MENTOR MemoryGel Xtra Breast Implant SMHX375Recalls & reported problems

Mentor Texas L.P. · Smooth Moderate High Profile Xtra, 375 cc |Smooth|Round|Silicone| · Class 3 · Implant

All models of MENTOR MemoryGel Breast Implants (368) · MENTOR MemoryGel Xtra Breast Implant (88)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Mentor MemoryGel Xtra breast implant is a silicone gel-filled implant. Most of the 127 reports describe capsular contracture, where scar tissue tightens around the implant, and some of these led to replacement or removal surgery; others describe hematoma, deformity, and the implant rupturing. The reports include no deaths and 121 injuries. Reports fell to 15 in the last 12 months from 45 in the previous 12. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
126
Deaths reported
0
Injuries reported
120
Recalls naming it
0

Summary

The FDA's MAUDE database contains 126 adverse event reports that list this exact device (DI 10081317027840), received between Jul 2, 2021 and Sep 16, 2026 — classified as 120 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 5 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 5 reportsApr 2024: 1 reportsMay 2024: 9 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 6 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 5 reportsDec 2024: 7 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 5 reportsApr 2025: 1 reportsMay 2025: 8 reportsJun 2025: 8 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 4 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem113
  • Material Rupture11
  • Migration5
  • Appropriate Device Problem Term/Code Not Available4
  • Use of Device Problem1

Patient problems

  • Capsular Contracture95
  • Hematoma13
  • Deformity/ Disfigurement8
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available7
  • No Clinical Signs, Symptoms or Conditions6
  • Seroma2
  • Breast Discomfort/Pain1
  • Impaired Healing1

Recent adverse event reports

The 25 most recent of 126 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that a patient with MENTOR MemoryGel Xtra Breast Implants experienced capsular contracture Baker Grade III on the right side post-operatively, which was confirmed during a physical exam. As a result, the patient underwent device removal.

    Patient problems: Capsular Contracture

    MAUDE report 50787624 ↗
  • InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 35-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH 375CC ENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE; BAKER GRADE III ON HER RIGHT SIDE POSTOPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT RIGHT SIDE REPLACEMENT SURGERY WITH CATALOG NUMBER SMHX375; SERIAL NUMBER (B)(6) ON (B)(6) 2026.

    Patient problems: Capsular Contracture

    MAUDE report 24966629 ↗
  • InjuryMar 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE (BAKER GRADE 3) ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON AN UNSPECIFIED DATE.

    Patient problems: Capsular Contracture

    MAUDE report 24664835 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH MENTOR MEMORYGEL XTRA, 375CC SILICONE PROSTHESIS EXPERIENCED LEFT CAPSULAR CONTRACTURE GRADE III POST OPERATION. THE ISSUE WAS DIAGNOSED THROUGH PHYSICAL EXAMINATION. AS A RESULT, THE PATIENT UNDERWENT UNILATERAL REPLACEMENT WITH SN: (B)(6) ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 24381762 ↗
  • InjuryFeb 18, 2026Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE (BAKER GRADE 3) ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.

    Patient problems: Capsular Contracture

    MAUDE report 24378486 ↗
  • InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WITH MENTOR MEMORYGEL XTRA BREAST IMPLANTS EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE III ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT IS SCHEDULED FOR REMOVAL ON (B)(6) 2026.

    Patient problems: Capsular Contracture

    MAUDE report 24280494 ↗
  • InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE3-4 ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.

    Patient problems: Capsular Contracture

    MAUDE report 24259597 ↗
  • InjuryDec 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH A MENTOR GEL BREAST PROSTHESES DEVELOPED HEMATOMA IN BOTH BREASTS POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR GEL BREAST PROSTHESES. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT BREAST PROSTHESIS.

    Patient problems: Hematoma

    MAUDE report 23936210 ↗
  • InjuryDec 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH A MENTOR GEL BREAST PROSTHESES DEVELOPED HEMATOMA IN BOTH BREASTS POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR GEL BREAST PROSTHESES. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT BREAST PROSTHESIS.

    Patient problems: Hematoma

    MAUDE report 23936204 ↗
  • InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE 3 ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23707584 ↗
  • InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE IV). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 23652353 ↗
  • InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE IV). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THE REPLACEMENT DEVICES WERE TWO MENTOR GEL IMPLANTS. THIS MEDWATCH FORM IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 23652351 ↗
  • InjuryNov 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON THE RIGHT BREAST. POST-OPERATIVELY, THE PATIENT DEVELOPED RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE IV). AS A RESULT, THE PATIENT HAS BEEN SCHEDULED FOR BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23482500 ↗
  • InjuryNov 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT ON THE LEFT BREAST. POST-OPERATIVELY, THE PATIENT DEVELOPED LEFT BREAST HEMATOMA. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THE REPLACEMENT DEVICE WAS A MENTOR GEL IMPLANT.

    Patient problems: Hematoma

    MAUDE report 23480206 ↗
  • InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION SURGERY WITH IMPLANTATION OF A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH RIGHT BREAST BAKER GRADE III CAPSULAR CONTRACTURE USING ULTRASOUND IMAGING. AS A RESULT, THE PATIENT UNDERWENT RIGHT-SIDED EXCHANGE WITH A MENTOR GEL IMPLANT ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23066709 ↗
  • InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR MEMORYGEL XTRA BREAST IMPLANT 375CC GEL BREAST PROSTHESIS DEVELOPED BAKER GRADE III CAPSULAR CONTRACTURE IN HER RIGHT BREAST POST IMPLANTATION. THE CAPSULAR CONTRACTURE WAS DIAGNOSED VIA A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION AND REPLACEMENT WITH A GEL BREAST PROSTHESIS ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23064074 ↗
  • InjurySep 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH TWO MENTOR MEMORYGEL XTRA BREAST IMPLANTS EXPERIENCED BILATERAL CAPSULAR CONTRACTURE (BAKER GRADE 3) POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23017673 ↗
  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH RIGHT BREAST BAKER GRADE IV CAPSULAR CONTRACTURE BY A MEDICAL PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT IMPLANT REMOVAL ON (B)(6) 2025.

    Patient problems: Capsular Contracture; Impaired Healing

    MAUDE report 22739116 ↗
  • InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 375CC (LEFT) MENTOR MEMORYGEL XTRA BREAST IMPLANT AND A 330CC (RIGHT) MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE (BAKER GRADE 3) POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025 AND REPLACED WITH A 295CC MENTOR MEMORYGEL XTRA BREAST IMPLANT (CATALOG NUMBER SMPX295) ON THE LEFT AND A 270CC MENTOR MEMORYGEL XTRA BREAST IMPLANT (CATALOG NUMBER SMPX270) ON THE RIGHT SIDE. THIS REPORT IS FOR THE LEFT SIDE DEVICE.

    Patient problems: Capsular Contracture

    MAUDE report 22671919 ↗
  • MalfunctionJul 30, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT BEING USED IN A PRIMARY BREAST AUGMENTATION WAS FOUND TO BE RUPTURED DURING OPERATION. NO ADVERSE EVENT WAS EXPERIENCED BY THE PATIENT. THE SURGEON USED BACKUP DEVICES, AND THERE WERE NO REPORTED PATIENT CONSEQUENCES OR PROCEDURAL DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22669058 ↗
  • InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT AND EXPERIENCED A CAPSULAR CONTRACTURE (BAKER GRADE 3) ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2024 AND REPLACED WITH A 375CC MENTOR MEMORYGEL XTRA BREAST IMPLANT (CATALOG NUMBER SMHX375) WITH SERIAL NUMBER (B)(6).

    Patient problems: Capsular Contracture

    MAUDE report 22667820 ↗
  • MalfunctionJun 27, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT AN UNKNOWN GEL IMPLANT BEING USED FOR A BREAST SURGERY RUPTURED INTRAOPERATIVELY. THERE WERE NO PATIENT CONSEQUENCES OR ADDITIONAL INFORMATION REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22351847 ↗
  • InjuryJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION WITH TWO MENTOR MEMORYGEL XTRA BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA ULTRASOUND, WITH BILATERAL BREAST CAPSULAR CONTRACTURES (BAKER GRADE III). AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2025. THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 22294445 ↗
  • InjuryJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH MENTOR MEMORYGEL XTRA BREAST IMPLANT 375CC GEL BREAST PROSTHESES DEVELOPED BAKER GRADE III CAPSULAR CONTRACTURE IN BOTH BREASTS POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL XTRA BREAST IMPLANT 295CC GEL BREAST PROSTHESES ON (B)(6) 2025. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 22294129 ↗
  • InjuryJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION PROCEDURE WITH MENTOR MEMORYGEL XTRA BREAST IMPLANT 375CC GEL BREAST PROSTHESES DEVELOPED BAKER GRADE III CAPSULAR CONTRACTURE IN BOTH BREASTS POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL XTRA BREAST IMPLANT 295CC GEL BREAST PROSTHESES ON (B)(6) 2025. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 22290078 ↗

Other Prosthesis, breast, noninflatable, internal, silicone gel-filled devices

All Prosthesis, breast, noninflatable, internal, silicone gel-filled devices →