, IT WAS NOT POSSIBLE TO RUN THE IN-SITU MEASUREMENTS BECAUSE OF AN ERROR MESSAGE FROM THE MAESTRO SOFTWARE SAYING THAT THE DEVICE WAS NOT DETECTED. THE AUDIO PROCESSOR'S DL-COIL, THE MAX S-COIL AND THE COUPLING CHECK TOOL WERE TRIED. A DIFFERENT MAX AND A DIFFERENT LAPTOP WERE ALSO TRIED, BUT THE RESULT WAS STILL THE SAME. THE EXTERNAL PART WAS CHECKED, AND EVERYTHING WAS WORKING. THE USER DOES NOT HAVE HEARING SENSATION ON THE CONCERNED LEFT SIDE, AS ALSO CONFIRMED BY THE AUDIOMETRIC TEST. REIMPLANTATION IS CONSIDERED BUT NOT DATE HAS BEEN SCHEDULED AS YET.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21430524 ↗
  • InjuryFeb 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER HAS BEEN EXPLANTED DUE TO AN INFECTION IN THE MASTOID BOWL. THE SURGEON OPTED TO REMOVE THE INFECTION AND THE IMPLANT TO REDUCE POSSIBILITY OF FURTHER INFECTION, AND NOT TO REIMPLANT THE USER UNTIL THE INFECTION WAS TAKEN CARE OF. THE SURGEON REPORTED INFECTION WAS CONFIRMED BY CULTURES, WHEN MEETING WITH THE SURGEON HE WAS UNSURE WHAT THE RESULT WAS. THE INFECTION BEGAN A FEW WEEKS PRIOR TO THE EXPLANTATION AND THE PATIENT UNDERWENT ANTIBIOTICS TO TRY AND TREAT THE INFECTION.

    Patient problems: Mastoiditis

    MAUDE report 21382971 ↗
  • InjuryJan 15, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    WOUND DEHISCENCE RETRO AURICULAR STATUS POST COCHLEA-IMPLANTATION. EXPLANTATION DUE TO SKIN DEFECT OVER THE IMPLANT. DEVICE WAS EXTRUDED RETROAURICULLARY, CABLE WAS EXTRUDED IN THE CAVITY. A RETROAURICULAR INFECTION WAS CAUSED BY THE EXTRUSION. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Skin Tears

    MAUDE report 21157302 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge

Other Implant, Cochlear devices

All Implant, Cochlear devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page