Skip to content

MEDLINE DYNJ04049Recalls & reported problems

Medline Industries, Inc. · SCISSORS,IRIS,CURVED,STD,4.5',STERILE · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10080196847273), received between Sep 9, 2026 and Sep 17, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 21, 2024Open, Classified

    Tip protector may fall off within the packaging, resulting in compromised sterile barrier system and loss of sterility.

    Z-0465-2025 · Root cause: Under Investigation by firm · FDA record ↗

  2. Apr 30, 2024Open, Classified

    There is the potential of the tip protector to fall off 4.5" sterile, curved, Iris Scissors

    Z-2024-2024 · Root cause: Under Investigation by firm · FDA record ↗

  3. Aug 3, 2023Open, Classified

    4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.

    Z-2595-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Misassembled During Manufacturing /Shipping1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Detachment of Device or Device Component

    It was reported that "the tip protector off in the packaging of only 1 pair of scissors."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803352 ↗
  • MalfunctionSep 9, 2026Detachment of Device or Device Component; Device Misassembled During Manufacturing /Shipping

    It was reported that the scissor cap is falling off of the scissors causing the sharp tip to puncture the packaging and potentially cause harm to providers.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50690272 ↗

Other SCISSORS, GENERAL, SURGICAL devices

All SCISSORS, GENERAL, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page