Fisher & Paykel Healthcare OPT970Recalls & reported problems

Fisher & Paykel Healthcare Limited · Optiflow Tracheostomy Direct Connection · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Fisher & Paykel Optiflow OPT970 tracheostomy direct connection is tubing that attaches a breathing humidifier to a patient's trach tube. The 43 reports mostly describe the tubing being cracked, split, broken, or coming apart from the connector, sometimes while in use on a patient. No deaths or injuries were reported. There are no recalls for this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 09420012436917), received between Dec 4, 2020 and May 22, 2026 — classified as 43 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn32
  • Break14
  • Detachment of Device or Device Component10
  • Stretched4
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions41
  • Low Oxygen Saturation2

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2026Detachment of Device or Device Component

    A HEALTHCARE FACILITY IN THE NETHERLANDS REPORTED THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DETACHED FROM THE 3-WAY CONNECTOR DURING PATIENT USE. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25250998 ↗
  • MalfunctionMar 30, 2026Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24736593 ↗
  • MalfunctionMar 4, 2026Stretched; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24508394 ↗
  • MalfunctionMar 3, 2026Stretched; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING PATIENT USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24503221 ↗
  • MalfunctionOct 28, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN CHINA REPORTED THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING PATIENT USE. IT WAS FURTHER REPORTED BY THE HEALTHCARE FACILITY THAT THE SUBJECT DEVICE WAS IN USE FOR APPROXIMATELY THREE TO FOUR DAYS. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23401012 ↗
  • MalfunctionSep 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22949297 ↗
  • MalfunctionJun 4, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22145316 ↗
  • MalfunctionApr 29, 2025Stretched; Material Split, Cut or Torn; Stretched; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21931838 ↗
  • MalfunctionMar 9, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WERE NO PATIENT CONSEQUENCES AS THE ISSUE WAS FOUND WHILST THE SUBJECT DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21556184 ↗
  • MalfunctionNov 7, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED AFTER ONE DAY OF USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20636413 ↗
  • MalfunctionOct 25, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20539735 ↗
  • MalfunctionOct 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED AND DETACHED AFTER ONE HOUR OF USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20404456 ↗
  • MalfunctionMay 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19402401 ↗
  • MalfunctionJan 31, 2024Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGE AS A RESULT OF BEING PULLED BY THE PATIENT DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18620096 ↗
  • MalfunctionAug 2, 2023Stretched; Material Split, Cut or Torn; Stretched; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17446402 ↗
  • MalfunctionMar 23, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16606479 ↗
  • MalfunctionMar 23, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16606434 ↗
  • MalfunctionJan 15, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16172280 ↗
  • MalfunctionDec 11, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    A HEALTHCARE FACILITY IN CHINA REPORTED THAT THE AFTER FIVE DAYS OF USE THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DETACHED FROM 3-WAY CONNECTOR. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15956416 ↗
  • MalfunctionDec 8, 2022Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + ADULT NASAL CANNULA WAS FOUND DAMAGED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15947080 ↗
  • MalfunctionDec 6, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN CHINA REPORTED THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING USE. THE HEALTHCARE FACILITY ALSO REPORTED THAT THE PATIENT DESATURATED, THE SUBJECT DEVICE WAS REPLACED AND THE PATIENT RECOVERED. NO FURTHER PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 15928314 ↗
  • MalfunctionJun 17, 2022Material Split, Cut or Torn; Break; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14725754 ↗
  • MalfunctionJun 2, 2022Break; Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN CHINA THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING PATIENT USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14583009 ↗
  • MalfunctionMay 4, 2022Break; Material Split, Cut or Torn; Material Split, Cut or Torn

    A DISTRIBUTOR REPORTED ON BEHALF OF A HEALTHCARE FACILITY IN (B)(4) THAT THE TUBING OF A OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS DAMAGED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14275164 ↗
  • MalfunctionApr 25, 2022Break; Material Split, Cut or Torn; Material Split, Cut or Torn

    A HEALTHCARE FACILITY IN (B)(6) REPORTED THAT THE TUBING OF AN OPT970 OPTIFLOW + TRACHEOSTOMY DIRECT CONNECTION WAS FOUND DAMAGED DURING PATIENT USE. THERE WERE NO REPORTED PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14198070 ↗