1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10081317028274), received between Feb 10, 2022 and Jan 16, 2026 — classified as 13 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (13 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K161176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 18, 2026Class IIOpen, Classified
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
Z-1686-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Rupture13
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Deformity/ Disfigurement12
- No Clinical Signs, Symptoms or Conditions1
- Post Operative Wound Infection1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryJan 16, 2026Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION SURGERY WITH IMPLANTATION OF A 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE BREAST TISSUE EXPANDER PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A DEFLATED RIGHT EXPANDER BY A MEDICAL PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT RIGHT-SIDED EXCHANGE AND LEFT-SIDED IMPLANTATION WITH UNDISCLOSED BREAST TISSUE EXPANDERS ON (B)(6) 2025. POST-REVISION, THE SURGEON OBSERVED THE EXPLANTED DEVICE WAS COLLAPSED BUT COULD NOT FIND ANY TEARS OR LEAKS ON THE EXPANDER.
Patient problems: Deformity/ Disfigurement
MAUDE report 24099384 ↗InjurySep 11, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST PROSTHESIS IMPLANTATION SURGERY WITH A 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE SALINE BREAST PROSTHESIS ON AN UNSPECIFIED BREAST SIDE. POST-OPERATIVELY, THE PATIENT SUFFERED BREAST IMPLANT DEFLATION. AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 23028470 ↗InjuryMay 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST RECONSTRUCTION PROCEDURE WITH A MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE 750CC TISSUE EXPANDER THAT DEFLATED POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION AND REPLACEMENT WITH A MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE 850CC TISSUE EXPANDER ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 22064471 ↗InjuryJun 21, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST RECONSTRUCTION WITH TWO 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE SALINE BREAST PROSTHESES. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH POSSIBLE BREAST IMPLANT LEAKING. AS A RESULT, THE PATIENT WAS SCHEDULED FOR BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2024.
Patient problems: Deformity/ Disfigurement
MAUDE report 19580997 ↗MalfunctionApr 2, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT AN UNSPECIFIED BREAST SURGERY WITH A MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE, 770CC, TISSUE EXPANDER EXPERIENCED UNKNOWN-SIDED DEFLATION INTRAOPERATIVE. NO PATIENT INVOLVEMENT OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19028538 ↗InjuryNov 14, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH TWO UNSPECIFIED MENTOR TISSUE EXPANDERS EXPERIENCED BILATERAL DEFLATIONS POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2023. THIS REPORT IS FOR THE RIGHT SIDE DEVICE.
Patient problems: Deformity/ Disfigurement
MAUDE report 18133573 ↗InjuryNov 14, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT IMPLANTED WITH TWO UNSPECIFIED MENTOR TISSUE EXPANDERS EXPERIENCED BILATERAL DEFLATIONS POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2023. THIS REPORT IS FOR THE LEFT SIDE DEVICE.
Patient problems: Deformity/ Disfigurement
MAUDE report 18133570 ↗InjuryAug 14, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A 40-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT AN UNSPECIFIED BREAST SURGERY WITH A MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE 750CC TISSUE EXPANDER THAT DEFLATED POST IMPLANTATION. DEFLATION OF THE PATIENT¿S RIGHT TISSUE EXPANDER WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION AND REPLACEMENT WITH A MENTOR MEMORYGEL BOOST BREAST IMPLANT 520CC GEL BREAST PROSTHESIS ON (B)(6) 2023.
Patient problems: Deformity/ Disfigurement
MAUDE report 17545853 ↗InjuryMay 24, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH A 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE BREAST TISSUE EXPANDER. POST-OPERATIVELY, THE PATIENT EXPERIENCED A DEFLATION OF HER RIGHT PROSTHESIS, WHICH WAS CONFIRMED BY A MEDICAL PROFESSIONAL. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. SEVERAL FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.
Patient problems: Deformity/ Disfigurement
MAUDE report 16994394 ↗InjuryMay 3, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH IMPLANTATION OF A 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE TISSUE EXPANDER PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A DEFLATED RIGHT TISSUE EXPANDER BY A MEDICAL PROFESSIONAL. AS A RESULT, THE PATIENT UNDERWENT UNILATERAL REMOVAL AND REPLACEMENT WITH AN UNDISCLOSED RIGHT-SIDED PROSTHESIS ON (B)(6) 2023.
Patient problems: Deformity/ Disfigurement
MAUDE report 16860610 ↗InjuryMar 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A 65-YEAR-OLD AFRICAN AMERICAN PATIENT WHO UNDERWENT AN UNSPECIFIED BREAST SURGERY WITH A MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE 750CC TISSUE EXPANDER DEVELOPED AN INFECTION IN HER RIGHT BREAST POST IMPLANTATION. ERYTHEMA AND DRAINAGE OF THE RIGHT BREAST MOUND WERE OBSERVED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION WITH NO REPLACEMENT DEVICES ON (B)(6) 2023.
Patient problems: Post Operative Wound Infection
MAUDE report 16469305 ↗InjuryMar 1, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A 48-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION WITH A 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE TISSUE EXPANDER AND EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2023.
Patient problems: Deformity/ Disfigurement
MAUDE report 16460842 ↗InjuryNov 23, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT BREAST RECONSTRUCTION WITH A 750CC SMOOTH HIGH PROFILE MENTOR ARTOURA PLUS TISSUE EXPANDER EXPERIENCED DEFLATION ON AN UNKNOWN SIDE POST PROCEDURE. AS A RESULT, EXPLANT WAS PERFORMED ON (B)(6) 2022.
Patient problems: Deformity/ Disfigurement
MAUDE report 15854133 ↗InjuryFeb 10, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT A (B)(6) YEAR-OLD ASIAN FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH A 750CC MENTOR ARTOURA PLUS, SMOOTH, HIGH PROFILE TISSUE EXPANDER AND EXPERIENCED A DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED VIA AN OFFICE EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2022.
Patient problems: Deformity/ Disfigurement
MAUDE report 13507373 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K161176. One submission often covers many models, so check the product and lot details in each record.
- Z-1687-2026 · Feb 18, 2026 · Open, ClassifiedArtoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc
Other Expander, skin, inflatable devices
- B.BRAUN 7B3050B. Braun Medical Inc. · EXPANDER, SKIN, INFLATABLE1 recall40 reports
- MENTOR CPX 4 Breast Tissue Expander with Suture Tabs 3549314Mentor Texas L.P. · EXPANDER, SKIN, INFLATABLE1 recall22 reports
- MENTOR CPX4 PLUS, Smooth, Medium Height SCPX146MHMentor Texas L.P. · Expander, skin, inflatable1 recall18 reports
- Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MX-13-TAllergan, Inc. · Expander, Skin, Inflatable1 recall16 reports
- MENTOR ARTOURA PLUS, Smooth, Ultra High Profile SDC135UHMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- MENTOR ARTOURA PLUS, Smooth, Ultra High Profile SDC120UHMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MV-14-TAllergan, Inc. · Expander, Skin, Inflatable1 recall13 reports
- MENTOR CPX4 PLUS, Smooth, Medium Height SCPX127MHMentor Texas L.P. · Expander, skin, inflatable1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page