Medline HCS8715MARecalls & reported problems
Medline Industries, Inc. · REGULATOR,MINI,0-15LPM,870 CGA · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10080196193905), received between May 8, 2026 and Sep 29, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak7
- Inaccurate Flow Rate2
- Detachment of Device or Device Component1
- Improper Flow or Infusion1
- Mechanical Problem1
- Noise, Audible1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936645 ↗MalfunctionSep 29, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936431 ↗MalfunctionSep 29, 2026Gas/Air Leak
It was reported that the customer is having issues with the units leaking from the flow adjustment dial.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933028 ↗MalfunctionSep 29, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931352 ↗MalfunctionSep 24, 2026Inaccurate Flow Rate; Mechanical Problem; Gas/Air Leak
This VMSR report summarizes 6 malfunction events that occurred between July 1 and September 30, 2023. The account reported that oxygen regulators (HCS8715MA) exhibited loose flow control dials and air leaks during use. The account further reported unstable flow settings and leakage when changing liters per minute settings. No patient information was available. No patient or user injuries were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883758 ↗MalfunctionSep 22, 2026Gas/Air Leak; Improper Flow or Infusion
It was reported that "1 MORE REGULATOR NOT PUTTING OUT THE PROPER LITER FLOW".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50856397 ↗MalfunctionSep 21, 2026Gas/Air Leak; Noise, Audible
This summary report covers malfunction events that occurred between January 1 and March 31, 2024. This VMSR summarizes one malfunction event involving 30 Medline oxygen regulators (catalog number HCS8715MA). The account reported that the units appeared to be leaking from the flow adjustment dial. Evaluation of the returned devices found oxygen escaping from multiple units; two units were confirmed to leak from the flow dial, eight units exhibited a rattling sound, some units did not leak during testing, and some units could not be fully tested because components were missing. Patient demographic information was not available. No patient or user injuries were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50842598 ↗MalfunctionAug 31, 2026Inaccurate Flow Rate
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50530730 ↗MalfunctionMay 8, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE METAL PIECE THAT ALLOWS THE TUBING TO CONNECT TO THE REGULATOR BROKE OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25129172 ↗
Other REGULATOR, PRESSURE, GAS CYLINDER devices
- ISPAN 8065797303Alcon Laboratories, Inc. · REGULATOR, PRESSURE, GAS CYLINDER56 reports
- Medline HCS8708MMedline Industries, Inc. · REGULATOR, PRESSURE, GAS CYLINDER10 reports
- EM L280-220KAllied Medical, LLC · Regulator, Pressure, Gas Cylinder2 reports
- Mini CGA 870 style regulator; 0-15 lpm RES043Sunset Healthcare Solutions, Inc. · Regulator, Pressure, Gas Cylinder2 reports
- Drive 18302GDELDrive Devilbiss Healthcare · REGULATOR, PRESSURE, GAS CYLINDER1 death1 report
- Western/Scott Fetzer M1-540-15FMWestern Enterprises · Regulator, Pressure, Gas Cylinder1 report
- MediTec 0721109USGCE, s.r.o. · Regulator, Pressure, Gas Cylinder1 report
- Cavagna VIPROXY 1 TouchCavagna North America Inc · Regulator, Pressure, Gas Cylinder1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page