Skip to content

Vital Signs™ AFPXXXXXRecalls & reported problems

Airlife · Vital Signs™ Limb-O™ Single Limb Anesthesia Breathing Circuit · Class 1

All models of Vital Signs (48)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10190752122338), received between Feb 28, 2020 and Jun 23, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement2
  • Loose or Intermittent Connection1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2026Unintended Movement

    ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 1314417-2026-00042 FOR THE FIRST REPORT. REFER TO 1314417-2026-00043 FOR THE SECOND REPORT. REFER TO 1314417-2026-00044 FOR THE THIRD REPORT. IT WAS REPORTED, THAT THE CONNECTION CAME APART ON THE ADULT ANESTHESIA CIRCUIT DUE TO THE RUBBER BAND SOMEHOW STARTED TO ROLL UP DUE TO EXTERNAL PRESSURE. IT WAS ADDITIONALLY REPORTED, THE MEDICAL STAFF ACTED QUICKLY, THE TUBE WAS REPLACED, AND NO HARM CAME TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25602608 ↗
  • MalfunctionJun 23, 2026Unintended Movement

    ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 1314417-2026-00042 FOR THE FIRST REPORT. REFER TO 1314417-2026-00043 FOR THE SECOND REPORT. REFER TO 1314417-2026-00044 FOR THE THIRD REPORT. IT WAS REPORTED, THAT THE CONNECTION CAME APART ON THE ADULT ANESTHESIA CIRCUIT DUE TO THE RUBBER BAND SHIFTING BACKWARDS. IT WAS ADDITIONALLY REPORTED, THE MEDICAL STAFF ACTED QUICKLY, THE TUBE WAS REPLACED, AND NO HARM CAME TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25602607 ↗
  • MalfunctionOct 24, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE BLUE CONNECTOR ON THE INHALE SIDE AND THE WHITE CONNECTOR ON THE EXHALATION SIDE (AFPXXXXX-ANES CIRCUIT, ADULT, 108 IN LIMBO) ARE CONNECTED AND USED, RESPECTIVELY. THE LEAK TEST PASSED WITHOUT ANY PROBLEMS. HOWEVER, THE WHITE CONNECTOR LOOSENED DURING THE PROCEDURE. HOWEVER, THE CONNECTOR CONTINUED TO PULL OUT IMMEDIATELY WHEN PULLED UNTIL THE END OF THE OP. THERE WAS NO PARTICULAR PROBLEM WITH THE BLUE CONNECTOR. ON FOLLOW-UP, THE CUSTOMER CONFIRMED THAT THE OPERATION WENT SUCCESSFULLY, BUT UNTIL THE END OF THE OPERATION, THE TUBE KEPT LOOSE WHEN IT WAS PULLED. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT HARM/INJURY ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17993374 ↗
  • MalfunctionFeb 28, 2020Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    THE CUSTOMER REPORTED THAT THE VITAL SIGNS¿ LIMB-O¿ SINGLE LIMB ANESTHESIA BREATHING CIRCUIT IS LEAKING WHILE CONNECTED TO PATIENT DURING SURGERY HOWEVER THE CUSTOMER CONFIRMED THAT NO PATIENT HARM WAS ASSOCIATED ON THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9771807 ↗

Other Circuit, breathing (w connector, adaptor, y piece) devices

All Circuit, breathing (w connector, adaptor, y piece) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page