MENTOR MemoryGel BOOST Breast Implant SMHB545Recalls & reported problems

Mentor Texas L.P. · Smooth Moderate High Profile, 545cc |Smooth|Round|Silicone| · Class 3 · Implant

All models of MENTOR MemoryGel Breast Implants (368) · MENTOR MemoryGel BOOST Breast Implant (53)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
21
Deaths reported
0
Injuries reported
19
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 10081317031151), received between Sep 19, 2024 and Mar 30, 2026 — classified as 19 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem15
  • Material Rupture4
  • Migration3
  • Material Deformation2
  • Appropriate Device Problem Term/Code Not Available1
  • Material Discolored1

Patient problems

  • Capsular Contracture14
  • Deformity/ Disfigurement6
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Material Deformation

    IT WAS REPORTED THAT A MENTOR MEMORYGEL BOOST BREAST IMPLANT WAS FOUND TO CONTAIN SHELL IRREGULARITIES DURING AN OPERATION. NO ADVERSE EVENT WAS EXPERIENCED BY THE PATIENT. THE SURGEON USED BACKUP DEVICES, AND THERE WERE NO REPORTED PATIENT CONSEQUENCES OR PROCEDURAL DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24720527 ↗
  • InjuryFeb 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT A BREAST RECONSTRUCTION REVISION PROCEDURE WITH A MENTOR MEMORYGEL BOOST BREAST IMPLANT GEL BREAST PROSTHESIS DEVELOPED BAKER GRADE IV CAPSULAR CONTRACTURE IN HER RIGHT BREAST POST IMPLANTATION. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION WITH NO REPLACEMENT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 24310003 ↗
  • InjuryJan 16, 2026Migration

    A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION PRIMARY WITH MENTOR MEMORYGEL BOOST, 545CC SILICONE PROSTHESIS EXPERIENCED LEFT SIDED CAPSULAR CONTRACTURE GRADE III, AND RIGHT SIDED PSEUDOPTOSIS, BREAST ASYMMETRY POSTOPERATION. THE ISSUE WAS DIAGNOSED THROUGH PHYSICAL EXAMINATION. AS A RESULT, THE PATIENT UNDERWENT RIGHT SIDED IMPLANT REPLACEMENT WITH MENTOR 650CC HIGH PROFILE SMOOTH SILICONE GEL IMPLANT, AND LEFT REVISION PERI AREOLAR SCAR ON (B)(6) 2025. THIS REPORT RELATES TO THE RIGHT SIDE.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 24096369 ↗
  • InjuryJan 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 32-YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE (BAKER GRADE UNKNOWN) ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025 AND REPLACED WITH A 775CC MENTOR MEMORYGEL BOOST BREAST IMPLANT.

    Patient problems: Capsular Contracture

    MAUDE report 24031303 ↗
  • InjuryJan 2, 2026Adverse Event Without Identified Device or Use Problem

    A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION PRIMARY WITH MENTOR MEMORYGEL BOOST, 545CC SILICONE PROSTHESIS EXPERIENCED LEFT SIDED CAPSULAR CONTRACTURE GRADE III POST OPERATION. THE ISSUE WAS DIAGNOSED THROUGH PHYSICAL EXAMINATION. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REPLACEMENTS WITH UNKNOWN GEL PROSTHESIS ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23948342 ↗
  • InjuryDec 16, 2025Material Discolored; Material Rupture; Appropriate Device Problem Term/Code Not Available; Material Discolored; Material Rupture; Appropriate Device Problem Term/Code Not Available

    A FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH MENTOR MEMORYGEL BOOST 545CC SILICONE PROSTHESIS EXPERIENCED LEFT SIDED SILENT RUPTURE, BUBBLES, AND DISCOLORATION IN THE IMPLANT POST OPERATION. THE DOCTOR CONFIRMED THE RUPTURE. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REPLACEMENTS PER FOLLOWING: R) REF: SMHB545 SN: (B)(6), L) REF: SMHB545 SN: (B)(6) ON (B)(6) 2025.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 23819363 ↗
  • MalfunctionDec 5, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT BEING USED IN A BREAST AUGMENTATION PROCEDURE WAS FOUND TO BE DEFECTIVE DURING THE OPERATION. THE IMPLANT HAD A STICKY FEEL TO IT AND WAS CRINKLED. NO ADVERSE EVENT WAS EXPERIENCED BY THE PATIENT. THE SURGEON USED A BACKUP DEVICE, AND THERE WERE NO REPORTED PATIENT CONSEQUENCES OR PROCEDURAL DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23729809 ↗
  • InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 49-YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE (BAKE GRADE 3) ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2026.

    Patient problems: Capsular Contracture

    MAUDE report 23662859 ↗
  • InjuryNov 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BAKER GRADE IV CAPSULAR CONTRACTURE DURING A FOLLOW-UP PHYSICAL EXAM WITH ASSOCIATED BREAST FIRMNESS AND TIGHTNESS IN ADDITION TO HIGH IMPLANTS. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. FOLLOW-UPS ARE BEING CONDUCTED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.

    Patient problems: Capsular Contracture

    MAUDE report 23489995 ↗
  • InjuryNov 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BILATERAL BAKER GRADE IV CAPSULAR CONTRACTURE DURING A FOLLOW-UP PHYSICAL EXAM WITH ASSOCIATED BREAST FIRMNESS AND TIGHTNESS IN ADDITION TO HIGH IMPLANTS. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. FOLLOW-UPS ARE BEING CONDUCTED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.

    Patient problems: Capsular Contracture

    MAUDE report 23489991 ↗
  • InjuryOct 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 32-YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT AND EXPERIENCED CAPSULAR CONTRACTURE (BAKER GRADE 4) ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23426166 ↗
  • InjurySep 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH LEFT BREAST CAPSULAR CONTRACTURE (BAKER GRADE III). AS A RESULT, THE PATIENT HAS BEEN SCHEDULED FOR BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 23024094 ↗
  • InjurySep 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 57-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION WITH TWO 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANTS AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE (BAKER GRADE 4 ON THE LEFT SIDE AND BAKER GRADE UNKNOWN ON THE RIGHT SIDE) POST-OPERATIVELY. THE PATIENT HAD PROGRESSIVE HARDENING, SUPERIOR DISPLACEMENT, FIRMNESS, AND TENDER TO MANIPULATION OF THE LEFT BREAST. THE PATIENT ALSO EXPERIENCED FIRMNESS ON THE RIGHT SIDE. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025, AND REPLACED WITH 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANTS (CATALOG NUMBER SMHB545) WITH LOT NUMBER 2045659 AND 2081966. THIS REPORT IS FOR THE RIGHT SIDE DEVICE.

    Patient problems: Capsular Contracture

    MAUDE report 23020530 ↗
  • InjurySep 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 57-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION WITH TWO 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANTS AND EXPERIENCED BILATERAL CAPSULAR CONTRACTURE (BAKER GRADE 4 ON THE LEFT SIDE AND BAKER GRADE UNKNOWN ON THE RIGHT SIDE) POST-OPERATIVELY. THE PATIENT HAD PROGRESSIVE HARDENING, SUPERIOR DISPLACEMENT, FIRMNESS, AND TENDER TO MANIPULATION OF THE LEFT BREAST. THE PATIENT ALSO EXPERIENCED FIRMNESS ON THE RIGHT SIDE. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025, AND REPLACED WITH 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANTS (CATALOG NUMBER SMHB545) WITH LOT NUMBER 2045659 AND 2081966. THIS REPORT IS FOR THE LEFT SIDE DEVICE.

    Patient problems: Capsular Contracture

    MAUDE report 23020529 ↗
  • InjuryApr 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 33-YEAR-OLD FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH SMOOTH MOD HIGH BOOST, 545CC ON BOTH SIDES AND EXPERIENCED CAPSULAR CONTRACTURE; BAKER GRADE III ON HER LEFT SIDE POSTOPERATIVELY; DIAGNOSED AFTER PHYSICAL EXAM. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT IS SCHEDULED FOR LEFT SIDE REPLACEMENT SURGERY ON (B)(6), 2025. IT WAS FURTHER REPORTED THAT THE SURGERY WAS PERFORMED ON PATIENT'S RIGHT SIDE TO CORRECT THE ASYMMETRY PRODUCED BY THE CAPSULAR CONTRACTURE ON THE LEFT SIDE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE FOR THE RIGHT SIDE. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 21872834 ↗
  • InjuryApr 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A 33-YEAR-OLD FEMALE PATIENT UNDERWENT REVISION BREAST AUGMENTATION WITH SMOOTH MOD HIGH BOOST, 545CC ON BOTH SIDES AND EXPERIENCED CAPSULAR CONTRACTURE; BAKER GRADE III ON HER LEFT SIDE POSTOPERATIVELY; DIAGNOSED AFTER PHYSICAL EXAM. THE PATIENT HAS CONSULTED WITH THE HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT IS SCHEDULED FOR LEFT SIDE REPLACEMENT SURGERY ON (B)(6) 2025. IT WAS FURTHER REPORTED THAT THE SURGERY WAS PERFORMED ON PATIENT'S RIGHT SIDE TO CORRECT THE ASYMMETRY PRODUCED BY THE CAPSULAR CONTRACTURE ON THE LEFT SIDE. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE FOR THE RIGHT SIDE. THIS MEDWATCH REPORT IS FOR THE PATIENT¿S LEFT-SIDED DEVICE.

    Patient problems: Capsular Contracture

    MAUDE report 21872832 ↗
  • InjuryMar 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT WHO UNDERWENT BREAST AUGMENTATION REVISION WITH A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT EXPERIENCED LEFT SIDED BAKER GRADE III CAPSULAR CONTRACTURE POST PROCEDURE. AS A RESULT, EXPLANT IS PLANNED FOR (B)(6) 2025.

    Patient problems: Capsular Contracture

    MAUDE report 21663588 ↗
  • InjuryJan 14, 2025Migration; Migration

    IT WAS REPORTED THAT A 54-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION WITH A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT AND EXPERIENCED DEVICE MIGRATION WITH ASYMMETRY ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2024, AND REPLACED WITH A MENTOR BREAST PROSTHESIS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 21150362 ↗
  • InjuryDec 11, 2024Material Rupture; Migration; Material Rupture; Migration

    IT WAS REPORTED THAT A 54-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST RECONSTRUCTION WITH A 545CC MENTOR MEMORYGEL BOOST BREAST IMPLANT AND EXPERIENCED A RUPTURE ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED DURING SURGERY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON(B)(6) 2024, AND REPLACED WITH A MENTOR BREAST PROSTHESIS.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 20909116 ↗
  • InjurySep 19, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH MENTOR MEMORYGEL BOOST BREAST IMPLANTS 545CC AND EXPERIENCED BILATERAL RUPTURE ALONG WITH CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED VIA MRI. AS A RESULT, THE PATIENT IS SCHEDULED FOR REMOVAL ON (B)(6) 2024. THIS REPORT IS FOR THE LEFT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture; Deformity/ Disfigurement

    MAUDE report 20259835 ↗
  • InjurySep 19, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH MENTOR MEMORYGEL BOOST BREAST IMPLANTS 545CC AND EXPERIENCED BILATERAL RUPTURE ALONG WITH CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED VIA MRI. AS A RESULT, THE PATIENT IS SCHEDULED FOR REMOVAL ON (B)(6) 2024. THIS REPORT IS FOR THE RIGHT BREAST PROSTHESIS.

    Patient problems: Capsular Contracture

    MAUDE report 20259656 ↗

Other Prosthesis, breast, noninflatable, internal, silicone gel-filled devices

All Prosthesis, breast, noninflatable, internal, silicone gel-filled devices →