Temperature Probe 900MR869Recalls & reported problems

Fisher & Paykel Healthcare Limited · Sensor for MR850 · Class 2

Quick summary

This Fisher & Paykel temperature probe is a sensor used with a breathing-gas humidifier. Reports mostly describe the probe coming loose or falling out of the breathing-circuit port during setup before use, without patient harm. There are 29 reports naming its identifier, mentioning 1 death and 1 injury. No recalls name this device, with reports falling to 5 in the last 12 months from 23 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 09420012401311), received between Apr 23, 2021 and Jan 19, 2026 — classified as 1 death, 1 injury, 27 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 5 reportsJan 2025: 4 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated26
  • Connection Problem1
  • Loose or Intermittent Connection1
  • Mechanical Problem1
  • Moisture or Humidity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Low Oxygen Saturation1
  • Obstruction/Occlusion1
  • Pneumonia1
  • Respiratory Failure1
  • Stroke/CVA1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2026Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE EASILY DISLODGED FROM A TEMPERATURE PROBE PORT OF AN RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24110467 ↗
  • MalfunctionDec 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE EASILY DISLODGED FROM A TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23752231 ↗
  • MalfunctionDec 2, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT 97 900MR869 TEMPERATURE/FLOW PROBES WERE FOUND DISLODGING FROM AN RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SET-UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23701163 ↗
  • MalfunctionNov 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT TWO 900MR869 TEMPERATURE/FLOW PROBES WERE IDENTIFIED TO BE EASILY DISLODGED FROM A TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23677413 ↗
  • MalfunctionNov 17, 2025Connection Problem; Device Dislodged or Dislocated; Connection Problem; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT THE TEMPERATURE/FLOW PROBE DISLODGED FROM A TEMPERATURE PROBE PORT OF AN RT265 INFANT EVAQUA2 BREATHING CIRCUITS DURING SET UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23577923 ↗
  • MalfunctionOct 9, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT SIX 900MR869 TEMPERATURE/FLOW PROBES WERE IDENTIFIED TO BE EASILY DISLODGED FROM A TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23262844 ↗
  • MalfunctionSep 24, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    A HEALTHCARE FACILITY IN CANADA REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE REPRESENTATIVE THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE LOOSE AT THE PATIENT-END TEMPERATURE PROBE PORT OF AN RT132 INFANT BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23140175 ↗
  • MalfunctionSep 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE EASILY DISLODGED FROM THE PATIENT-END TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23092700 ↗
  • MalfunctionSep 18, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE EASILY DISLODGED FROM THE PATIENT-END TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23092699 ↗
  • MalfunctionAug 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE EASILY DISLODGED FROM THE PATIENT-END TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22919470 ↗
  • MalfunctionAug 24, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE WAS IDENTIFIED TO BE EASILY DISLODGED FROM THE PATIENT-END TEMPERATURE PROBE PORT OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SETUP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22881666 ↗
  • MalfunctionJul 31, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT SEVEN 900MR869 TEMPERATURE/FLOW PROBES WERE FOUND DISLODGING FROM AN RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT DURING SET-UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22683612 ↗
  • MalfunctionMay 29, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN THAT THREE 900MR869 TEMPERATURE/FLOW PROBES WERE IDENTIFIED TO DISLODGE FROM THE PATIENT-END TEMPERATURE PROBE PORT OF AN INFANT BREATHING CIRCUIT PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22102877 ↗
  • MalfunctionMay 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT THIRTY 900MR869 TEMPERATURE/FLOW PROBES WERE FOUND DISLODGING FROM RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS DURING SET-UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21990914 ↗
  • MalfunctionApr 21, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN, VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE DISLODGED FROM THE PATIENT-END TEMPERATURE PROBE PORT OF AN INFANT BREATHING CIRCUIT DURING SET UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878706 ↗
  • MalfunctionApr 10, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT THREE 900MR869 TEMPERATURE/FLOW PROBES WERE FOUND DISLODGING FROM EIGHT RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS DURING SET-UP AND PRIOR TO PATIENT USE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21822657 ↗
  • MalfunctionFeb 19, 2025Mechanical Problem; Mechanical Problem

    A DISTRIBUTOR IN (B)(4) REPORTED THAT A 900MR869 TEMPERATURE PROBE WAS IDENTIFIED TO BE LOOSE AT THE PATIENT-END TEMPERATURE PROBE PORT OF AN INFANT BREATHING CIRCUIT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21417579 ↗
  • MalfunctionFeb 10, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN PENNSYLVANIA REPORTED THAT A TEMPERATURE PROBE WAS FOUND DISLODGING FROM THE BREATHING CIRCUIT. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21351852 ↗
  • MalfunctionFeb 4, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT TEMPERATURE PROBES ARE DISLODGING FROM RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS. THE HEALTHCARE FACILITY HAS NOT CONFIRMED THE NUMBER OF EVENTS OR THE QUANTITY OF UNITS AFFECTED. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21309340 ↗
  • MalfunctionJan 21, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN WISCONSIN REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT 900MR869 TEMPERATURE/FLOW PROBES WERE DISLODGING FROM THE PATIENT-END TEMPERATURE PROBE PORTS OF RT266 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS. THE HEALTHCARE FACILITY DID NOT CONFIRM THE NUMBER OF EVENTS OR THE QUANTITY OF UNITS AFFECTED. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21204398 ↗
  • MalfunctionJan 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT A 900MR869 TEMPERATURE PROBE WAS FOUND DISLODGING FROM THE PATIENT-END TEMPERATURE PROBE PORT OF A F&P HEALTHCARE INFANT BREATHING CIRCUIT DURING PATIENT USE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21175149 ↗
  • MalfunctionJan 16, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN COLORADO REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE THAT ELEVEN 900MR869 TEMPERATURE PROBES WERE FOUND DISLODGING FROM THE PATIENT-END TEMPERATURE PROBE PORT OF F&P HEALTHCARE INFANT BREATHING CIRCUITS DURING SET UP AND PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21175148 ↗
  • MalfunctionJan 7, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN MARYLAND REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE FIELD REPRESENTATIVE, THAT A 900MR869 TEMPERATURE/FLOW PROBE DISLODGED FROM A BREATHING CIRCUIT. UPON FURTHER EVALUATION BY THE HEALTHCARE FACILITY, THEY REPORTED NINE ADDITIONAL 900MR869 TEMPERATURE/FLOW PROBES "FIT LESS SECURELY", HOWEVER WHEN FULLY INSERTED, DID NOT COME LOOSE. THERE WAS NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21098269 ↗
  • MalfunctionDec 16, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN CANADA REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE REPRESENTATIVE THAT NINE 900MR869 TEMPERATURE/FLOW PROBES ARE DISLODGING FROM THE PATIENT-END TEMPERATURE PROBE PORT OF RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS. THIS WAS IDENTIFIED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20948530 ↗
  • MalfunctionDec 16, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    A HEALTHCARE FACILITY IN ILLINOIS REPORTED VIA A FISHER & PAYKEL (F&P) HEALTHCARE REPRESENTATIVE THAT TWO 900MR869 TEMPERATURE PROBES ARE DISLODGING FROM THE PATIENT-END TEMPERATURE PROBE PORT OF BREATHING CIRCUITS. THIS WAS IDENTIFIED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20948529 ↗

Other Humidifier, Respiratory Gas, (Direct Patient Interface) devices

All Humidifier, Respiratory Gas, (Direct Patient Interface) devices →