Vital Signs™ A4U12214Recalls & reported problems
Airlife · Vital Signs™ Adult Anesthesia Breathing Circuit · Class 1
All models of Vital Signs (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10190752119871), received between Jul 12, 2022 and Aug 24, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Defective Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Defective Component1
- Fitting Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2022Fitting Problem; Fitting Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE ADULT MASK IN A4U12214 - ANES CIRCUIT, ADULT, 108 IN EXP, 3L BAG DOES NOT PROVIDE AN APPROPRIATE SEAL AROUND THE PATIENT'S AIRWAY DESPITE SEVERAL ATTEMPTS TO RE INFLATE. THERE WAS PATIENT INVOLVEMENT. AT THIS TIME, THE CUSTOMER HAS NOT PROVIDED ANY INFORMATION REGARDING PATIENT HARM OR INJURY ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15286996 ↗MalfunctionJul 12, 2022Defective Component; Defective Component
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE MASK FOR VITAL SIGNS® ADULT ANESTHESIA BREATHING CIRCUIT DOES NOT PROVIDE AN APPROPRIATE SEAL AROUND THE PATIENT'S AIRWAY AND THIS POSSESSES A GREAT RISK TO PATIENT CARE AND OUTCOMES. THERE WAS PATIENT INVOLVEMENT BUT NO KNOWN CONSEQUENCE OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14988154 ↗
Other Circuit, breathing (w connector, adaptor, y piece) devices
- Medline Industries, Inc. DYNJ905503FMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. PHS972096014BMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. DYNJ0185290GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)3 recalls0 reports
- Medline Industries, Inc. DYNJ905503GMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJAA262AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYNJ905567AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ86657AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
- Medline Industries, Inc. DYKMBNDL199AMedline Industries, Inc. · Circuit, breathing (w connector, adaptor, y piece)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page