Fisher & Paykel Healthcare RT125Recalls & reported problems
Fisher & Paykel Healthcare Limited · CIRCUIT NEO BIAS-FLOW - Infant Ventilator Circuit (> 4 L/min) Single Heated · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 09420012431141), received between Feb 16, 2017 and Apr 13, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Crack” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K020332); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Crack3
- Leak/Splash3
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 13, 2023Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA, VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE RT125 INFANT BIAS FLOW BREATHING CIRCUIT WAS FOUND BROKEN AFTER 10 MINUTES OF USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16740533 ↗MalfunctionOct 31, 2022Leak/Splash; Leak/Splash
A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN THAILAND VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE RT125 INFANT BIAS FLOW BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15705167 ↗MalfunctionMay 31, 2022Crack
A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN JAPAN THAT THE Y-PIECE OF A RT125 INFANT BIAS FLOW BREATHING CIRCUIT WAS FOUND CRACKED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14554516 ↗MalfunctionOct 11, 2017Crack; Crack
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE SWIVEL WYE OF THREE RT125 INFANT BIAS FLOW BREATHING CIRCUIT HAD A CRACK. THIS WAS FOUND WHEN THE CIRCUITS KITS WERE REMOVED FROM THE PACKAGING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 6943168 ↗MalfunctionApr 4, 2017Crack; Crack
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A CRACK WAS FOUND ON THE SWIVEL WYE OF AN RT125 INFANT BIAS FLOW BREATHING CIRCUIT. THIS WAS DISCOVERED BEFORE PATIENT USE.
Patient problems: No Patient Involvement
MAUDE report 6459146 ↗MalfunctionFeb 16, 2017Leak/Splash; Leak/Splash
A HOSPITAL IN (B)(4) REPORTED VIA A DISTRIBUTOR THAT THE RT125 INFANT BIAS FLOW BREATHING CIRCUIT DID NOT PASS THE INITIAL LEAK TEST OF THE VENTILATOR WHEN CHECKED BEFORE USE. THE MEDICAL ENGINEER NOTICED THE AIR LEAK WAS AT THE DUCK BILL. NO PATIENT CONSEQUENCE WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6337754 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K020332. One submission often covers many models, so check the product and lot details in each record.
- Z-0553-05 · Dec 21, 2004 · TerminatedRT130 and RT131 Neonatal/ Infant Breathing Circuits, heated wire. For use with MR850 Respiratory Humidifier.
Other Humidifier, Respiratory Gas, (Direct Patient Interface) devices
- Fisher & Paykel Healthcare PT101USFisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)2 recalls5 deaths1,068 reports
- HUDSON RCI IPN913776Teleflex Incorporated · HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)436 reports
- F&P Optiflow+ OPT944Fisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)5 deaths392 reports
- Fisher & Paykel Healthcare PT100USFisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)2 recalls59 reports
- MR850 MR850JHUFisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)93 reports
- HUDSON RCI HUD42500Medline Industries, Inc. · HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE)87 reports
- F&P Optiflow Junior 2 OJR412Fisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)80 reports
- F&P Airvo 900PT561Fisher & Paykel Healthcare Limited · Humidifier, Respiratory Gas, (Direct Patient Interface)1 death56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page