Skip to content

Medline DYNJP3102ARecalls & reported problems

Medline Industries, Inc. · DRAPE,LAP,CHOLE,W/TROUGHS,STERILE · Class 2

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
3

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10080196778362), received between Sep 19, 2025 and Sep 12, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Component Missing” (1 report).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 7, 2026Open, Classified

    Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

    Z-2128-2026 · Root cause: Process control · FDA record ↗

  2. Jan 7, 2026Open, Classified

    Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

    Z-1423-2026 · Root cause: Under Investigation by firm · FDA record ↗

  3. Feb 13, 2025Open, Classified

    Potential breach in pouch packaging which could lead to loss of sterility.

    Z-1363-2025 · Root cause: Packaging process control · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Component Missing1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 12, 2026Component Missing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50745061 ↗
  • InjurySep 19, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE CUSTOMER, THE VELCRO FROM THE DRAPE "BECAME DETACHED AND ADHERED TO THE PATIENT'S BOWEL" ON (B)(6) 2025 AND EXTENDED THE LENGTH OF THE PROCEDURE IN ORDER TO "REMOVE SAFELY".

    Patient problems: Foreign Body In Patient

    MAUDE report 23097425 ↗

Other Drape, surgical, exempt devices

All Drape, surgical, exempt devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.