Fisher & Paykel Healthcare OPT844Recalls & reported problems

Fisher & Paykel Healthcare Limited · Nasal Cannula for Airvo, MR880 & MR850 Medium · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Fisher & Paykel Optiflow adult nasal cannula, a soft tube used to deliver oxygen. The 39 reports mostly describe tubing found cracked, torn or split, plus hooks breaking or coming loose and parts detaching, generally noticed before or during use with no patient harm mentioned. No deaths or injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 09420012432124), received between Oct 18, 2018 and Jun 11, 2025 — classified as 39 malfunctions.

The most frequently reported specific device problem is “Break” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break24
  • Material Split, Cut or Torn8
  • Device Handling Problem6
  • Detachment of Device or Device Component4
  • Disconnection1
  • Gas/Air Leak1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions34
  • No Known Impact Or Consequence To Patient4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN CHINA REPORED THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22181946 ↗
  • MalfunctionMar 9, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN CHINA REPORTED THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16517018 ↗
  • MalfunctionOct 5, 2022Break; Break

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED DURING USE. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15547766 ↗
  • MalfunctionSep 29, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE HEADGEAR HOOK OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND BROKEN DURING USE. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15506095 ↗
  • MalfunctionAug 14, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN CHINA REPORTED THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15223047 ↗
  • MalfunctionJun 17, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    A HEALTHCARE FACILITY IN CHINA REPORTED THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DETACHED FROM THE MANIFOLD DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14725595 ↗
  • MalfunctionJan 19, 2022Material Split, Cut or Torn; Break; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13298891 ↗
  • MalfunctionJan 19, 2022Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13298826 ↗
  • MalfunctionJan 11, 2022Material Split, Cut or Torn; Break; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(6) THAT THE TUBING OF AN OPT844 NASAL CANNULA BROKE DURING PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13232430 ↗
  • MalfunctionDec 23, 2021Break

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(6), THAT THE CONNECTOR HOOK OF THE OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13082567 ↗
  • MalfunctionDec 2, 2021Break; Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE CONNECTOR HOOK OF THE OPT844 OPTIFLOW ADULT NASAL CANNULA WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12918923 ↗
  • MalfunctionOct 21, 2021Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(4) THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED DURING PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12677691 ↗
  • MalfunctionOct 14, 2021Break; Break

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(4), THAT THE CONNECTOR HOOK OF THE OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12637273 ↗
  • MalfunctionOct 12, 2021Break; Break

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(6), THAT THE CONNECTOR HOOK OF THE OPT844 OPTIFLOW ADULT NASAL CANNULA WAS DAMAGED DURING PATIENT USE. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12620494 ↗
  • MalfunctionOct 5, 2021Break; Break

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(4) THAT THE OPT844 OPTIFLOW ADULT NASAL CANNULA BROKE DURING PATIENT USE. IT WAS FURTHER REPORTED THAT THE PATIENT TRIED TO REMOVE THE SUBJECT OPT844 OPTIFLOW ADULT NASAL CANNULA. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12580565 ↗
  • MalfunctionAug 17, 2021Break; Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT ONE SIDE OF THE NASAL INTERFACE OF THE OPT844 OPTIFLOW ADULT NASAL CANNULA WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12334586 ↗
  • MalfunctionJun 15, 2021Break; Break

    A DISTRIBUTOR IN (B)(6) REPORTED ON BEHALF OF A HEALTHCARE FACILITY THAT THE OPT844 OPTIFLOW ADULT NASAL CANNULA WAS FOUND DAMAGED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12005476 ↗
  • MalfunctionApr 21, 2021Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT A OPT844 NASAL CANNULA WAS FOUND BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11704223 ↗
  • MalfunctionMar 8, 2021Break; Break; Detachment of Device or Device Component

    A DISTRIBUTOR IN CHINA REPORTED ON BEHALF OF A HEALTHCARE FACILITY, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA DISCONNECTED DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11441199 ↗
  • MalfunctionMar 8, 2021Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR IN (B)(6) REPORTED ON BEHALF OF A HEALTHCARE FACILITY, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA HAD A HOLE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11441123 ↗
  • MalfunctionFeb 25, 2021Break; Break

    A DISTRIBUTOR IN (B)(6) REPORTED ON BEHALF OF A HEALTHCARE FACILITY, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE NASAL INTERFACE OF A OPT844 ADULT NASAL CANNULA WAS FOUND BROKEN BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11381563 ↗
  • MalfunctionFeb 8, 2021Break

    A REGULATORY BODY IN JAPAN REPORTED ON BEHALF OF A HEALTHCARE FACILITY, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE TUBING OF A OPT844 OPTIFLOW ADULT NASAL CANNULA BROKE DURING PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11298946 ↗
  • MalfunctionDec 21, 2020Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT THE TUBING OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS BROKEN DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11052176 ↗
  • MalfunctionDec 21, 2020Break; Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED ON BEHALF OF A DISTRIBUTOR, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE NASAL INTERFACE OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS BROKEN DURING PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11052144 ↗
  • MalfunctionDec 21, 2020Break; Break

    A HEALTHCARE FACILITY IN (B)(6) REPORTED ON BEHALF OF A DISTRIBUTOR, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE NASAL INTERFACE OF AN OPT844 OPTIFLOW ADULT NASAL CANNULA WAS BROKEN DURING PATIENT USE. THERE WAS NO REPORTED PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11051995 ↗

Other Cannula, Nasal, Oxygen devices

All Cannula, Nasal, Oxygen devices →