Skip to content

Kosmos OTS iOS Software Package P007639Recalls & reported problems

Echonous, Inc. · Kosmos OTS iOS Software Package · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 09353278000717), received between May 30, 2025 and May 30, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 9, 2024Open, Classified

    Ultrasound system has a bug in affected iOS software, which after DICOM export, can cause inaccuracy in postprocessed measurements done in 3rd party software, such as PACS or other post-processing software, which may lead to inaccurate diagnoses or treatment.

    Z-2325-2024 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 30, 2025Computer Software Problem; Computer Software Problem

    ON (B)(6) 2024, ECHONOUS DETECTED A SOFTWARE ANOMALY IN THE MODULE RESPONSIBLE FOR CREATING DICOM FILES FOR EXPORT OR ARCHIVE. ALTHOUGH THE IMAGES AND CLIPS ARE BEING STORED CORRECTLY IN THE DICOM FILES CREATED BY OUR IOS SOFTWARE, THE PIXEL INCREMENT VALUES PUT IN THE METADATA ARE INCORRECT. AS A RESULT, THE MEASUREMENTS PERFORMED ON ARCHIVED OR EXPORTED DICOM IMAGES/CLIPS WILL BE INCORRECT (10% TO 15% WITH B-MODE, ~ 2% IN M-MODE DISTANCE, CW/PW >10X). THE DEFECT IS ONLY PRESENT WHEN PERFORMING MEASUREMENTS ON EXPORTED/ARCHIVED FILES USING THIRD-PARTY DICOM SOFTWARE. MEASUREMENTS PERFORMED WITHIN THE KOSMOS APP ARE UNAFFECTED BY THIS DEFECT. ADDITIONALLY, THIS SOFTWARE BUG IS ONLY PRESENT IN THE IOS VERSION OF THE KOSMOS APP. KOSMOS AND THE KOSMOS ANDROID APP REMAIN UNAFFECTED BY THIS DEFECT.

    Patient problems: Insufficient Information

    MAUDE report 22118943 ↗

Other System, Imaging, Pulsed Echo, Ultrasonic devices

All System, Imaging, Pulsed Echo, Ultrasonic devices →

Sources and updates