Fisher & Paykel Healthcare PT301USRecalls & reported problems

Fisher & Paykel Healthcare Limited · Respiratory Support Device · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 09420012466662), received between Jul 8, 2025 and Apr 30, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 7, 2024Class IIOpen, Classified

    Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

    Z-3262-2024 · Root cause: Software change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Audible Alarm6
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1
  • Low Oxygen Saturation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2026No Audible Alarm

    A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A PT301 AIRVO 3 HUMIDIFIER (AIRVO 3) GENERATED A VISUAL ALARM. THE NATURE OF THE VISUAL ALARM SUGGESTS THERE MAY BE A POTENTIAL ISSUE WITH THE SUBJECT DEVICE'S AUDIBLE ALARM. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25053621 ↗
  • MalfunctionMar 25, 2026No Audible Alarm

    A healthcare facility in (b)(6) reported that a PT301 Airvo 3 generated an alarm. The nature of the alarm suggests there may be a potential issue with the subject device's audible alarm.There was no patient involvement as the issue was found whilst the subject device was not in use on a patient.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24694621 ↗
  • MalfunctionFeb 26, 2026No Audible Alarm

    A HEALTHCARE FACILITY REPORTED VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A PT301 AIRVO 3 GENERATED A VISUAL ALARM. THE NATURE OF THE VISUAL ALARM SUGGESTS THERE MAY BE A POTENTIAL ISSUE WITH THE SUBJECT DEVICE'S AUDIBLE ALARM. THE HEALTHCARE FACILITY REPORTED THAT THE PATIENT EXPERIENCED DESATURATION. F&P HAVE REQUESTED FOR FURTHER INFORMATION FROM THE HEALTHCARE FACILITY, INCLUDING MORE DETAILS ON PATIENT CONSEQUENCES, THE SEQUENCE OF EVENTS, AND FOR THE RETURN OF THE SUBJECT DEVICE. F&P WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF F&P'S INVESTIGATION.

    Patient problems: Insufficient Information; Low Oxygen Saturation

    MAUDE report 24465974 ↗
  • MalfunctionAug 13, 2025No Audible Alarm; No Audible Alarm

    A HEALTHCARE FACILITY IN NEW JERSEY REPORTED THAT A PT301 AIRVO 3 HUMIDIFIER (AIRVO 3) GENERATED A VISUAL ALARM. THE NATURE OF THE VISUAL ALARM SUGGESTS THERE MAY BE A POTENTIAL ISSUE WITH THE SUBJECT DEVICE'S AUDIBLE ALARM. DURING DEVICE EVALUATION AT THE FISHER & PAYKEL HEALTHCARE (F&P) REGIONAL OFFICE IN THE USA, IT WAS FOUND THAT THERE WAS NO AUDIBLE ALARM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22801292 ↗
  • MalfunctionJul 23, 2025No Audible Alarm; Insufficient Device Problem Information; No Audible Alarm; Insufficient Device Problem Information

    A HEALTHCARE FACILITY IN GEORGIA REPORTED VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A PT301 AIRVO 3 HUMIDIFIER (AIRVO 3) GENERATED AN ALARM. THE NATURE OF THE ALARM SUGGESTS THERE MAY BE A POTENTIAL ISSUE WITH THE SUBJECT DEVICE'S AUDIBLE ALARM. THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILST THE SUBJECT DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22584183 ↗
  • MalfunctionJul 8, 2025No Audible Alarm; No Audible Alarm

    A HEALTHCARE FACILITY IN NEW JERSEY REPORTED VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A PT301 AIRVO 3 DID NOT HAVE AN AUDIBLE ALARM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22454709 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page