AirLife™ 2K8000Recalls & reported problems

Airlife · AirLife™ Adult Manual Resuscitator Oxygen Reservoir Bag, without Mask · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 10, 2024Open, Classified

    Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

    Z-1057-2024 · Root cause: Process control · FDA record ↗

Other Anesthesia breathing circuit kit (adult & pediatric) devices

All Anesthesia breathing circuit kit (adult & pediatric) devices →