POSEY 2510Recalls & reported problems
Posey Products, LLC · POSEY ECONOMY LIMB · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10190676005298), received between Aug 18, 2022 and Aug 23, 2024 — classified as 1 death, 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Cardiac Arrest1
- Insufficient Information1
- Irritability1
- Loss of consciousness1
- Ventilator Dependent1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2024Break; Break
COMPLAINTS REPORTED PATIENT TORE THE TIES OFF THE RESTRAINT WITH BRUTE FORCE. THE PATIENT WAS ABLE TO SELF EXTUBATE. RESULTED IN PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 20057300 ↗MalfunctionAug 23, 2024Break; Break
COMPLAINTS REPORTED ECONOMY LIMB HOLDER SEPARATED AT THE POINT WHERE THE TIEDOWN CONNECTS TO THE SOFT WRIST RESTRAINT. NYLON RIPPED EASILY FROM SOFT LIMB RESTRAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20055282 ↗InjuryAug 23, 2024Break; Break
COMPLAINTS REPORTED WRIST RESTRAINTS ON PATIENT TO PREVENT SELF EXTUBATION/CLINICAL LINE REMOVAL, MATERIALS RIPPED AT VELCRO/PADDING WHEN PATIENT PULLED AGAINST DEVICE, DEVICE FAILURE ALLOWED PATIENT TO REMOVE CRITICAL LINE RESULTING IN CARDIAC ARREST.
Patient problems: Cardiac Arrest
MAUDE report 20053997 ↗MalfunctionAug 23, 2024Break; Break
A CUSTOMER INFORMED US OF A CRITICAL ISSUE IDENTIFIED WITH THE RESTRAINT WHERE THEY WILL NO LONGER BE USING PART NUMBER 2510 WILL BE REPLACING WITH PART NUMBER 2532.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20051238 ↗DeathAug 18, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
WE RECEIVED AN FDA MEDWATCH INDICATING THAT A PATIENT WHO WAS VENTILATOR-DEPENDENT WITH PAST MEDICAL HISTORY OF ACUTE LEFT MCA STROKE WITH RIGHT HEMIPARESIS AND SEVERE APHASIA WAS ADMITTED FROM A TRANSITIONAL CARE FACILITY FOR AGITATION AND CLOSE OBSERVATION. PATIENT SELF-REMOVED THEIR TRACHEOSTOMY TUBE AND STAFF WAS ABLE TO REPLACE TRACH. MEDS WERE ADMINISTERED TO CALM PATIENT DOWN FROM AN AGITATED STATE. THE PATIENT WAS TRANSFERRED TO THE ICU FOR CLOSE MONITORING WHERE BILATERAL WRIST RESTRAINTS WERE INITIATED AS PATIENT WAS PULLING AT TUBES/TRACH. THE RIGHT RESTRAINT WAS REMOVED DUE TO RIGHT HEMIPARESIS. THE PATIENT WAS TRANSFERRED OUT OF THE ICU. THE PATIENT WAS AGITATED, KICKING, CRYING AND YANKING AT THE RESTRAINT TRYING TO PULL THE TRACH OUT. THE PATIENT WAS REPOSITIONED AND THE RESTRAINT WAS NOTED AS BEING REALLY TIGHT AND THE NURSE HAD RETIED THE RESTRAINT. PATIENT WAS REDIRECTED, REORIENTED, NO LONGER PULLING AT TRACH. RESTING COMFORTABLY IN BED. AT THE TIME OF THE NEXT HOURLY CHECK, THE PATIENT WAS FOUND UNCONSCIOUS, CPR INITIATED AND PATIENT EXPIRED. PATIENT HAD INTENTIONALLY PULLED OUT TRACH. NURSE REPORTED THAT THE RESTRAINT WAS STILL ON THE PATIENT'S WRIST AND WAS TI
Patient problems: Irritability; Loss of consciousness; Ventilator Dependent
MAUDE report 15250684 ↗
Other Restraint, Protective devices
- DeRoyal M8138Deroyal Industries, Inc. · Restraint, protective1 recall1 report
- DeRoyal M8125Deroyal Industries, Inc. · Restraint, protective1 recall0 reports
- DeRoyal M8137Deroyal Industries, Inc. · Restraint, protective1 recall0 reports
- DeRoyal M8136Deroyal Industries, Inc. · Restraint, protective1 recall0 reports
- DeRoyal M8124Deroyal Industries, Inc. · Restraint, protective1 recall0 reports
- POSEY 2532Posey Products, LLC · Restraint, Protective36 reports
- POSEY 2551Posey Products, LLC · Restraint, Protective14 reports
- POSEY 2799Posey Products, LLC · Restraint, Protective9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page