Medline DYNJP2001SRecalls & reported problems
Medline Industries, Inc. · GOWN,SIRUS,NON REINFRCD,LARGE,SET IN SL · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10080196304660), received between Apr 7, 2021 and Sep 17, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jan 7, 2026Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Z-2129-2026 · Root cause: Process control · Recall details · FDA record ↗
- Jan 7, 2026Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Z-1424-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available3
- Material Integrity Problem2
- Contamination1
- Device Contaminated During Manufacture or Shipping1
- Fluid/Blood Leak1
- Material Fragmentation1
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Post Operative Wound Infection2
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Device Contaminated During Manufacture or Shipping
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50810340 ↗MalfunctionSep 14, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50762991 ↗MalfunctionSep 8, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50674947 ↗MalfunctionSep 1, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50541854 ↗MalfunctionJun 12, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE GOWN FAILED TO PROTECT SCRUB NURSE FROM BLOOD EXPOSURE. A SMALL AMOUNT OF BLOOD WAS SPLASHED ONTO THE SLEEVE AND SOAKED THROUGH THE GOWN ON TO THE SKIN OF THE SCRUB NURSE.
Patient problems: Insufficient Information
MAUDE report 25492819 ↗InjuryFeb 12, 2026Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE CUSTOMER, THREE GOWNS INCLUDED IN A LATER RECALL WERE USED DURING A "TOTAL KNEE ARTHROPLASTY" PROCEDURE ON (B)(6) 2026. THE CUSTOMER REPORTED THAT THE PATIENT DEVELOPED AN "INFECTED KNEE[S] AND REQUIRED A SECOND SURGERY ON (B)(6) 2026."
Patient problems: Post Operative Wound Infection
MAUDE report 24327958 ↗InjuryFeb 12, 2026Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE CUSTOMER, THREE GOWNS INCLUDED IN A LATER RECALL WERE USED DURING A "TOTAL KNEE ARTHROPLASTY" PROCEDURE ON (B)(6) 2026. THE CUSTOMER REPORTED THAT THE PATIENT DEVELOPED AN "INFECTED KNEE[S] AND REQUIRED A SECOND SURGERY ON (B)(6) 2026."
Patient problems: Post Operative Wound Infection
MAUDE report 24327383 ↗MalfunctionJun 7, 2021Material Fragmentation
IT WAS REPORTED THAT SURGICAL GOWN MATERIAL WAS COMING OFF THE PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11946902 ↗MalfunctionApr 7, 2021Contamination; Material Frayed
IT WAS REPORTED, A PHYSICIAN AND TECHNICIAN WERE REMOVING STERILE SURGICAL GOWN "STRANDS AS LONG AS 2-3 CENTIMETERS THAT HAD MADE ITS WAY TO IMPLANTS, INSTRUMENTS, AND EVEN INSIDE THE BRAIN."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11631047 ↗
Other GOWN, SURGICAL devices
- Medline DYNJP2202SMedline Industries, Inc. · GOWN, SURGICAL2 recalls5 reports
- MEDLINE DYNJP2002SMedline Industries, Inc. · GOWN, SURGICAL2 recalls4 reports
- Medline DYNJP2602Medline Industries, Inc. · GOWN, SURGICAL2 recalls3 reports
- Medline Industries, Inc. DYNJP2402Medline Industries, Inc. · GOWN, SURGICAL2 recalls2 reports
- MEDLINE DYNJP2002Medline Industries, Inc. · GOWN, SURGICAL2 recalls2 reports
- Medline DYNJP2103Medline Industries, Inc. · GOWN, SURGICAL2 recalls1 report
- Medline DYNJP2203Medline Industries, Inc. · GOWN, SURGICAL2 recalls1 report
- Medline DYNJP2701Medline Industries, Inc. · GOWN, SURGICAL2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page