i-STAT 05p58-25Recalls & reported problems

Abbott Point Of Care Inc. · CARTRIDGE, Total B-hCG · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Abbott i-STAT Total B-hCG cartridge is a point-of-care test for human chorionic gonadotropin, a pregnancy hormone. Its 105 reports are all malfunction reports, and nearly all describe false positive results, where the test showed a positive result that other tests did not match. There were no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
105
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 105 adverse event reports that list this exact device (DI 10054749001009), received between Jul 10, 2017 and Apr 16, 2024 — classified as 105 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (103 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 12 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result103
  • False Negative Result2

Patient problems

  • Abdominal Pain26
  • No Information22
  • No Clinical Signs, Symptoms or Conditions14
  • Nausea10
  • Vomiting9
  • Pain6
  • Headache5
  • No Code Available5

Recent adverse event reports

The 25 most recent of 105 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Anxiety

    MAUDE report 19116994 ↗
  • MalfunctionApr 2, 2024False Positive Result; False Positive Result

    ON 27-MAR-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 13 YEAR OLD FEMALE PATIENT WITH OSTEOGENESIS IMPERFECTA. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE TESTED RESULTS SAMPLE I-STAT (B)(6) 2024 09:26 30.3 1 I-STAT (B)(6) 2024 09:50 23.7 2 THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. I-STAT POSITIVE CUT-OFF VALUES: B-HCG < 5.0 IU/L NEGATIVE 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE B-HCG >25.0 IU/L POSITIVE INTENDED USE: THE I-STAT® TOTAL BETA-HUMAN CHORIONIC GONADOTROPIN (B-HCG) TEST IS AN IN VITRO DIAGNOSTIC TEST FOR THE QUANTITATIVE AND QUALITATIVE DETERMINATION OF B-HCG IN VENOUS WHOLE BLOOD OR PLASMA SAMPLES USING THE I-STAT 1 ANALYZER SYSTEMS. THE TEST IS INTENDED TO BE USED AS AN AID IN THE EARLY DETECTION OF PREGNANCY AND IS FOR PRESCRIPTION USE ONLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19023292 ↗
  • MalfunctionMar 19, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Anxiety; Lymphoma

    MAUDE report 18934397 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    ON 19-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 39 YEAR OLD FEMALE PATIENT WITH ABDOMINAL PAIN, NAUSEA, DIARRHEA. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD; DATE; COLLECTED; TESTED & RESULT. I-STAT, (B)(6) 2022, 2:03PM, 2:14PM & 52.6 IU/ML. ROCHE, (B)(6) 2022, NI, NI & <5.0 IU/ML. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. I-STAT POSITIVE CUT-OFF VALUES: B-HCG < 5.0 IU/L NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE. B-HCG >25.0 IU/L POSITIVE. INTENDED USE: THE I-STAT® TOTAL BETA-HUMAN CHORIONIC GONADOTROPIN (B-HCG) TEST IS AN IN VITRO DIAGNOSTIC TEST FOR THE QUANTITATIVE AND QUALITATIVE DETERMINATION OF B-HCG IN VENOUS WHOLE BLOOD OR PLASMA SAMPLES USING THE I-STAT 1 ANALYZER SYSTEMS. THE TEST IS INTENDED TO BE USED AS AN AID IN THE EARLY DETECTION OF P

    Patient problems: Abdominal Pain; Diarrhea; Nausea

    MAUDE report 18702697 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Abdominal Pain; Decreased Appetite; Dysuria; Nausea; Vomiting

    MAUDE report 18702696 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Abdominal Pain; Nausea; Vomiting

    MAUDE report 18702680 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Dysphasia; Hallucination; Memory Loss/Impairment; Nausea; Numbness

    MAUDE report 18702677 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    ON 19-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 47 YEAR OLD FEMALE PATIENT WITH ALCOHOL INTOXICATION. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD; DATE; COLLECTED; TESTED; RESULT (IU/ML). I-STAT 1, (B)(6) 2023, 8:59PM, 9:14PM & 49.7 (POSITIVE). ROCHE, NI, NI, NI & <5.0 (NEGATIVE). THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. I-STAT POSITIVE CUT-OFF VALUES: B-HCG < 5.0 IU/L, NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L, INDETERMINATE. B-HCG >25.0 IU/L, POSITIVE. INTENDED USE: THE I-STAT® TOTAL BETA-HUMAN CHORIONIC GONADOTROPIN (B-HCG) TEST IS AN IN VITRO DIAGNOSTIC TEST FOR THE QUANTITATIVE AND QUALITATIVE DETERMINATION OF B-HCG IN VENOUS WHOLE BLOOD OR PLASMA SAMPLES USING THE I-STAT 1 ANALYZER SYSTEMS. THE TEST IS INTENDED TO BE USED AS AN AID IN THE EARLY DETECTION OF

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18702673 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    ON 19-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 51 YEAR OLD FEMALE PATIENT WITH FLANK PAIN AND A COUGH. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD; DATE; COLLECTED; TESTED & RESULT (IU/ML). I-STAT, (B)(6) 2023, 02:53AM, 03:05AM & 31.3. I-STAT, (B)(6) 2023, 03:24AM, 03:36AM & 36.3. ROCHE, (B)(6) 2023, NI, NI & <5.0. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. I-STAT POSITIVE CUT-OFF VALUES: B-HCG < 5.0 IU/L; NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L; INDETERMINATE. B-HCG >25.0 IU/L; POSITIVE . INTENDED USE: THE I-STAT® TOTAL BETA-HUMAN CHORIONIC GONADOTROPIN (B-HCG) TEST IS AN IN VITRO DIAGNOSTIC TEST FOR THE QUANTITATIVE AND QUALITATIVE DETERMINATION OF B-HCG IN VENOUS WHOLE BLOOD OR PLASMA SAMPLES USING THE I-STAT 1 ANALYZER SYSTEMS. THE TEST IS INTENDED TO BE USED

    Patient problems: Cough; Pain

    MAUDE report 18702668 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Abdominal Pain; Nausea; Vomiting

    MAUDE report 18702667 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    ON 19-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 43 YEAR OLD FEMALE PATIENT WITH VOMITING, EMESIS, DIARRHEA, AND ABDOMINAL PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD; DATE; COLLECTED; TESTED; RESULT (IU/ML) I-STAT, (B)(6) 2023, 10:59AM, 11:10AM & 35.2. ROCHE, (B)(6) 2023, NI, NI & <5.0. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. I-STAT POSITIVE CUT-OFF VALUES: B-HCG < 5.0 IU/L, NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L, INDETERMINATE. B-HCG >25.0 IU/L, POSITIVE. INTENDED USE: THE I-STAT® TOTAL BETA-HUMAN CHORIONIC GONADOTROPIN (B-HCG) TEST IS AN IN VITRO DIAGNOSTIC TEST FOR THE QUANTITATIVE AND QUALITATIVE DETERMINATION OF B-HCG IN VENOUS WHOLE BLOOD OR PLASMA SAMPLES USING THE I-STAT 1 ANALYZER SYSTEMS. THE TEST IS INTENDED TO BE USED AS AN AID IN THE EARLY D

    Patient problems: Abdominal Pain; Diarrhea; Nausea; Vomiting

    MAUDE report 18702665 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Pain

    MAUDE report 18702664 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Chest Pain; Decreased Respiratory Rate

    MAUDE report 18702663 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Abdominal Pain; Decreased Appetite; Nausea; Vomiting

    MAUDE report 18702662 ↗
  • MalfunctionFeb 14, 2024False Positive Result; False Positive Result

    ON 19-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 23 YEAR OLD FEMALE PATIENT WITH LEG PAIN, AND NUMBNESS RIGHT TOE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULT (IU/ML) I-STAT ON (B)(6) 2023 01:04 PM 01:15 PM 33.8 ROCHE ON (B)(6 2023 NI NI <5.0 THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. I-STAT POSITIVE CUT-OFF VALUES: B-HCG < 5.0 IU/L NEGATIVE 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE B-HCG >25.0 IU/L POSITIVE INTENDED USE: THE I-STAT® TOTAL BETA-HUMAN CHORIONIC GONADOTROPIN (B-HCG) TEST IS AN IN VITRO DIAGNOSTIC TEST FOR THE QUANTITATIVE AND QUALITATIVE DETERMINATION OF B-HCG IN VENOUS WHOLE BLOOD OR PLASMA SAMPLES USING THE I-STAT 1 ANALYZER SYSTEMS. THE TEST IS INTENDED TO BE USED AS AN AID IN THE EARLY DETECTION OF PREGNANCY AND IS

    Patient problems: Limb Pain; Numbness

    MAUDE report 18702661 ↗
  • MalfunctionJan 16, 2024False Positive Result; False Positive Result

    NA.

    Patient problems: Abdominal Pain; Pancreatitis

    MAUDE report 18523076 ↗
  • MalfunctionJan 10, 2024False Positive Result; False Positive Result

    ON 02-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 45 YEAR OLD FEMALE PATIENT WITH ABDOMINAL PAIN, PELVIC PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULT (UNITS) SAMPLE I-STAT (B)(6) 2023 8:32 AM 8:32 AM 37.6 IU/L VENOUS #1 I-STAT (B)(6) 2023 8:32 AM 9:02 AM 73.0 IU/L VENOUS #1 VITROS (B)(6) 2023 8:28 AM 9:03 AM <2 MIU/ML VENOUS #2 I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L NEGATIVE 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE B-HCG = 25.0 IU/L POSITIVE THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: Abdominal Pain

    MAUDE report 18482974 ↗
  • MalfunctionJan 10, 2024False Positive Result; False Positive Result

    ON 02-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 43 YEAR OLD FEMALE PATIENT WITH MALIGNANT NEOPLASM. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULT (UNITS) SAMPLE I-STAT (B)(6) 23 8:47 AM 8:47 AM 77.5 IU/L VENOUS #1 VITROS (B)(6) 23 8:55 AM 9:29 AM <2 MIU/ML VENOUS #2 I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L NEGATIVE 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE B-HCG = 25.0 IU/L POSITIVE THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: Non-malignant Neoplasm

    MAUDE report 18482973 ↗
  • MalfunctionJan 10, 2024False Positive Result; False Positive Result

    ON 02-JAN-2024, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A 60 YEAR OLD FEMALE PATIENT WITH POST-MENOPAUSAL BLEEDING. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULT SAMPLE I-STAT (B)(6) 2023 09:50 09:50 102.4 IU/L VENOUS #1 VITROS (B)(6) 2023 13:33 14:02 <2 MIU/ML VENOUS #2 I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L NEGATIVE 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE B-HCG = 25.0 IU/L POSITIVE THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: Intermenstrual Bleeding

    MAUDE report 18482972 ↗
  • MalfunctionNov 14, 2023False Positive Result; False Positive Result

    ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULTS ON A 34 YEAR OLD FEMALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULT (UNITS) SAMPLE I-STAT, (B)(6) 2023, 16:14, 16:14, >2000 IU/L, VENOUS #1. NI, (B)(6) 2023, 16:14, NI, <2.4 MIU/ML , VENOUS #1. I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L , NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L, INDETERMINATE. B-HCG = 25.0 IU/L, POSITIVE. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18131478 ↗
  • MalfunctionNov 14, 2023False Positive Result; False Positive Result

    ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULTS ON A 22 YEAR OLD FEMALE PATIENT WITH HEADACHES. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE TESTED RESULT SAMPLE I-STAT ON(B)(6) 2023 0819 1367 IU/L A ROCHE ON (B)(6) 2023 0814 <1 IU/L B I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L NEGATIVE 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE B-HCG = 25.0 IU/L POSITIVE THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: Headache

    MAUDE report 18131476 ↗
  • MalfunctionNov 14, 2023False Positive Result; False Positive Result

    ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT OF >2000 IU/L ON A 22 YEAR OLD FEMALE PATIENT WITH GENERALIZED ABDOMINAL PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. METHOD, DATE, COLLECTED, TESTED, RESULT, SAMPLE; I-STAT, (B)(6) 2023, 0149, 0149, >2000 IU/L, VENOUS #1; NI, (B)(6) 2023, 0149, NI, <2.4 MIU/ML, VENOUS #1. I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L, NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L, INDETERMINATE. B-HCG = 25.0 IU/L POSITIVE THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: Abdominal Pain

    MAUDE report 18131473 ↗
  • MalfunctionNov 14, 2023False Positive Result; False Positive Result

    ON 07-NOV-2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULTS ON A 25 YEAR OLD FEMALE PATIENT WITH HEADACHES. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD, DATE, COLLECTED, TESTED, RESULT, SAMPLE; I-STAT, (B)(6) 2023, 6:22 PM, 6:22 PM, 96.6 IU/L, VENOUS #1; NI, (B)(6) 2023, 6:22 PM, NI, <2.4 MIU/ML, VENOUS #1; I-STAT, (B)(6) 2023, 8:19 PM, 8:19 PM, 162 IU/L, VENOUS #2; I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L, NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L, INDETERMINATE. B-HCG = 25.0 IU/L, POSITIVE. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: Headache

    MAUDE report 18131472 ↗
  • MalfunctionNov 14, 2023False Positive Result; False Positive Result

    ON (B)(6)2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULTS ON A 12 YEAR OLD FEMALE PATIENT WITH SICKLE CELL. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE TESTED RESULT SAMPLE I-STAT, (B)(6)2023, 1214, 373 IU/L, A. ROCHE, (B)(6)2023, 1218, <1 IU/L , B. I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L , NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L , INDETERMINATE. B-HCG = 25.0 IU/L , POSITIVE . THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18131470 ↗
  • MalfunctionNov 8, 2023False Positive Result; False Positive Result

    ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A POSITIVE RESULT ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED RESULT SAMPLE: I-STAT (B)(6) 2023 3:49 PM 166.2 WHOLE BLOOD. LAB COBAS (B)(6) 2023 4:08 PM <0.6 SERUM. LAB COBAS (B)(6) 2023 5:43PM <0.6 SERUM. I-STAT (B)(6) 2023 6:01 PM <5.0 WHOLE BLOOD . I-STAT POSITIVE CUT-OFF VALUES: B-HCG = 5.0 IU/L NEGATIVE. 5.0 < SS-HCG < 25.0 IU/L INDETERMINATE. B-HCG = 25.0 IU/L POSITIVE . THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18091752 ↗

FDA clearance and classification

Cleared through 510(k) K133002 on Apr 16, 2015 (decision: Substantially Equivalent). Submitted by Abbott Point of Care, Inc. as “i-STAT Total beta-hCG Test, i-STAT Total beta-hCG Controls, i-STAT Total beta-hCG Calibration Verification Materials”. The FDA review took 569 days. FDA record ↗

The FDA regulates product code DHA (system, test, human chorionic gonadotropin) under 21 CFR 862.1155 as a Class II device. 21 CFR 862.1155 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the i-STAT 05p58-25

Has the i-STAT 05p58-25 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10054749001009), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the i-STAT 05p58-25?

The FDA's MAUDE database lists 105 adverse event reports for this device, received between Jul 10, 2017 and Apr 16, 2024. In those reports, the most common specific device problems are “False Positive Result” (103) and “False Negative Result” (2); the most common patient problems are “Abdominal Pain” (26), “Nausea” (10) and “Vomiting” (9). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the i-STAT 05p58-25?

None of the reports are classified as deaths, 0 as injuries and 105 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the i-STAT 05p58-25 FDA approved?

It was cleared through 510(k) K133002 on Apr 16, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the i-STAT 05p58-25, and is it still sold?

The FDA's device database lists Abbott Point Of Care Inc. as the company responsible for it (brand i-STAT, model 05p58-25). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, Test, Human Chorionic Gonadotropin devices

All System, Test, Human Chorionic Gonadotropin devices →