Medline DYNJP2411Recalls & reported problems
Medline Industries, Inc. · SHEET,DRAPE,ECONOMY,40X58,STERILE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10080196745487), received between Apr 9, 2026 and Apr 13, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Contamination” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 7, 2026Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Z-1423-2026 · Root cause: Under Investigation by firm · FDA record ↗
- Jan 7, 2026Open, Classified
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Z-2128-2026 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Contamination2
Patient problems
- Emotional Changes2
- Impaired Healing2
- Post Operative Wound Infection2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 13, 2026Contamination
ACCORDING TO THE FACILITY ON (B)(6) 2025, "PRODUCT WAS USED AS A STERILE DRAPE FOR A BREAST AUGMENTATION PROCEDURE." THE FACILITY REPORTED, "THE PATIENT'S LEFT BREAST HAS BEEN INFECTED EVER SINCE THE SURGERY AND WILL REQUIRE AN ADDITIONAL SURGERY TO CLEAR THE INFECTION OUT."
Patient problems: Emotional Changes; Impaired Healing; Post Operative Wound Infection
MAUDE report 24858298 ↗InjuryApr 9, 2026Contamination
ACCORDING TO THE FACILITY ON (B)(6) 2025, "PRODUCT WAS USED AS A STERILE DRAPE FOR A BREAST AUGMENTATION PROCEDURE." THE FACILITY REPORTED, "THE PATIENT'S LEFT BREAST HAS BEEN INFECTED EVER SINCE THE SURGERY AND WILL REQUIRE AN ADDITIONAL SURGERY TO CLEAR THE INFECTION OUT." THE FACILITY FURTHER REPORTED, "THERE HAS BEEN ONGOING MEDICAL CARE SINCE THE SURGERY" AS A RESULT OF "DELAYED WOUND HEALING AND NON-HEALING WITH INFECTION OF IMPLANT ON THE LEFT SIDE." THE FACILITY STATED, "THE IMPLANT WILL NEED TO BE REMOVED, THE WOUND WILL NEED TO BE CLOSED, AND THE IMPLANT REPLACED AT LATER DATE." AND "THE PATIENT HAS SUFFERED INJURY TO HER BREAST DURING THE INFECTION." PATIENT IS "FRUSTRATED AND ANGRY." PER THE FACILITY "THE DRAPE WAS USED AND DISCARDED."
Patient problems: Emotional Changes; Impaired Healing; Post Operative Wound Infection
MAUDE report 24833226 ↗
Other Drape, surgical devices
- Medline DYNJP8002Medline Industries, Inc. · Drape, surgical2 recalls2 reports
- Medline DYNJP2414RMedline Industries, Inc. · Drape, surgical2 recalls2 reports
- Medline DYNJP2410Medline Industries, Inc. · Drape, surgical2 recalls2 reports
- Medline DYNJP2416Medline Industries, Inc. · Drape, surgical2 recalls1 report
- Medline DYNJP3008UGMedline Industries, Inc. · Drape, surgical2 recalls1 report
- Medline DYNJP8201UGMedline Industries, Inc. · Drape, surgical2 recalls0 reports
- Medline DYNJP2412TMedline Industries, Inc. · Drape, surgical2 recalls0 reports
- Medline DYNJP4003Medline Industries, Inc. · Drape, surgical2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page