This Med-El SYNCHRONY cochlear implant is a surgically placed hearing implant. The 46 reports mostly describe the implant shifting out of position or its electrode array sitting outside the cochlea, along with hearing not working after activation, ringing in the ear, and facial nerve stimulation. Reports mention 17 injuries and no deaths, with 29 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
51
Deaths reported
0
Injuries reported
21
Recalls naming it
0

Summary

The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 09008737395445), received between Oct 5, 2020 and Sep 28, 2026 — classified as 21 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Migration” (14 reports).

4 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 2 reportsJul 2026: 2 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration14
  • Device Appears to Trigger Rejection11
  • Impedance Problem11
  • Adverse Event Without Identified Device or Use Problem10
  • Device Handling Problem3
  • Insufficient Device Problem Information3
  • Output Problem3
  • Circuit Failure1

Patient problems

  • Failure of Implant16
  • No Clinical Signs, Symptoms or Conditions15
  • Skin Tears6
  • Tissue Breakdown6
  • Unspecified Infection4
  • Implant Pain3
  • Post Operative Wound Infection3
  • Bacterial Infection2

Recent adverse event reports

The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Impedance Problem

    The user came to the clinic because of low hearing performance. Re-implantation was done on (b)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911260 ↗
  • InjuryAug 20, 2026Device Appears to Trigger Rejection

    Swelling at the receiver/stimulator site of the left cochlear implant was observed one month post implantation surgery. Despite antibiotic therapy, the swelling did not improve, and the device eventually extruded and was therefore explanted.

    Patient problems: Skin Tears

    MAUDE report 50338442 ↗
  • InjuryAug 13, 2026Migration

    There was a migration of the electrode array out of the cochlea with 5 extra-cochlear channels. The user was reimplanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50234351 ↗
  • MalfunctionJul 17, 2026Migration

    IN SITU TESTING SHOW AFFECTED CHANNELS. REVISION SURGERY IS CONSIDERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25852083 ↗
  • MalfunctionJul 1, 2026Impedance Problem

    In-situ measurements show an increase in channels affected by high impedances since may 2023. Due to complex medical history, re-implantation is not recommended; The Audiologist is re-programming the right processor for sound awareness, disabling the HI electrodes and electrodes showing significant fluctuation.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25675789 ↗
  • InjuryJun 25, 2026Adverse Event Without Identified Device or Use Problem

    THE USER HAS BEEN SUFFERING FROM AN INFECTION AROUND THE IMPLANT'S STIMULATOR AND HAS BEEN HOSPITALIZED. IT IS PLANNED TO EXPLANT THE USER AND LEAVE THE ELECTRODE ARRAY INSIDE THE COCHLEA, SO THAT THE USER CAN BE REIMPLANTED ONCE THE INFECTION IS SORTED.

    Patient problems: Medical device site infection; Unspecified Infection

    MAUDE report 25619050 ↗
  • InjuryJun 2, 2026Adverse Event Without Identified Device or Use Problem

    THERE WAS NO HEARING IMPRESSION. THE USER HAS BEEN EXPLANTED.

    Patient problems: Wound Dehiscence

    MAUDE report 25360497 ↗
  • MalfunctionMay 12, 2026Migration

    NO HEARING SENSATION WITH THE IMPLANT AT FIRST FITTING. REPOSITIONING SURGERY IS SCHEDULED ON (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25146568 ↗
  • MalfunctionMar 23, 2026Migration

    USER HAS BEEN EXPERIENCING TINNITUS SINCE IMPLANTATION, BOTH WITH AND WITHOUT AP. USER IS SCHEDULED ON (B)(6) 2026 FOR RECEIVER/STIMULATOR ADJUSTMENT.

    Patient problems: Tinnitus

    MAUDE report 24660302 ↗
  • InjuryMar 3, 2026Device Appears to Trigger Rejection

    REPORTEDLY, THE MUSCLE FLAP WAS OPEN DUE TO A TOO STRONG MAGNET. THE DEVICE HAS EXTRUDED THROUGH THE SKIN. THE DEVICE WAS EXPLANTED.

    Patient problems: Skin Tears

    MAUDE report 24495922 ↗
  • MalfunctionFeb 23, 2026Migration

    AT THE INITIAL ACTIVATION THE USER COULD ONLY HEAR ON THE FIRST 4 ELECTRODES. AN X-RAY IS PLANNED AND THEN IT WILL BE DECIDED HOW TO PROCEED. HOWEVER, RE-POSITIONING IS BEING CONSIDERED, IF THE IMAGES SHOWS THAT THE ELECTRODE ARRAY IS OUTSIDE THE COCHLEA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413750 ↗
  • MalfunctionDec 19, 2025Migration; Migration

    THE USER HAS HAD POORER PERFORMANCE THAN EXPECTED SINCE ACTIVATION AND FACIAL STIMULATION. PATIENT HAS REPORTED ABNORMAL SOUND QUALITY MUMBLING WITH NO CLARITY OF SPEECH. ALSO REPORTS PINGING WHEN TOUCHING PINNA AND WITH JAW MOVEMENTS. RE-INSERTION WAS SUCCESSFULLY PERFORMED. THE FACIAL NERVE WAS DEHISCENT IN THE FACIAL RECESS. THE SURGEON PLACED A BIT OF TISSUE OVER THE FACIAL NERVE TO PROTECT IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Undesired Nerve Stimulation

    MAUDE report 23847855 ↗
  • MalfunctionNov 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THERE WAS A REVISION SURGERY 5 DAYS AFTER IMPLANTATION (DUE TO ELECTRODE DISLOCATION/INCORRECT POSITION) IN WHICH THE DEVICE WAS REMOVED TO A SALINE BATH AND RE-IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23448230 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER WAS EXPLANTED DUE TO A WOUND INFECTION WHICH STARTED A FEW WEEKS PRIOR. THE USER WILL BE REIMPLANTED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 23307478 ↗
  • MalfunctionSep 10, 2025Migration; Migration

    THE USER IS GETTING VERY LIMITED BENEFIT FROM HIS DEVICE. IMAGING HAS BEEN REQUESTED.

    Patient problems: Discomfort

    MAUDE report 23014404 ↗
  • MalfunctionAug 20, 2025Impedance Problem; Impedance Problem

    THE USER CAN NO LONGER HEAR WITH THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22847509 ↗
  • MalfunctionAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER IS HAVING SOME COMPLICATIONS AND LOTS OF PAIN AT THE SURGERY SITE. THE USER WILL UNDERGO EXPLANT SURGERY DUE TO LACK OF BENEFIT AND WILL NOT BE REIMPLANTED. SURGERY IS NOT YET SCHEDULED.

    Patient problems: Implant Pain

    MAUDE report 22791324 ↗
  • InjuryJul 7, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE ELECTRODE LEAD EXTRUDED INTO THE AUDITORY CANAL. THE USER HAS BEEN EXPLANTED.

    Patient problems: Skin Tears

    MAUDE report 22411085 ↗
  • MalfunctionMay 26, 2025Migration; Migration

    THE USER REPORTED A CHANGING/DETERIORATING HEARING IMPRESSION AND UNDERWENT A RE-POSITIONING SURGERY ON (B)(6) 2025 BECAUSE OF AN ELECTRODE MIGRATION. NOT ALL CHANNELS COULD BE REINSERTED DURING REVISION SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084681 ↗
  • InjuryMay 2, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    REPORTEDLY, THE ELECTRODE LEAD EXTRUDED THROUGH THE EAC. THE USER HAS BEEN RE-IMPLANTED.

    Patient problems: Skin Tears

    MAUDE report 21947813 ↗
  • MalfunctionMar 28, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    REPORTEDLY, THE USER COULD NOT HEAR WITH THE DEVICE AND A DEVICE DEFECT WAS ASSUMED.THE USER WAS REIMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21717260 ↗
  • MalfunctionMar 5, 2025Impedance Problem; Impedance Problem

    6 MONTHS AFTER IMPLANTATION, IN (B)(6) 2025, THE USER'S FATHER REPORTED THAT THE LEFT SIDE DEVICE WASN'T WORKING FOR 2 MONTHS AND THE USER WAS ONLY WEARING THE RIGHT SIDE. RE-IMPLANTATION IS CONSIDERED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21525570 ↗
  • InjuryFeb 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER HAS BEEN EXPLANTED DUE TO AN INFECTION OVER THE IMPLANT HOUSING.

    Patient problems: Post Operative Wound Infection; Unspecified Infection

    MAUDE report 21456887 ↗
  • InjuryFeb 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE USER WAS SENT TO THE EMERGENCY DEPARTMENT ON (B)(6) 2025 AND THEN FOLLOWED UP WITH A PHYSICIAN. THEY DID A CT AND GAVE A PRESSURE DRESSING AS WELL AS ORAL ANTIBIOTICS. THE USER CONTINUED TO EXPERIENCE ONGOING PAIN AND SWELLING; THUS, HE UNDERWENT A PROCEDURE ON (B)(6) 2025 TO EVACUATE INFECTION AND BLOOD FROM UNDER THE SKIN FLAP AND ASSESS IMPLANT INTEGRITY, WHICH WAS CONFIRMED AS FUNCTIONING NORMALLY, AND THE CLINIC DECIDED TO PRESERVE THE CURRENT DEVICE WITH PLANS FOR POTENTIAL USE IN 3-4 WEEKS IF HEALING PROGRESSES. THIS REPORT REFERS TO 9710014-2025-00123. FOR FURTHER INFORMATION PLEASE REFER TO THIS REPORT.

    Patient problems: Unspecified Infection

    MAUDE report 21371673 ↗
  • MalfunctionFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER WAS SENT TO THE EMERGENCY DEPARTMENT ON (B)(6) 2025 AND THEN FOLLOWED UP WITH A PHYSICIAN DU TO SWELLING AND PAIN. THEY DID A CT AND GAVE A PRESSURE DRESSING AS WELL AS ORAL ANTIBIOTICS. THE USER CONTINUED TO EXPERIENCE ONGOING PAIN AND SWELLING; THUS, HE UNDERWENT A PROCEDURE ON (B)(6) 2025 TO EVACUATE INFECTION AND BLOOD FROM UNDER THE SKIN FLAP AND ASSESS IMPLANT INTEGRITY, WHICH WAS CONFIRMED AS FUNCTIONING NORMALLY, AND THE CLINIC DECIDED TO PRESERVE THE CURRENT DEVICE WITH PLANS FOR POTENTIAL USE IN 3-4 WEEKS IF HEALING PROGRESSES.

    Patient problems: Unspecified Infection

    MAUDE report 21353397 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2356-2024 · Apr 18, 2024 · Open, Classified
    Mi1250 SYNCHRONY 2, Item Numbers: a) 39541, FLEXSOFT S-Vector; b) 39545, FLEX28 S-Vector; c) 39547, FLEX26 S-Vector; Cochlear Implant
  • Z-2136-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
  • Z-2137-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
  • Z-2138-2019 · Oct 16, 2018 · Terminated
    MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge

Other Implant, Cochlear devices