Skip to content

MEDLINE DYND13508Recalls & reported problems

Medline Industries, Inc. · CATHETER,URETHRAL,REDRUBBER,STERILE,8FR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10080196031139), received between Jun 18, 2018 and Jul 18, 2018 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. May 4, 2026Open, Classified

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

    Z-2479-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Malposition of Device1
  • Off-Label Use1

Patient problems

  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 18, 2018Off-Label Use

    IT WAS REPORTED THAT THE URETHRAL CATHETER WAS BEING USED TO MAINTAIN TRACHEOSTOMY STOMA PATENCY. REPORTEDLY, THE DEVICE BECAME LODGED IN THE PATIENT'S LEFT BRONCHUS AND AN UNIDENTIFIED PROCEDURE WAS REQUIRED TO REMOVE THE URETHRAL CATHETER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7697751 ↗
  • MalfunctionJun 18, 2018Malposition of Device; Device Dislodged or Dislocated

    RED RUBBER CATHETER WAS PLACED IN PATIENT'S TRACHEOSTOMY STOMA TO MAINTAIN PATENCY. THE CATHETER BECAME LODGED IN THE LEFT BRONCHUS AND WAS SUBSEQUENTLY REMOVED IN OR.

    Patient problems: No Code Available

    MAUDE report 7609368 ↗

Other CATHETER, URETERAL, GASTRO-UROLOGY devices

All CATHETER, URETERAL, GASTRO-UROLOGY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page