Cochlear® Nucleus® CI624 cochlear implant with Slim 20 electrodeRecalls & reported problems
Cochlear Limited · Class 3 · Implant
All models of Cochlear Nucleus Implant (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Cochlear Nucleus CI624 cochlear implant is a surgically placed hearing implant with a 20-electrode array. The 67 reports mostly describe problems around surgery and healing: infections at the incision site, skin breakdown, swelling and fluid buildup under the skin, wound opening, pain, and in some cases the electrode array pushing out of the cochlea or sitting outside it, leading to repeat surgery to remove or replace the device. Reports describe 66 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 69
- Deaths reported
- 0
- Injuries reported
- 68
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 69 adverse event reports that list this exact device (DI 09321502046329), received between Oct 19, 2020 and Sep 24, 2026 — classified as 68 injuries, 1 malfunction.
The most frequently reported specific device problem is “Expulsion” (13 reports).
3 other recalls cover devices cleared under the same FDA submission (P970051); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available33
- Expulsion13
- Adverse Event Without Identified Device or Use Problem4
- Malposition of Device4
- Migration4
- Circuit Failure2
- Failure to Deliver Shock/Stimulation2
- Insufficient Device Problem Information2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available20
- Unspecified Infection15
- Tissue Breakdown10
- Wound Dehiscence10
- Pain7
- Post Operative Wound Infection5
- No Clinical Signs, Symptoms or Conditions4
- Swelling/ Edema4
Recent adverse event reports
The 25 most recent of 69 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Appropriate Device Problem Term/Code Not Available
Per the clinic, the patient experienced pain/non auditory sensations with the device. Subsequently, the device was explanted on (b)(6)2026, and the patient was reimplanted with another cochlear device during the same surgery.
Patient problems: Pain
MAUDE report 50878481 ↗InjurySep 24, 2026Appropriate Device Problem Term/Code Not Available
Per the clinic, the patient experienced pain/non auditory sensations with the device. Subsequently, the device was explanted on (b)(6) 2026, and the patient was reimplanted with another cochlear device during the same surgery.
Patient problems: Pain
MAUDE report 50878480 ↗InjuryApr 20, 2026Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT WAS PLACED UNDER GENERAL ANESTHESIA AND UNDERWENT REVISION SURGERY ON (B)(6) 2025 (SPECIFIC DATE NOT REPORTED); IN ORDER TO DRAIN SEROMA AT INCISION SITE.
Patient problems: Pain; Swelling/ Edema
MAUDE report 24926643 ↗InjuryMar 18, 2026Therapeutic or Diagnostic Output Failure
PER THE CLINIC, THE PATIENT EXPERIENCED A PERFORMANCE DECREMENT WITH THE DEVICE. REPROGRAMMING ATTEMPTS WERE MADE; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. DURING FOLLOW-UP EXAMINATION, THE PATIENT UNDERWENT A CT SCAN WHICH REVEALED 5-6 ELECTRODES PLACED EXTRA-COCHLEARLY WHICH LED TO THE DECISION TO EXPLANT THE DEVICE. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED ON (B)(6) 2026, AND THE PATIENT WAS RE-IMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Extraskeletal ossification; Scar Tissue
MAUDE report 24635714 ↗InjuryMar 18, 2026Therapeutic or Diagnostic Output Failure
PER THE CLINIC, THE PATIENT EXPERIENCED A PERFORMANCE DECREMENT WITH THE DEVICE. REPROGRAMMING ATTEMPTS WERE MADE; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. DURING FOLLOW-UP EXAMINATION, THE PATIENT UNDERWENT A CT SCAN WHICH REVEALED 5-6 ELECTRODES PLACED EXTRA-COCHLEARLY WHICH LED TO THE DECISION TO EXPLANT THE DEVICE. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED ON (B)(6) 2026, AND THE PATIENT WAS RE-IMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Extraskeletal ossification
MAUDE report 24635713 ↗InjuryFeb 4, 2026Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT EXPERIENCED PAIN AND SWELLING AT THE MAGNET SITE, LEADING TO NON-USE OF THE SOUND PROCESSOR IN (B)(6) 2025 (SPECIFIC DATES NOT REPORTED). SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION SURGERY IN (B)(6) 2025 (SPECIFIC DATE NOT REPORTED), TO DRAIN A SEROMA AT THE INCISION SITE. THE PATIENT WAS ALSO TREATED WITH MULTIPLE ROUNDS OF ANTIBIOTICS (SPECIFIC DATES, DURATION AND TYPES NOT REPORTED). ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.
Patient problems: Pain; Seroma; Swelling/ Edema
MAUDE report 24255703 ↗InjuryNov 4, 2025Expulsion; Expulsion
PER THE CLINIC, THE PATIENT DEVELOPED AN INFECTION, EXPERIENCED SKIN BREAKDOWN AT IMPLANT SITE AND EXTRUSION OF THE ELECTRODE ARRAY. SUBSEQUENTLY, THE PATIENT WAS TREATED WITH MULTIPLE ROUNDS OF ORAL ANTIBIOTICS (SPECIFIC DATE AND DURATION NOT REPORTED); HOWEVER, THE ISSUE DID NOT RESOLVE. THE DEVICE WAS EXPLANTED ON (B)(6) 2025. THERE ARE NO PLANS TO REIMPLANT THE PATIENT WITH A NEW DEVICE AS OF THE DATE OF THIS REPORT.
Patient problems: Tissue Breakdown; Unspecified Infection
MAUDE report 23472994 ↗InjuryNov 4, 2025Expulsion; Expulsion
PER THE CLINIC, THE PATIENT DEVELOPED AN INFECTION, EXPERIENCED SKIN BREAKDOWN AT IMPLANT SITE AND EXTRUSION OF THE ELECTRODE ARRAY. SUBSEQUENTLY, THE PATIENT WAS TREATED WITH MULTIPLE ROUNDS OF ORAL ANTIBIOTICS (SPECIFIC DATE AND DURATION NOT REPORTED); HOWEVER, THE ISSUE DID NOT RESOLVE. THE DEVICE WAS EXPLANTED ON (B)(6) 2025. THERE ARE NO PLANS TO REIMPLANT THE PATIENT WITH A NEW DEVICE AS OF THE DATE OF THIS REPORT.
Patient problems: Medical device site infection; Tissue Breakdown
MAUDE report 23472992 ↗InjurySep 10, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT EXPERIENCED SWELLING, POOR WOUND HEALING, DEVELOPED A WOUND DEHISCENCE AND INFECTION AT THE INCISION SITE. SUBSEQUENTLY, THE PATIENT WAS TREATED WITH ORAL ANTIBIOTICS FOR A DURATION OF 10 DAYS (SPECIFIC DATE NOT REPORTED). THE IMPLANTED DEVICE REMAINS.
Patient problems: Impaired Healing; Medical device site infection; Swelling/ Edema; Wound Dehiscence
MAUDE report 23012604 ↗InjurySep 10, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT EXPERIENCED SWELLING, POOR WOUND HEALING, DEVELOPED A WOUND DEHISCENCE AND INFECTION AT THE INCISION SITE. SUBSEQUENTLY, THE PATIENT WAS TREATED WITH ORAL ANTIBIOTICS FOR A DURATION OF 10 DAYS (SPECIFIC DATE NOT REPORTED). THE IMPLANTED DEVICE REMAINS.
Patient problems: Impaired Healing; Medical device site infection; Swelling/ Edema; Wound Dehiscence
MAUDE report 23012602 ↗InjuryJun 24, 2025Expulsion; Expulsion
PER THE CLINIC, THE PATIENT EXPERIENCED AN EXTRUSION OF THE ELECTRODE INTO THE EAR CANAL. IT WAS ALSO REPORTED THAT THERE WAS CHRONIC TYMPANIC MEMBRANE PERFORATION WITH INTERMITTENT OTORRHEA AND BREAKDOWN OF CANAL CLOSURE. THE DEVICE WAS EXPLANTED ON (B)(6) 2025 AND RE-IMPLANTATION IS PLANNED BUT HAD NOT TAKEN PLACE AT THE TIME OF THIS REPORT.
Patient problems: Fluid Discharge; Perforation; Tissue Breakdown
MAUDE report 22313049 ↗InjuryJun 24, 2025Expulsion; Expulsion
PER THE CLINIC, THE PATIENT EXPERIENCED AN EXTRUSION OF THE ELECTRODE INTO THE EAR CANAL. IT WAS ALSO REPORTED THAT THERE WAS CHRONIC TYMPANIC MEMBRANE PERFORATION WITH INTERMITTENT OTORRHEA AND BREAKDOWN OF CANAL CLOSURE. THE DEVICE WAS EXPLANTED ON (B)(6) 2025 AND RE-IMPLANTATION IS PLANNED BUT HAD NOT TAKEN PLACE AT THE TIME OF THIS REPORT.
Patient problems: Fluid Discharge; Perforation; Tissue Breakdown
MAUDE report 22313047 ↗InjuryFeb 9, 2025Circuit Failure; Circuit Failure
PER THE CLINIC, THE DEVICE WAS EXPLANTED ON (B)(6) 2025 AND THE PATIENT WAS REIMPLANTED WITH A NEW COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21341316 ↗InjuryFeb 9, 2025Circuit Failure; Circuit Failure
PER THE CLINIC, THE PATIENT EXPERIENCED OPEN CIRCUITS WITH THE ELECTRODES (SPECIFIC DATE NOT REPORTED). THERE ARE PLANS TO EXPLANT THE DEVICE HOWEVER, THIS HAS NOT OCCURRED AS OF THE DATE OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21341312 ↗InjuryNov 12, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
PER THE CLINIC, THE PATIENT EXPERIENCED LOSS OF CONNECTION TO THE INTERNAL DEVICE. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED (DATE NOT REPORTED), AND IT IS UNKNOWN IF THERE ARE PLANS TO REIMPLANT THE PATIENT AS OF THE DATE OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20670255 ↗InjuryNov 12, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
PER THE CLINIC, THE PATIENT EXPERIENCED LOSS OF CONNECTION TO THE INTERNAL DEVICE. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED (DATE NOT REPORTED), AND IT IS UNKNOWN IF THERE ARE PLANS TO RE-IMPLANT THE PATIENT AS OF THE DATE OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20670230 ↗InjurySep 18, 2024Positioning Failure; Positioning Failure
PER THE CLINIC, THE DEVICE WAS EXPLANTED ON (B)(6) 2024 DUE TO IMPROPER PLACEMENT OF THE ELECTRODE ARRAY AND POOR PERFORMANCE WITH THE DEVICE. THE PATIENT WAS REIMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20243339 ↗InjurySep 18, 2024Positioning Failure; Positioning Failure
PER THE CLINIC, THE DEVICE WAS EXPLANTED ON (B)(6) 2024 DUE TO IMPROPER PLACEMENT OF THE ELECTRODE ARRAY AND POOR PERFORMANCE WITH THE DEVICE. THE PATIENT WAS REIMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20243337 ↗MalfunctionApr 24, 2024Migration; Migration
PER THE CLINIC, THE PATIENT EXPERIENCED A MIGRATION OF THE ELECTRODE ARRAY AND TINGLING SENSATION WITH STIMULATION SUBSEQUENT TO SUSTAINING A HEAD INJURY. SUBSEQUENTLY, THE DEVICE WAS EXPLANTED ON (B)(6) 2024 UNDER GENERAL ANESTHESIA AND THE PATIENT WAS REIMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Head Injury
MAUDE report 19170201 ↗InjuryApr 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PER THE CLINIC, THE DEVICE WAS EXPLANTED ON (B)(6) 2024 IN ORDER TO PERFORM MRI PROCEDURE. REIMPLANTATION IS PLANNED BUT HAS NOT TAKEN PLACE AT THE TIME OF THIS REPORT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19150757 ↗InjuryApr 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PER THE CLINIC, THE DEVICE WAS EXPLANTED ON (B)(6) 2024 IN ORDER TO PERFORM MRI PROCEDURE. REIMPLANTATION IS PLANNED BUT HAS NOT TAKEN PLACE AT THE TIME OF THIS REPORT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19150749 ↗InjuryApr 5, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT DEVELOPED AN INFECTION AT THE IMPLANT SITE AND SUBSEQUENTLY WAS TREATED WITH ORAL ANTIBIOTICS, HOWEVER, THE ISSUE DID NOT RESOLVE. THE DEVICE WAS EXPLANTED ON (B)(6) 2024. THE PATIENT WAS REIMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Unspecified Infection
MAUDE report 19048499 ↗InjuryApr 5, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT DEVELOPED AN INFECTION AT THE IMPLANT SITE AND SUBSEQUENTLY WAS TREATED WITH ORAL ANTIBIOTICS, HOWEVER, THE ISSUE DID NOT RESOLVE. THE DEVICE WAS EXPLANTED ON (B)(6) 2024. THE PATIENT WAS RE-IMPLANTED WITH ANOTHER COCHLEAR DEVICE DURING THE SAME SURGERY.
Patient problems: Unspecified Infection
MAUDE report 19048494 ↗InjuryJan 15, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
PER THE CLINIC, THE PATIENT WAS HOSPITALIZED IN (B)(6) 2023 (SPECIFIC DATE NOT REPORTED) FOR VERTIGO. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION SURGERY (SPECIFIC DATE NOT REPORTED) FOR UNSPECIFIED REASONS. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE AS OF THE DATE OF THIS REPORT.
Patient problems: Vertigo
MAUDE report 18509981 ↗InjuryJan 11, 2024Migration; Migration
PER THE CLINIC, THE PATIENT EXPERIENCED A MIGRATION OF ELECTRODE ARRAY. THE PATIENT UNDERWENT REVISION SURGERY UNDER GENERAL ANESTHESIA ON (B)(6) 2022 IN ORDER TO REINSERT THE ELECTRODE ARRAY. THE IMPLANTED DEVICE REMAINS.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18490480 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P970051. One submission often covers many models, so check the product and lot details in each record.
- Z-2155-2015 · Jun 5, 2015 · TerminatedCochlear Nucleus Sterile Silicone Template Product Usage: The Cochlear Nucleus Sterile Silicone Template is used in the sterile field to check the size of the periosteal pocket, t
- Z-0003-2012 · Sep 16, 2011 · TerminatedCochlear Nucleus CI512 Cochlear Implant, REF Z209051, New with Nucleus 5, Sterile EO, Made in Australia, Cochlear Ltd 14 Mars Road, Lane Cove 2066, Australia. The cochlear implant
- Z-0119-2010 · Sep 30, 2009 · TerminatedNucleus Non-Magnetic Plug for removable magnet implants, REF Z50100, STERILE EO, Cochlear Limited.
Other Implant, cochlear devices
- HiRes™ Ultra CI HiFocus™ MS Electrode CI-1600-04Advanced Bionics AG · Implant, Cochlear2,918 reports
- Cochlear® Nucleus® CI512 cochlear ImplantCochlear Limited · IMPLANT, COCHLEAR2,771 reports
- Cochlear® Nucleus® CI622 cochlear Implant with Slim Straight electrodeCochlear Limited · Implant, cochlear2,115 reports
- Cochlear® Nucleus® CI632 cochlear Implant with Slim Modiolar electrodeCochlear Limited · Implant, cochlear2,057 reports
- Nucleus Z60353Cochlear Limited · IMPLANT, COCHLEAR2 deaths1,815 reports
- Cochlear® Nucleus® CI522 cochlear Implant with Slim Straight electrodeCochlear Limited · IMPLANT, COCHLEAR1,475 reports
- Cochlear® Nucleus® CI612 cochlear ImplantCochlear Limited · Implant, cochlear1,150 reports
- Nucleus Z190574Cochlear Limited · IMPLANT, COCHLEAR981 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.