Mi1250 SYNCHRONY 2 FLEX26 S-Vector Magnet 39547Recalls & reported problems
Med-El Elektromedizinische Geräte Gesellschaft m.b.H. · Cochlear Implant · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a MED-EL Synchrony 2 cochlear implant, a hearing implant surgically placed inside the ear. The 32 reports mostly describe hearing performance dropping out, electrode channels shifting out of the inner ear, and implant-related skin infections and pain, with some users needing revision surgery or a new implant. There are no deaths and 6 injuries. One open recall, started April 2024, covers a small number of units that may crack inside their electronics and stop working.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 09008737395476), received between Sep 1, 2021 and Sep 15, 2026 — classified as 8 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Impedance Problem” (13 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (P000025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 18, 2024Class IIOpen, Classified
A small number of devices may have a higher probability of a crack within the internal electronics substrate which could result in a non-functional device.
Z-2356-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Impedance Problem13
- Adverse Event Without Identified Device or Use Problem9
- Migration8
- Insufficient Device Problem Information2
- Circuit Failure1
- Device Appears to Trigger Rejection1
- Electrical Shorting1
- Inadequacy of Device Shape and/or Size1
Patient problems
- Failure of Implant14
- No Clinical Signs, Symptoms or Conditions11
- Skin Infection3
- Head Injury2
- Implant Pain2
- Tissue Breakdown2
- Bacterial Infection1
- Granuloma1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
The device has been explanted because of wound infection. Due to wound dehiscence and very thin skin coverage, the implant body was visible. Therefore, revision of the implant bed was indicated. The electrode was left in situ.
Patient problems: Wound Dehiscence
MAUDE report 50769721 ↗InjuryAug 12, 2026Device Appears to Trigger Rejection
The user has been wearing her processor 24 hours a day for weeks on end and has developed an infection that has caused the implant to extrude. This report refers to 9710014-2026-00748. For further information please refer to this report.
Patient problems: Tissue Breakdown
MAUDE report 50218817 ↗InjuryJul 16, 2026Impedance Problem
IT HAS BEEN REPORTED A HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. CHANNELS 1 TO 4 SHOWED HIGH IMPEDANCE AND WERE THEREFORE DEACTIVATED ON (b)(6) 2026. EXPERT MEASUREMENTS IN CONTINUOUS MODE SHOWED FLUCTUATING IMPEDANCE VALUES ON CHANNELS 1, 6, 7, 11. REIMPLANTATION CONSIDERED BUT NO TIME FRAME HAS BEEN SET YET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25836950 ↗MalfunctionJun 26, 2026Migration
HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25632751 ↗MalfunctionApr 10, 2026Impedance Problem
HEARING PERFORMANCE WITH THE DEVICE HAS BEEN AFFECTED. THE USER HAS BEEN REIMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24842529 ↗MalfunctionMar 24, 2026Impedance Problem
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED AFTER FALLING OUT OF THE BED. THE CLINIC WILL SCHEDULE A DATE FOR REVISION SURGERY.
Patient problems: Head Injury
MAUDE report 24672161 ↗InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem
THE USER SUFFERED A SKIN INFECTION NEAR THE IMPLANT AREA AND SKIN FLAP NECROSIS. DEVICE WAS EXPLANTED.
Patient problems: Skin Infection
MAUDE report 24559402 ↗MalfunctionMar 6, 2026Migration
TWO ELECTRODE CHANNELS MIGRATED OUT OF COCHLEA AFTER IMPLANTATION. REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 AND THE USER WAS IMPLANTED WITH A NEW DEVICE. AT EXPLANTATION 6 ELECTRODE CHANNELS WERE FOUND EXTRA-COCHLEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24529271 ↗InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
EXPLANTATION ON (B)(6) 2025 DUE TO GRANULATIONS IN THE AREA OF THE IMPLANT BEARING THAT LED TO SWELLING AND PAIN ABOVE THE IMPLANT. REIMPLANTATION IS PLANNED BUT NOT SCHEDULED YET.
Patient problems: Granuloma
MAUDE report 23645883 ↗MalfunctionOct 24, 2025Migration; Migration
THE USER'S HEARING PERFORMANCE WITH THE DEVICE IS AFFECTED. THE USER HAS BEEN REIMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376914 ↗MalfunctionAug 21, 2025Impedance Problem; Impedance Problem
THE USER UNDERWENT RE-IMPLANTATION SURGERY. IN SITU MEASUREMENTS REPORTEDLY SHOWED ONE CHANNEL WITH HIGH IMPEDANCE; THEREFORE, A BACK UP DEVICE WAS IMPLANTED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22860234 ↗MalfunctionJul 9, 2025Migration; Migration
REPORTEDLY THERE WAS A DECREASE IN PERFORMANCE AND INCREASINGLY HIGHER IMPEDANCE VALUES. INITIALLY THE USER HAD BENEFIT, BUT HAS SEEN A GRADUAL DECREASE IN PERFORMANCE SOON AFTER THEIR FOLLOW UP FROM ACTIVATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22457548 ↗InjuryMay 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE USER HAD AN INFECTION AT THE IMPLANT AREA. USER WAS EXPLANTED ON THE LEFT SIDE.
Patient problems: Bacterial Infection; Unspecified Infection
MAUDE report 22112506 ↗MalfunctionMay 15, 2025Impedance Problem; Impedance Problem
THE SURGEON ELECTED TO IMPLANT A BACK-UP IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22024069 ↗MalfunctionMay 8, 2025Impedance Problem; Impedance Problem
IN SITU TESTING SHOWED AFFECTED CHANNELS.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21983219 ↗MalfunctionDec 23, 2024Impedance Problem; Impedance Problem
THE USER'S HEARING WITH THE DEVICE IS AFFECTED. SINCE APPROXIMATELY 2-3 MONTHS, HE HAS INTERMITTENCIES AND VOLUME DECREASES WHEN MOVING THE JAW. THE USER ALSO REPORTED PAIN IN THE EAR AND AROUND THE EAR WHEN LAYING ON THAT SIDE. PER CLINICAL SUPPORT, FURTHER TROUBLESHOOTING SHOULD BE APPLIED, LIKE WEARING A TIGHT HEADBAND OR RECOMMENDING THE USER TO PRESS/MASSAGE THE STIMULATOR AREA IN CASE INTERMITTENCY IS RECOGNIZED.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 20999221 ↗MalfunctionOct 14, 2024Inadequacy of Device Shape and/or Size; Adverse Event Without Identified Device or Use Problem; Inadequacy of Device Shape and/or Size; Adverse Event Without Identified Device or Use Problem
THE USER DID NOT EXPERIENCE SATISFACTORY SPEECH PERCEPTION SINCE IMPLANTATION. AS PER EXPLANT REPORT FORM INFORMATION, 1 OR 2 CHANNELS WERE FOUND OUTSIDE FROM THE COCHLEA AT EXPLANTATION. THE USER WAS RE-IMPLANTED WITH A LONGER ELECTRODE ARRAY.
Patient problems: Failure of Implant
MAUDE report 20437576 ↗MalfunctionOct 11, 2024Circuit Failure; Circuit Failure
AFTER CHANGING THE RONDO 3 AUDIO PROCESSOR, THE USER DID NOT HAVE ANY SOUND FROM THE DEVICE. THE USER HAS BEEN RE-IMPLANTED ON (B)(6) 2024.
Patient problems: Failure of Implant
MAUDE report 20425422 ↗MalfunctionSep 26, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE EXPLANTED DEVICE WAS RECEIVED FOR INVESTIGATION.
Patient problems: Implant Pain
MAUDE report 20307471 ↗MalfunctionSep 24, 2024Electrical Shorting; Electrical Shorting
AFTER IMPLANTING THE DEVICE IN THE INITIAL SURGERY ON (B)(6) 2024, INTRAOPERATIVE MEASUREMENTS SHOWED 7 SHORT-CIRCUITED CHANNELS. THERE WERE ALSO NO RESPONSES ON AUTOART. IT WAS DECIDED TO EXPLANT THE DEVICE AND USE THE BACK-UP INSTEAD. THE BACK-UP WAS SUCCESSFULLY IMPLANTED ON THE SAME DAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20290544 ↗MalfunctionSep 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE DEVICE HAS BEEN RECEIVED AT HQ FOR INVESTIGATION. AS PER INFORMATION ON THE DEVICE EXPLANT REPORT FORM THE DEVICE WAS EXPLANTED DUE TO INCREASING PAIN AND DIZZINESS, ELEVATED IMPEDANCES AND SHORT CIRCUITS. THE USER HAS BEEN RE-IMPLANTED WITH A NEW DEVICE.
Patient problems: Implant Pain
MAUDE report 20244911 ↗MalfunctionAug 23, 2024Impedance Problem; Lack of Effect; Impedance Problem; Lack of Effect
THE USER SUSTAINED A HEAD TRAUMA. THE USER INITIALLY REPORTED NO SIGNIFICANT CHANGE TO HEARING IMPRESSION AT THE TIME OF TESTING, WHICH WAS SHORTLY AFTER THE TRAUMA. IT WAS ADVISED TO MONITOR FOR CHANGES TO HEARING IMPRESSION AND ELECTRODE STATUS. AS PER FURTHER CLARIFICATION RECEIVED FROM THE FIELD, THERE WAS A DECREASE IN HEARING PERFORMANCE OVER TIME. RE-IMPLANTATION HAS BEEN DONE ON (B)(6) 2024.
Patient problems: Failure of Implant; Head Injury
MAUDE report 20046650 ↗MalfunctionJul 25, 2024Migration; Migration
THE USER'S HEARING PERFORMANCE WITH THE DEVICE WAS AFFECTED. IT WAS REPORTED THAT UPON ACTIVATION, THE USER EXHIBITED NO AUDITORY RESPONSE FOR 5-6 ELECTRODES. AN X-RAY REVEALED THAT AT LEAST SEVEN ELECTRODES WERE OUTSIDE THE COCHLEA. A RE-INSERTION SURGERY HAS BEEN SUCCESSFULLY PERFORMED.
Patient problems: Failure of Implant
MAUDE report 19826492 ↗MalfunctionJun 10, 2024Migration; Migration
IT WAS REPORTED THAT THE ELECTRODE MIGRATED FROM THE COCHLEA WHICH WAS CONFIRMED BY IMAGING LEADING O A REVISION SURGERY. DURING THE REPOSITIONING SURGERY, THE FIRM ELECTRODE LEAD WAS ENCASED BY TISSUE ON THE UPPER MASTOIDECTOMY. SMALL MOVEMENTS OF THE TISSUE, CAUSED BY THE SURGICAL INSTRUMENTS, SHOWED SHIFTING MOVEMENTS ON THE LOWER PART OF THE LEAD TOWARDS THE POSTERIOR TYMPANOTOMY. NO FIXATION CLIP COULD BE USED INITIALLY DUE TO THE ANATOMY. DURING THE EXPOSURE OF THE IMPLANT BED, THE OUTGOING LEAD WAS MOST LIKELY DAMAGED BY THE SURGEON. THEREFORE, THE USER WAS REIMPLANTED WITH A NEW DEVICE IN THE SAME SURGERY.
Patient problems: Failure of Implant
MAUDE report 19496222 ↗MalfunctionApr 10, 2024Migration; Migration
THE USER HAS NO AUDITORY IMPRESSION IN SINGLE-CHANNEL STIMULATION <40 QU IN THE INITIAL FITTING. THE USER HAS BEEN RE-IMPLANTED. ACCORDING TO THE EXPLANTATION REPORT THE DEVICE WAS DISLOCATED. IT IS ASSUMED THAT THE DISPLACEMENT OF THE IMPLANT HOUSING HAS CAUSED THE ELECTRODE TO MOVE OUT OF THE COCHLEAR IMPLANT.
Patient problems: Failure of Implant
MAUDE report 19078625 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000025. One submission often covers many models, so check the product and lot details in each record.
- Z-2137-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard
- Z-2138-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used toge
- Z-2136-2019 · Oct 16, 2018 · TerminatedMED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil000 CONCERTO PIN +FLEX28; b. Mi1000 CONCERTO +FLEX28; c. Mi100
Other Implant, Cochlear devices
- HiRes™ Ultra CI HiFocus™ MS Electrode CI-1600-04Advanced Bionics AG · Implant, Cochlear2,918 reports
- Cochlear® Nucleus® CI512 cochlear ImplantCochlear Limited · IMPLANT, COCHLEAR2,771 reports
- Cochlear® Nucleus® CI622 cochlear Implant with Slim Straight electrodeCochlear Limited · Implant, cochlear2,115 reports
- Cochlear® Nucleus® CI632 cochlear Implant with Slim Modiolar electrodeCochlear Limited · Implant, cochlear2,057 reports
- Nucleus Z60353Cochlear Limited · IMPLANT, COCHLEAR2 deaths1,815 reports
- Cochlear® Nucleus® CI522 cochlear Implant with Slim Straight electrodeCochlear Limited · IMPLANT, COCHLEAR1,475 reports
- Cochlear® Nucleus® CI612 cochlear ImplantCochlear Limited · Implant, cochlear1,150 reports
- Nucleus Z190574Cochlear Limited · IMPLANT, COCHLEAR981 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page