Skip to content

Baha 93338Recalls & reported problems

Cochlear Bone Anchored Solutions AB · BIA400 Implant 4mm w abutment 14mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Cochlear Baha bone-anchored hearing implant is an implant in the skull that holds a sound processor. All 128 reports are injuries, with no deaths; many describe skin problems and infection around the post that passes through the skin, and some describe the implant losing its bond with the bone and needing removal. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
128
Deaths reported
0
Injuries reported
128
Recalls naming it
0

Summary

The FDA's MAUDE database contains 128 adverse event reports that list this exact device (DI 09321502033442), received between Aug 1, 2017 and Apr 15, 2026 — classified as 128 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (48 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 2 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 0 reportsMay 2025: 7 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 6 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 4 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem48
  • Loss of Osseointegration35
  • Appropriate Device Problem Term/Code Not Available30
  • Adverse Event Without Identified Device or Use Problem7
  • Failure to Osseointegrate3
  • Therapeutic or Diagnostic Output Failure3
  • Detachment of Device or Device Component2
  • Loosening of Implant Not Related to Bone-Ingrowth2

Patient problems

  • Skin Infection28
  • Inadequate Osseointegration26
  • Impaired Healing17
  • Unspecified Infection13
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available8
  • Head Injury8
  • No Code Available8
  • Pain8

Recent adverse event reports

The 25 most recent of 128 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 15, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN OVERGROWTH AT THE ABUTMENT SITE. THE PATIENT SUBSEQUENTLY UNDERWENT REVISION SURGERY ON (B)(6) 2025 UNDER LOCAL ANESTHESIA TO EXCISE THE SKIN. HOWEVER, THE ISSUE COULD NOT BE RESOLVED SO THE PATIENT UNDERWENT ABUTMENT REMOVAL PROCEDURE AND A NEW BAHA CONNECT SYSTEM WAS IMPLANTED MORE SUPERIORLY IN AN AREA WITH LESS FATTY TISSUE UNDER GENERAL ANESTHESIA ON (B)(6) 2026.

    Patient problems: Impaired Healing

    MAUDE report 24880755 ↗
  • InjuryApr 15, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN OVERGROWTH AT THE ABUTMENT SITE. THE PATIENT SUBSEQUENTLY UNDERWENT REVISION SURGERY ON (B)(6) 2025 UNDER LOCAL ANESTHESIA TO EXCISE THE SKIN. HOWEVER, THE ISSUE COULD NOT BE RESOLVED SO THE PATIENT UNDERWENT ABUTMENT REMOVAL PROCEDURE AND A NEW BAHA CONNECT SYSTEM WAS IMPLANTED MORE SUPERIORLY IN AN AREA WITH LESS FATTY TISSUE UNDER GENERAL ANESTHESIA ON (B)(6) 2026.

    Patient problems: Impaired Healing

    MAUDE report 24880754 ↗
  • InjuryMar 11, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN ISSUES WITH GRANULATION TISSUE AT THE ABUTMENT SITE AND OTORRHEA. SUBSEQUENTLY, THE PATIENT UNDERWENT A SKIN REVISION SURGERY OF TREATMENT WITH SILVER NITRATE ON (B)(6) 2025.

    Patient problems: Fluid Discharge; Impaired Healing

    MAUDE report 24566293 ↗
  • InjuryMar 11, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN ISSUES WITH GRANULATION TISSUE AT THE ABUTMENT SITE AND OTORRHEA. SUBSEQUENTLY, THE PATIENT UNDERWENT A SKIN REVISION SURGERY OF TREATMENT WITH SILVER NITRATE ON (B)(6) 2025.

    Patient problems: Fluid Discharge; Impaired Healing

    MAUDE report 24566292 ↗
  • InjuryMar 9, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN ISSUES AROUND ABUTMENT SITE (SPECIFIC DATE NOT REPORTED). THE PATIENT UNDERWENT REVISION SURGERY TO DEBRIDE THE AFFECTED SITE ON JUNE 10, 2025. DURING THE SAME PROCEDURE ON (B)(6) 2025, THE PATIENT ALSO RECEIVED A STEROID INJECTION (DURATION NOT REPORTED). LATER, THE PATIENT DEVELOPED BLEEDING AT ABUTMENT SITE. THE IMPLANTED DEVICE REMAINS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24542350 ↗
  • InjuryMar 9, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN ISSUES AROUND ABUTMENT SITE (SPECIFIC DATE NOT REPORTED). THE PATIENT UNDERWENT REVISION SURGERY TO DEBRIDE THE AFFECTED SITE ON (B)(6) 2025. DURING THE SAME PROCEDURE ON (B)(6) 2025, THE PATIENT ALSO RECEIVED A STEROID INJECTION (DURATION NOT REPORTED). LATER, THE PATIENT DEVELOPED BLEEDING AT ABUTMENT SITE. THE IMPLANTED DEVICE REMAINS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24542349 ↗
  • InjuryFeb 26, 2026Loss of Osseointegration

    PER THE CLINIC, THE PATIENT EXPERIENCED A LOSS OF OSSEOINTEGRATION RESULTING IN FIXTURE AND ABUTMENT LOSS (SPECIFIC DATE NOT REPORTED). A REVISION SURGERY IS PLANNED BUT HAD NOT TAKEN PLACE AT THE TIME OF THIS REPORT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24451521 ↗
  • InjuryFeb 25, 2026Loss of Osseointegration

    PER THE CLINIC, THE PATIENT EXPERIENCED A LOSS OF OSSEOINTEGRATION RESULTING IN FIXTURE AND ABUTMENT LOSS (SPECIFIC DATE NOT REPORTED). A REVISION SURGERY IS PLANNED BUT HAD NOT TAKEN PLACE AT THE TIME OF THIS REPORT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24449951 ↗
  • InjuryFeb 6, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN OVERGROWTH AT THE IMPLANT SITE, REQUIRING USE OF A LONGER ABUTMENT. THE ORIGINAL ABUTMENT WAS REPORTEDLY LOST FOLLOWING A PATIENT FALL, DURING WHICH THE ABUTMENT DETACHED AND THE FIXATION SCREW FRACTURED. ON (B)(6) 2025, THE PATIENT UNDERWENT REVISION SURGERY UNDER GENERAL ANESTHESIA. DURING THE PROCEDURE, IT WAS OBSERVED THAT PART OF THE ABUTMENT THREAD REMAINED STUCK WITHIN THE IMPLANT SCREW, AND THE SURGEON WAS UNABLE TO REMOVE THE REMAINING ABUTMENT THREAD OUT OF IMPLANT. SUBSEQUENTLY, ON (B)(6) 2025, A FURTHER SURGICAL PROCEDURE WAS PERFORMED UNDER GENERAL ANESTHESIA, DURING WHICH THE ABUTMENT WAS REPLACED. THE IMPLANTED DEVICE REMAINS.

    Patient problems: Impaired Healing

    MAUDE report 24277247 ↗
  • InjuryFeb 6, 2026Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED SKIN OVERGROWTH AT THE IMPLANT SITE, REQUIRING USE OF A LONGER ABUTMENT. THE ORIGINAL ABUTMENT WAS REPORTEDLY LOST FOLLOWING A PATIENT FALL, DURING WHICH THE ABUTMENT DETACHED AND THE FIXATION SCREW FRACTURED. ON (B)(6) 2025, THE PATIENT UNDERWENT REVISION SURGERY UNDER GENERAL ANESTHESIA. DURING THE PROCEDURE, IT WAS OBSERVED THAT PART OF THE ABUTMENT THREAD REMAINED STUCK WITHIN THE IMPLANT SCREW, AND THE SURGEON WAS UNABLE TO REMOVE THE REMAINING ABUTMENT THREAD OUT OF IMPLANT. SUBSEQUENTLY, ON (B)(6) 2025, A FURTHER SURGICAL PROCEDURE WAS PERFORMED UNDER GENERAL ANESTHESIA, DURING WHICH THE ABUTMENT WAS REPLACED. THE IMPLANTED DEVICE REMAINS.

    Patient problems: Impaired Healing

    MAUDE report 24277243 ↗
  • InjuryJan 21, 2026Detachment of Device or Device Component; Therapeutic or Diagnostic Output Failure

    PER THE CLINIC, THE PATIENT EXPERIENCED A PERFORMANCE DECREMENT SUBSEQUENT TO SUSTAINING AN IMPACT TO THE ABUTMENT SIDE OF THE HEAD. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED UNDER A GENERAL ANAESTHETIC ON (B)(6) 2025. THE PATIENT WAS REIMPLANTED WITH A TRANSCUTANEOUS OSIA IMPLANT SYSTEM AT A NEW SITE.

    Patient problems: Head Injury

    MAUDE report 24127878 ↗
  • InjuryJan 21, 2026Detachment of Device or Device Component; Therapeutic or Diagnostic Output Failure

    PER THE CLINIC, THE PATIENT EXPERIENCED A PERFORMANCE DECREMENT SUBSEQUENT TO SUSTAINING AN IMPACT TO THE ABUTMENT SIDE OF THE HEAD. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED UNDER A GENERAL ANAESTHETIC ON (B)(6) 2025. THE PATIENT WAS REIMPLANTED WITH A TRANSCUTANEOUS OSIA IMPLANT SYSTEM AT A NEW SITE.

    Patient problems: Head Injury

    MAUDE report 24127874 ↗
  • InjuryJan 14, 2026Loss of Osseointegration

    PER THE CLINIC, THE PATIENT EXPERIENCED AN INFECTION AT THE IMPLANT SITE AND WAS SUBSEQUENTLY TREATED WITH ORAL ANTIBIOTICS (SPECIFIC DATE AND DURATION NOT REPORTED). THE EVENT RESULTED IN FIXTURE LOSS AND LOSS OF OSSEOINTEGRATION WITH MINOR BLEEDING. THE PATIENT DOES NOT WISH TO UNDERGO REIMPLANTATION AND WILL BECOME A NON-USER.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Inadequate Osseointegration; Skin Infection

    MAUDE report 24065295 ↗
  • InjuryJan 14, 2026Loss of Osseointegration

    PER THE CLINIC, THE PATIENT EXPERIENCED AN INFECTION AT THE IMPLANT SITE AND WAS SUBSEQUENTLY TREATED WITH ORAL ANTIBIOTICS (SPECIFIC DATE AND DURATION NOT REPORTED). THE EVENT RESULTED IN FIXTURE LOSS AND LOSS OF OSSEOINTEGRATION WITH MINOR BLEEDING. THE PATIENT DOES NOT WISH TO UNDERGO REIMPLANTATION AND WILL BECOME A NON-USER.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Inadequate Osseointegration; Skin Infection

    MAUDE report 24065293 ↗
  • InjuryDec 11, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT WAS TREATED WITH ORAL ANTIBIOTICS FOR THE INFECTION (SPECIFIC DATE AND DURATION NOT REPORTED). THE PATIENT WAS THEN PLACED UNDER GENERAL ANESTHESIA ON (B)(6) 2025 TO REMOVE THE ABUTMENT. THE INTERNAL FIXTURE REMAINS.

    Patient problems: Skin Infection

    MAUDE report 23786099 ↗
  • InjuryDec 11, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED RECURRENT INFECTIONS AT THE ABUTMENT SITE (SPECIFIC DATE NOT REPORTED). ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.

    Patient problems: Skin Infection

    MAUDE report 23786097 ↗
  • InjuryNov 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED A DISCOMFORT AND SKIN INFECTION AT THE ABUTMENT SITE. THE PATIENT WAS TREATED WITH SILVER NITRATE, ORAL AND TOPICAL ANTIBIOTICS, AND UNDERWENT A SKIN REVISION SURGERY TO RESOLVE THE INFECTION. SUBSEQUENTLY, THE ABUTMENT WAS REMOVED UNDER LOCAL ANESTHESIA ON (B)(6) 2025. THERE ARE PLANS TO REPLACE THE PATIENT WITH A NEW ABUTMENT HOWEVER, THIS HAS NOT OCCURRED AS OF THE DATE OF THIS REPORT.

    Patient problems: Discomfort; Skin Infection

    MAUDE report 23676989 ↗
  • InjuryNov 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED A DISCOMFORT AND SKIN INFECTION AT THE ABUTMENT SITE. THE PATIENT WAS TREATED WITH SILVER NITRATE, ORAL AND TOPICAL ANTIBIOTICS, AND UNDERWENT A SKIN REVISION SURGERY TO RESOLVE THE INFECTION. SUBSEQUENTLY, THE ABUTMENT WAS REMOVED UNDER LOCAL ANESTHESIA ON (B)(6) 2025. THERE ARE PLANS TO REPLACE THE PATIENT WITH A NEW ABUTMENT HOWEVER, THIS HAS NOT OCCURRED AS OF THE DATE OF THIS REPORT.

    Patient problems: Discomfort; Skin Infection

    MAUDE report 23676987 ↗
  • InjuryNov 19, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED CELLULITIS AT THE IMPLANT SITE AND WAS TREATED WITH ORAL AND TOPICAL ANTIBIOTICS AND TOPICAL STEROID OINTMENT (SPECIFIC DATE AND DURATION NOT REPORTED). ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.

    Patient problems: Cellulitis

    MAUDE report 23591049 ↗
  • InjuryNov 19, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED CELLULITIS AT THE IMPLANT SITE AND WAS TREATED WITH ORAL AND TOPICAL ANTIBIOTICS AND TOPICAL STEROID OINTMENT (SPECIFIC DATE AND DURATION NOT REPORTED). ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.

    Patient problems: Cellulitis

    MAUDE report 23591047 ↗
  • InjuryNov 17, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED RECURRENT FLUID DRAINAGE, BLEEDING AND GRANULATION TISSUE FORMATION AT THE ABUTMENT SITE. SUBSEQUENTLY, THE ABUTMENT WAS REMOVED UNDER GENERAL ANESTHESIA ON (B)(6) 2025.

    Patient problems: Fluid Discharge; Hemorrhage/Blood Loss/Bleeding; Impaired Healing

    MAUDE report 23566726 ↗
  • InjuryNov 17, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED RECURRENT FLUID DRAINAGE, BLEEDING AND GRANULATION TISSUE FORMATION AT THE ABUTMENT SITE. SUBSEQUENTLY, THE ABUTMENT WAS REMOVED UNDER GENERAL ANESTHESIA ON (B)(6) 2025.

    Patient problems: Fluid Discharge; Hemorrhage/Blood Loss/Bleeding; Impaired Healing

    MAUDE report 23566725 ↗
  • InjuryAug 29, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    PER THE CLINIC, THE PATIENT EXPERIENCED A LOSS OF OSSEOINTEGRATION SUBSEQUENT TO SUSTAINING A HEAD TRAUMA. REIMPLANTATION IS PLANNED BUT HAD NOT TAKEN PLACE AT THE TIME OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.

    Patient problems: Head Injury; Inadequate Osseointegration

    MAUDE report 22920314 ↗
  • InjuryAug 29, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    PER THE CLINIC, THE PATIENT EXPERIENCED A LOSS OF OSSEOINTEGRATION SUBSEQUENT TO SUSTAINING A HEAD TRAUMA. RE-IMPLANTATION IS PLANNED BUT HAD NOT TAKEN PLACE AT THE TIME OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.

    Patient problems: Head Injury; Inadequate Osseointegration

    MAUDE report 22920312 ↗
  • InjuryJun 12, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    PER THE CLINIC, THE PATIENT EXPERIENCED AN INFECTION AT THE ABUTMENT SITE AND SUBSEQUENTLY WAS TREATED WITH ORAL AND IV ANTIBIOTICS (SPECIFIC DATE AND DURATION NOT REPORTED). HOWEVER, THE ISSUE COULD NOT BE RESOLVED AND THE DEVICE WAS EXPLANTED UNDER GENERAL ANESTHESIA ON (B)(6) 2025. THERE ARE NO PLANS TO RE-IMPLANT THE PATIENT WITH A NEW DEVICE AS OF THE DATE OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.

    Patient problems: Skin Infection

    MAUDE report 22191815 ↗

Other Hearing aid, bone conduction, implanted devices

All Hearing aid, bone conduction, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page